The answer to the following question was supplied by LCG Europe's "GC Connections" columnist John Hinshaw.
The answer to the following question was supplied by LCGC Europe's “GC Connections” columnist John Hinshaw.
Q: What are the main causes of GC column degradation or decline in performance?
John Hinshaw: There are four main causes of GC column degradation and performance decline.
First, columns are degraded by the build-up of nonvolatile sample residues that interact with target analytes to cause peak tailing, catalytic decomposition, and ultimately loss of separation. Proper sample preparation, the use of a retention gap, and inlet liner maintenance will help alleviate this problem. Trimming the beginning of a contaminated column can recover some performance. Headspace sampling, when possible, completely eliminates this problem.
Second, chemically active substances in the sample such as derivatization reagents and their by-products can degrade the stationary phase and will lead to significant performance losses. Careful choice of derivatization reagents and avoidance of excessive amounts will help reduce this effect. Some stationary phases are less susceptible to these effects.
Third, the presence of oxygen in the carrier gas from leaks or poor quality gas cylinders will degrade some stationary phases, such as Carbowax and bonded Carbowax types. Attention to carrier gas quality, the use of fresh oxygen filters, and leak-free connections will ensure a longer lifetime for these columns.
Fourth, excessively high operating or conditioning temperatures will age a column prematurely. Never exceed the maximum operating temperature when performing separations, and never exceed the recommended conditioning temperature and duration.
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.