The Application Notebook
Agilent
This application note shows that unspecific reaction of adenosine triphosphate can be completely prevented due to the iron and steel-free design using the Agilent 1260 Infinity Bio-inert Quaternary LC System.
Severe peak tailing of phosphate compounds is a well described issue in HPLC analysis. Interaction between stainless steel and phosphate groups were described by Liu et al. leading to the formation of phosphopeptide-Fe(III) complexes (1). The 1260 Infinity Bio-inert Quaternary LC System provides a complete metalfree sample flow path throughout the system. With this system, the user can analyse a variety of phosphate compounds without the emersion of peak tailing or other unspecific reaction due to phosphate–iron complexes.
Solvents: 10 mM ammonium acetate with increasing amount of methanol
Sample: Adenosine triphosphate (ATP), solved in H2Odd (5 mg/mL)
A PEEK restriction capillary was used instead of a stainless steel column.
Significant peak tailing could be observed for ATP analysis in a stainless steel based system, the 1260 Infinity Quaternary LC System. With increasing amount of organic mobile phase, the retention of the phosphate compound was increasing to a huge extent, also resulting in relevant area reduction, see Figure 1.
Figure 1: ATP analysis on Agilent 1260 Infinity Quaternary System.
With the 1260 Infinity Bio-inert Quaternary LC System, the unspecific reaction of the used phosphate sample could be completely prevented, resulting in good peak shapes without substantial peak tailing or area reduction, see Figure 2.
Figure 2: ATP analysis on the Agilent 1260 Infinity Bio-inert Quaternary System.
Unspecific reaction and peak tailing of adenosine triphosphate could be completely prevented when using the Agilent 1260 Infinity Bioinert Quaternary LC System. Due to the iron and steelfree design of the 1260 Infinity Bio-inert Quaternary LC system, phosphate compounds can be analyzed without any issues regarding the formation of phosphate–iron complexes as found with stainless steel systems.
(1) Liu et al., Rapid Communications in Mass Spectrometry 19(19), 2747–2756 (2005).
Agilent Technologies Inc.
5301 Stevens Creek Blvd., Santa Clara, California 95051, USA
tel. (800) 227 9770 (Directory) fax (866) 497 1134
Website: www.agilent.com
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.