One of the most important attributes of flexible fused silica capillary tubing is its outstanding strength. When made properly with a protective abrasion resistant coating and handled appropriately this material has proven to maintain its strength remarkably well.
Joe Macomber and John Shannon, Polymicro Technologies, A subsidiary of Molex Incorporated
One of the most important attributes of flexible fused silica capillary tubing is its outstanding strength. When made properly with a protective abrasion resistant coating and handled appropriately this material has proven to maintain its strength remarkably well. Initial tensile strength and proper handling concepts have been discussed previously (1). Degradation of fused silica over time is certainly expected and significant work on the mechanical reliability of fused silica in the related fiber optic industry has been demonstrated in the form of fatigue testing (2). However, very little data exists on the impact of long term aging on capillary tubing tensile strength. Testing of 20 year old material has been conducted recently. This note discusses tensile strength results of aged capillary tubing as compared to recently produced material.
Aged fused silica capillary for this study was produced in 1988 and 1989 and was stored on white Styrofoam spools with a hub diameter of 8". Although the actual storage conditions were not controlled, all product used in this study were in their original sealed polymer bags. Only samples wherein the tubing had remained arc sealed from the external atmosphere were considered. Recently produced material was harvested directly from newly drawn tubing stock and was kept sealed until evaluated. Tensile strength testing was performed on a Tensile Tester, Instron Model 3344 (Instron, Norwood, Massachusetts).
Aged samples consisted of 3 sizes; these being 0.10 mm, 0.32 mm, and 0.53 mm i.d. capillary tubing (Polymicro Technologies, Phoenix, Arizona). Based on the spool hub diameter, the static stress levels during aging were 17 kpsi, 21 kpsi, and 32 kpsi, respectively. Material of corresponding sizes used for control samples was stored on the same 8" size spool, but none had aged more than 4 weeks.
Data from tensile strength testing is often displayed in the form of a Weibull Plot (3). Comparative display of all such plots is beyond the scope of this communication. A simple, and yet informative method of comparison is to summarize the mean tensile strength. Average values for all samples tested by product i.d. are displayed in Table I. For comparison the values of the control samples are included.
Table I: Mean Tensile Strength Comparisons (kpsi)
It is clear from the data that degradation in tensile strength has occurred after 20 years of aging. However, the material still retains good strength, with the mean values being well above the typical proof testing value of 100 kpsi. Data suggests this tubing would still pass a standard proof testing regime, but an elevated break rate may be expected. When the applied stress level of capillary of a typical 6" or 8" GC cage is calculated and compared against these finding, it is reasonable to conclude from a strength standpoint that this tubing could be used successfully to manufacture GC columns.
Polyimide coated fused silica capillary is a very strong material. With proper handling and storage, long lifetimes can be expected for most products. For specific questions, contact a Polymicro Technical Specialist.
(1) J. Macomber, LCGC Application Notebook, June 2009, p 58.
(2) B. Skutnick, T. Wei, Fiber Optics Reliability: Benign and Adverse Environments, SPIE Vol. 842 (1987), p 41.
(3) "Mechanical Stress & Fiber Strength," The Book on the Technologies of Polymicro, 2005, p 2.17.
Polymicro Technologies
A subsidiary of Molex Incorporated
18019 N 25th Ave., Phoenix, AZ 85023
tel. (602)375-4100; fax (602)375-4110
Website: www.polymicro.com
Automated Sample Preparation (ISO 20122) for MOSH/MOAH in Seasoning Oils
May 6th 2025This work presents an Automated Sample Preparation procedure for MOSH/MOAH analysis of Seasoning Oils. We compare results from a manual epoxidation procedure compliant with DIN 16995 with results based on fully automated sample preparation (epoxidation and saponification) compliant with ISO 20122. In both cases, online clean-up via activated aluminum oxide (AlOx) are used to remove interfering n-alkanes from the MOSH fraction during the HPLC run. Automated data evaluation using a dedicated software (GERSTEL ChroMOH) is presented.
Free Poster: NDSRI Risk Assessment and Trace-Level Analysis of N-Nitrosamines
April 25th 2025With increasing concern over genotoxic nitrosamine contaminants, regulatory bodies like the FDA and EMA have introduced strict guidelines following several high-profile drug recalls. This poster showcases a case study where LGC and Waters developed a UPLC/MS/MS method for quantifying trace levels of N-nitroso-sertraline in sertraline using Waters mass spectrometry and LGC reference standards.