Polyolefin (PO) is the largest volume industrial polymer in the world for making a wide range of commercial products that touch nearly every aspect of our daily lives, such as automobile parts, pipes, packaging films, household bottles, baby diapers, and so on.
Polyolefin (PO) is the largest volume industrial polymer in the world for making a wide range of commercial products that touch nearly every aspect of our daily lives, such as automobile parts, pipes, packaging films, household bottles, baby diapers, and so on. Polyolefin as a group includes high-density and low-density polyethylene (HDPE, LDPE), polypropylene (PP), EP rubber, and linear low density (LLDPE) copolymers of ethylene with alpha olefins (propylene, 1-butene, 1-hexene, 1-octene).
Though chemically simple, being made up of only carbon and hydrogen atoms, PO products derive their wide range of end-use properties from their semicrystalline structure. The ability to incorporate co-monomers in PE to create short chain branches (SCB) makes it possible to control the polymer crystallinity and crystalline morphology, and thus control the rigidity and flexibility of PO products. The controlling factor also depends on the SCB variations across the molar mass distribution (MMD).
Table I: Description of test samples used in this study
In dilute solution, the effect of SCB causes a reduction in polymer coil size with an increase of methyl groups off the backbone. Such structural changes can be studied by high temperature size exclusion chromatography (SEC, also known as gel permeation chromatography, or GPC), equipped with a multiwavelength infrared detector (IR5 MCT) (1), a light scattering (LS) detector, and a viscosity detector in the triple-detector GPC configuration (TD-GPC) (2). These techniques allow the analysis of SCB variation across MMD as it is illustrated by using the homopolymer PP and PE samples described in Table I together with their blend.
1. Ethylene-propylene Composition Analysis by GPC-IR5 MCT:
The MMD obtained by GPC-LS for the three samples in Table I are shown in Figure 1, where the IR5 MCT detector total CH channel is used to record the sample concentration across the GPC elution curve.
Figure 1: Molar mass distribution curves of the three samples.
For studying chemical composition variation across MMD, the combination of detector signals from the IR5 MCT methylene (CH2) and methyl (CH3) channels, shown in the bottom part of Figure 2, is used. Since the PP part in the blend sample is of higher MM than the PE part, a higher methyl signal is seen in the higher MM end of the sample, as expected.
The trend of changing ethylene percent, seen in Figure 2, can be predicted a priori from the co-adding of the original PE and PP GPC elution curves. This ethylene to propylene transition also can be determined from the difference between the CH2 and CH3 signals of the IR5 MCT detector (1). The validity of this analytical approach is supported by the agreement seen in the Figure between the predicted and the IR5 MCT results.
Figure 2: Analysis of the test blend sample. IR5 MCT CH2 and CH3 absorbance channels versus Log M (solid lines); ethylene % variation across MMD, predicted versus IR5 MCT measured (top part).
2. Ethylene-propylene Composition Analysis by TD-GPC:
In the triple detector GPC, the ratio of LS signal over sample concentration at each GPC elution volume gives a measure of the polymer weight-average MM (Mw). Similarly, the ratio of the viscosity detector signal over sample concentration gives the polymer intrinsic viscosity (IV). The plot of Log IV versus Log Mw by LS is called the Mark-Houwink (MH) plot, which is an often-used tool to reveal polymer branching and conformational structures (3).
Both light scattering and viscosity detector signals are more responsive to higher MM molecules. This is the reason that the LS and viscosity curves of the blend sample are skewed more to the higher MM region than the IR concentration curve shown in Figure 3.
Figure 3: Triple detector signal overlay of the test blend sample.
The corresponding MH plots for the three samples are shown in Figure 4. The MH line of the PP sample is seen to have a shift to higher MM from that of the PE sample. This is because of the extra methyl group off the chain backbone that leads to higher MM and lower IV for the PP sample.
Figure 4: Mark-Houwink plot of homopolymer PP, HDPE, and their blend.
