This application note describes the method for detection of nanogram levels of explosive constituents in seawater samples (e.g., after detonation of unexploded ordnance devices). The improved chromatographic method was developed using the Agilent 1290 Infinity LC Method Development system.
Detlef Wilhelm1, Edgar Nägele2, and Udo Huber2, 1 AnaTox GmbH & Co. KG and 2 Agilent Technologies R&D and Marketing GmbH & Co. KG
This application note describes the method for detection of nanogram levels of explosive constituents in seawater samples (e.g., after detonation of unexploded ordnance devices). The improved chromatographic method was developed using the Agilent 1290 Infinity LC Method Development system.
Long after World War II, unexploded ordnance devices are sometimes found, not only in the ground but also near the coasts, such as in the Baltic Sea. These old munitions are often corroded and cannot be deactivated; sometimes, the only way to deactivate is with a controlled detonation. The residues of the detonation that can be found in the water must be monitored because of their toxicity. This application note describes the method for detection of nanogram levels of explosive constituents in seawater samples. The improved chromatographic method was developed using the Agilent 1290 Infinity LC Method Development system.
Figure 1: Detection of the 5 µg/ml standard with the 60-mm high sensitivity cell at 235 nm. Chromatographic conditions: Stationary phase: Stable Bond CN-RRHT (2.1 à 100 mm, 1.8 µm); Mobile phase A: Water; Mobile phase B: Acetonitrile; Gradient: at 0 min 20 % B, at 6 min 40 % B, at 7 min 40 % B; Flow rate: 0.5 mL/min; Column temp.: 30 °C; Inj. vol.: 1 µl; UV Detection: 214 nm/4, Ref.: 450 nm/80; Data rate: 20 Hz.
Experimental
LC System
For method development and analysis of the samples, an Agilent 1290 Infinity LC system was used. The system consists of: 1290 Infinity Binary pump with integrated vacuum degasser, 1290 Infinity high performance autosampler, 1290 Infinity Thermostatted column compartment, and 1290 Infinity Diode Array Detector with 10- and 60-mm Max Light Cartridge flow cell.
Analytes
Components: Octogen (HMX), Hexogen (RDX), 2,4-Dintrotoluol (2,4-DNT), 2,6-Dintrotoluol (2,6-DNT), 2,4,6-Trintrotoluol (TNT), Nitropenta (PETN), Nitroglycerine, 2,4,6-Trinitrophenylmethylnitramine (Tetryl), 2,3-Dinitrodimethylbutane (2,3-DNDMB-internal standard, eluting at 3.35 min in Figure 1).
The explosives analyzed in this application are very polar but not water-soluble substances, and are not well separated on typical RP18 columns. Therefore, different RP columns, including more polar materials, were systematically screened using the Agilent 1290 Infinity Method Development system. Great selectivity differences can be seen between the unpolar phases like Eclipse Plus C18, Eclipse Plus Phenyl-Hexyl, Poroshell EC-C18, and Stable Bond C18. Only the Phenyl-Hexyl column shows separation of more than four of the explosives. The polar RP-phases are more useful. In particular, the critical pair 2,4- and 2,6-dinitrotoluol is only completely separated on the Bonus RP and the SB-AQ, but with the disadvantage of co-elution of TNT and Tetryl. The best separation with a simple gradient was achieved with the Stable Bond CN column, and therefore, this phase was used for further optimization. The final method separates all explosives within 7 min. A further improvement in sensitivity was achieved by using the new 60-mm high sensitivity flow cell. Figure 1 shows the chromatograms of the 5 µg/ml standard with the 10 mm cell (data not shown) and the 60 mm high sensitivity cell. If the detection wavelength is changed from 235 nm to 214 nm, the influence of the eluent might be increased. However, the sensitivity for 2,3-DNDMB, nitropenta, and nitroglycerine is further increased.
The new Agilent 1290 Infinity LC is designed to provide the highest speed, resolution, and sensitivity. The transfer of a standard method to determine eight explosives on sub-2-µm RRHD or RRHT columns could be done "overnight" by identifying the best selectivity using the Agilent Method Development system. The results show that the analysis could be shortened to 7 min without any matrix influences.
Agilent Technologies, Inc.
5301 Stevens Creek Blvd., Santa Clara, CA 95051
tel: (877) 424-4536, fax: (408) 345-8474
Website: www.agilent.com
New Study Uses MSPE with GC–MS to Analyze PFCAs in Water
January 20th 2025Scientists from the China University of Sciences combined magnetic solid-phase extraction (MSPE) with gas chromatography–mass spectrometry (GC–MS) to analyze perfluoro carboxylic acids (PFCAs) in different water environments.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ion used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
A Guide To Finding the Ideal Syringe and Needle
January 20th 2025Hamilton has produced a series of reference guides to assist science professionals in finding the best-suited products and configurations for their applications. The Syringe and Needle Reference Guide provides detailed information on Hamilton Company’s full portfolio of syringes and needles. Everything from cleaning and preventative maintenance to individual part numbers are available for review. It also includes selection charts to help you choose between syringe terminations like cemented needles and luer tips.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Oasis or Sand Dune? Isolation of Psychedelic Compounds
January 20th 2025Magic mushrooms, once taboo, have recently experienced a renaissance. This new awakening is partially due to new findings that indicate the effects of psilocybin, and its dephosphorylated cousin psilocin may produce long lasting results for patients who might be struggling with anxiety, depression, alcohol and drug abuse, and post-traumatic stress disorder. Hamilton Company has developed a methodology for the isolation and identification of 5 common psychedelic compounds used in the potential treatment of disease. The PRP-1 HPLC column resin remains stable in the harsh alkaline conditions ideal for better separations.