Analysis of Phosphate Compounds with Agilent 1260 Infinity Bio-inert Quaternary LC system

Article

The Application Notebook

The Application NotebookThe Application Notebook-09-02-2012
Volume 0
Issue 0

Agilent

This application note shows that unspecific reaction of adenosine triphosphate can be completely prevented due to the iron and steel-free design using the Agilent 1260 Infinity Bio-inert Quaternary LC System.

Severe peak tailing of phosphate compounds is a well described issue in HPLC analysis. Interaction between stainless steel and phosphate groups were described by Liu et al. leading to the formation of phosphopeptide-Fe(III) complexes (1). The 1260 Infinity Bio-inert Quaternary LC System provides a complete metalfree sample flow path throughout the system. With this system, the user can analyse a variety of phosphate compounds without the emersion of peak tailing or other unspecific reaction due to phosphate–iron complexes.

Experimental Conditions

Solvents: 10 mM ammonium acetate with increasing amount of methanol

Sample: Adenosine triphosphate (ATP), solved in H2Odd (5 mg/mL)

A PEEK restriction capillary was used instead of a stainless steel column.

Results

Significant peak tailing could be observed for ATP analysis in a stainless steel based system, the 1260 Infinity Quaternary LC System. With increasing amount of organic mobile phase, the retention of the phosphate compound was increasing to a huge extent, also resulting in relevant area reduction, see Figure 1.

Figure 1: ATP analysis on Agilent 1260 Infinity Quaternary System.

With the 1260 Infinity Bio-inert Quaternary LC System, the unspecific reaction of the used phosphate sample could be completely prevented, resulting in good peak shapes without substantial peak tailing or area reduction, see Figure 2.

Figure 2: ATP analysis on the Agilent 1260 Infinity Bio-inert Quaternary System.

Conclusions

Unspecific reaction and peak tailing of adenosine triphosphate could be completely prevented when using the Agilent 1260 Infinity Bioinert Quaternary LC System. Due to the iron and steelfree design of the 1260 Infinity Bio-inert Quaternary LC system, phosphate compounds can be analyzed without any issues regarding the formation of phosphate–iron complexes as found with stainless steel systems.

Reference

(1) Liu et al., Rapid Communications in Mass Spectrometry 19(19), 2747–2756 (2005).

Agilent Technologies Inc.

5301 Stevens Creek Blvd., Santa Clara, California 95051, USA

tel. (800) 227 9770 (Directory) fax (866) 497 1134

Website: www.agilent.com