The MH plot of the blend sample in Figure 4 is seen to go through a transition from the PE line to the PP line as the MM increases. The experimental data agrees with the predicted curve very well. This agreement supports the use of TD-GPC for studying polymer blends and block copolymers of EP, PE, EO (ethylene/octene), and so on.
The micro-structural difference of Polyethylene and Polypropylene as well as the compositional variation in EP can be detected by GPC-IR5 MCT and TD-GPC, as it has been validated by using a test sample of known composition created by a 50-50 PE-PP blend. This result of SCB variation across MMD provides the additional polymer structural information that complements perfectly the CRYSTAF (crystallization analysis fractionation), TREF (temperature rising elution fractionation), and CEF (crystallization elution fractionation) results which give SCB distribution across the crystallization temperatures (3–5) and to the cross-fractionation technique (6).
(1) J. Montesinos, R. Tarín, A. Ortín, and B. Monrabal. ICPC 2006 Houston.
(2) A. M. Striegel et al., Modern Size-Exclusion Liquid Chromatography (Wiley, New York, 2009).
(3) W. W. Yau and D. Gillespie, Polymer 42, 8947–8958 (2001).
(4) B. Monrabal, in Encyclopedia of Analytical Chemistry, R. A. Meyers, Ed. (John Wiley & Sons Ltd., 2000).
(5) B. Monrabal, L. Romero, N. Mayo, and J. Sancho-Tello, Macromolecular Symposia 282, 14–24 (2009).
(6) A. Ortin, B. Monrabal, and J. Sancho-Tello, Macromolecular Symposia 257, 13–28 (2007).
Polymer Char
Gustave Eiffel 8, Paterna, Valencia, 46980, Spain
tel. +34 96 131 8120, fax +34 96 131 8122
E-mail: info@polymerchar.com Website: http://www.polymerchar.com
2024 EAS Awardees Showcase Innovative Research in Analytical Science
November 20th 2024Scientists from the Massachusetts Institute of Technology, the University of Washington, and other leading institutions took the stage at the Eastern Analytical Symposium to accept awards and share insights into their research.
Inside the Laboratory: The Richardson Group at the University of South Carolina
November 20th 2024In this edition of “Inside the Laboratory,” Susan Richardson of the University of South Carolina discusses her laboratory’s work with using electron ionization and chemical ionization with gas chromatography–mass spectrometry (GC–MS) to detect DBPs in complex environmental matrices, and how her work advances environmental analysis.
AI and GenAI Applications to Help Optimize Purification and Yield of Antibodies From Plasma
October 31st 2024Deriving antibodies from plasma products involves several steps, typically starting from the collection of plasma and ending with the purification of the desired antibodies. These are: plasma collection; plasma pooling; fractionation; antibody purification; concentration and formulation; quality control; and packaging and storage. This process results in a purified antibody product that can be used for therapeutic purposes, diagnostic tests, or research. Each step is critical to ensure the safety, efficacy, and quality of the final product. Applications of AI/GenAI in many of these steps can significantly help in the optimization of purification and yield of the desired antibodies. Some specific use-cases are: selecting and optimizing plasma units for optimized plasma pooling; GenAI solution for enterprise search on internal knowledge portal; analysing and optimizing production batch profitability, inventory, yields; monitoring production batch key performance indicators for outlier identification; monitoring production equipment to predict maintenance events; and reducing quality control laboratory testing turnaround time.
Infographic: Be confidently audit ready, at any time and reduce failures in pharma QC testing
November 20th 2024Discover how you can simplify the audit preparation process with data integrity dashboards that provide transparency to key actions, and seamlessly track long-term trends and patterns, helping to prevent system suitability failures before they occur with waters_connect Data Intelligence software.
Critical Role of Oligonucleotides in Drug Development Highlighted at EAS Session
November 19th 2024A Monday session at the Eastern Analytical Symposium, sponsored by the Chinese American Chromatography Association, explored key challenges and solutions for achieving more sensitive oligonucleotide analysis.