Ultra fast separations of acetaminophen, caffeine, propofol and parabens are achieved in 30 s using the Thermo Scientific Accela UHPLC system with the 80 Hz PDA detector. The performance of the Accela PDA detector is highlighted by its low noise level, low drift, four orders of magnitude linear working range and high sensitivity.
Guifeng Jiang, Yenny Chan and Andre J. Nohl, Thermo Fisher Scientific Inc., San Jose, California, USA.
Ultra fast separations of acetaminophen, caffeine, propofol and parabens are achieved in 30 s using the Thermo Scientific Accela UHPLC system with the 80 Hz PDA detector. The performance of the Accela PDA detector is highlighted by its low noise level, low drift, four orders of magnitude linear working range and high sensitivity.
Detector noise and drift
The drift was measured at flow rate of 1000 μL/min with a constant flow of 100% methanol. The drift value for the PDA detector was 0.1 mAU/h.
The reduction in the level of noise increased detector sensitivity. The level of noise of the PDA detector was lower when the detector was operated at lower data sampling rates and/or higher rise times. The levels of noise of four different data rates (5, 20, 40, 80 Hz) were 1.1, 4.5, 14.8 and 43.1 μAU respectively.
Data sampling rate optimization
In general, about 20 data points per peak is required for reproducible quantification. High data sampling rates resulted in increased noise reducing sensitivity. Eight pharmaceutical small molecules were separated in 30 s using LC gradient method I and a BDS 50 × 4.6 mm column with elution order of uracil, acetaminophen, caffeine, methylparaben, ethylparaben, propylparaben, butylparaben and propofol (Figure 1). Similar peak width and peak resolution were obtained with 40 Hz and 80 Hz data sampling rates, indicating that both rates were suitable for this chromatographic separation. The 40 Hz data sampling rate was preferred here for its lower noise level and better sensitivity.
Figure 1
Detector linearity and sensitivity
High throughput UHPLC completes separations in tens of seconds and improves sensitivity resulting in sharper, taller and better resolved peaks.
Calibration curves of the four compounds were constructed over the concentration range of 39–300000 ng/mL with 11 calibration levels. Correlation coefficients were achieved 0.999 or better for all four compounds with four orders of magnitude linearity working range.
The Accela PDA detector acquires data at sampling rates up to 80 Hz for UHPLC high throughput separations with independent setting of diode rates, data sampling rates and rise time. Using the current UHPLC instrumentation and sub-2 μm particle columns, the PDA detector provided sufficient data point to define a chromatographic peak with peak width as narrow as 0.166 s. The four orders of magnitude linear working range and the ppb (ng/mL) level sensitivity for the PDA detector ensured the low level determination and quantification of the target compounds in UHPLC high throughput analyses.
Accela is a trademark of Thermo Fisher Scientific Inc.
Thermo Fisher Scientific Inc.
2215 Grand Avenue Parkway, Austin, Texas 78728, USA
tel. +1 800 532 4752 fax +1 561 688 8731
Website: www.thermo.com
New Study Uses MSPE with GC–MS to Analyze PFCAs in Water
January 20th 2025Scientists from the China University of Sciences combined magnetic solid-phase extraction (MSPE) with gas chromatography–mass spectrometry (GC–MS) to analyze perfluoro carboxylic acids (PFCAs) in different water environments.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ion used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
A Guide To Finding the Ideal Syringe and Needle
January 20th 2025Hamilton has produced a series of reference guides to assist science professionals in finding the best-suited products and configurations for their applications. The Syringe and Needle Reference Guide provides detailed information on Hamilton Company’s full portfolio of syringes and needles. Everything from cleaning and preventative maintenance to individual part numbers are available for review. It also includes selection charts to help you choose between syringe terminations like cemented needles and luer tips.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Oasis or Sand Dune? Isolation of Psychedelic Compounds
January 20th 2025Magic mushrooms, once taboo, have recently experienced a renaissance. This new awakening is partially due to new findings that indicate the effects of psilocybin, and its dephosphorylated cousin psilocin may produce long lasting results for patients who might be struggling with anxiety, depression, alcohol and drug abuse, and post-traumatic stress disorder. Hamilton Company has developed a methodology for the isolation and identification of 5 common psychedelic compounds used in the potential treatment of disease. The PRP-1 HPLC column resin remains stable in the harsh alkaline conditions ideal for better separations.