The U.S. EPA Office of Groundwater and Drinking Water released Method 524.3, "Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, Version 1.0," in the Federal Register on August 3rd, 2009 (1, 2).
Laura Chambers and Moklesur Rahman, OI Analytical
The U.S. EPA Office of Groundwater and Drinking Water released Method 524.3, "Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, Version 1.0," in the Federal Register on August 3rd, 2009 (1, 2). This new method contains an updated list of 76 VOCs that includes fuel oxygenates and two emerging contaminants of interest.
Method 524.3 allows the analyst to use purge-and-trap (P&T) operating conditions recommended by the instrument manufacturer, while restricting five parameters to prescribed ranges. This study was undertaken to optimize instrument parameters and validate that the resulting data meet or exceed new quality control and performance criteria in Method 524.3.
Instrumentation used in this study was an Eclipse 4660 Purge-and-Trap Sample Concentrator equipped with an Infra-Sparge™ sample heater, a 4551A Autosampler, and an Agilent 7890/5975 GC–MS.
Modified P&T parameters of 60 mL/min purge gas flow rate for 6 min with samples heated to 40 °C were used in this study.
The new method includes several modifications to the required quality control (QC), including changes to the way the initial calibration curve is evaluated, addition of acceptance criteria for the minimum reporting levels (MRLs), and elimination of the statistical method detection limit (MDL) requirement. These and all other QC requirements were included in the initial demonstration of capability (IDC).
Figure 1
Drinking water samples from three municipalities and one bottled-water facility were collected and analyzed. A chromatogram of a sample from College Station is shown in Figure 1. A summary of analytical results is presented in Table I.
Table I: Summary of analytical results from analysis of four finished drinking water samples, MS Scan mode. Only those compounds with concentrations above the minimum reporting level (MRL) of 0.5 ppb are listed.
U.S. EPA Method 524.3 permits greater latitude in setting instrument operating parameters. Parameters used in this study provided method validation and quality control data that consistently met or exceeded new performance criteria specified in U.S. EPA 524.3. The modified purge parameters also reduced the purge-and-trap cycle time for each sample by 5 min.
(1) Federal Register, Vol. 74, No. 147, August 3, 2009.
(2) Method 524.3. Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, Version 1.0; EPA-815/B-09-009.
(3) OI Application Note #3500, "U.S. EPA Method 524.3 for Analysis of Volatile Organic Compounds (VOCs) in Finished Drinking Water: Optimized Instrument Parameters and Method Validation," 2009.
OI Analytical
P.O. Box 9010, College Station, TX 77842
tel. (979)690-1711; fax (979)690-0440
Email: oimail@oico.com; Website: www.oico.com
Separation of Ultra-Short and Long Chain PFAS Compounds Using a Positive Charge Surface Column
December 11th 2024A separation of ultra-short and long chain PFAS (C1-C18) is performed on a HALO®PCS Phenyl-Hexyl column along with a HALO®PFAS Delay column which demonstrates excellent retention for both hydrophilic and hydrophobic analytes.