Cyanide, an environmental contaminant, can cause serious health effects including goiters, hypothyroidism, and some neuromuscular diseases.
Cyanide, an environmental contaminant, can cause serious health effects including goiters, hypothyroidism, and some neuromuscular diseases. Cyanide wastewater sources include the plating and mining industries, burning of coal and plastics, and effluent from publicly owned treatment works (POTW) (1). The U.S. EPA specifies 5.2 and 22 μg/L total cyanide for continuous and maximum limits, respectively, for POTW discharges into fresh water, and 1 μg/L total cyanide for both discharge limits into salt water (2). The EPA defines these continuous (4 d) and maximum (1 h average) limits to ensure that aquatic life is unharmed. POTW cyanide effluent is primarily flow-through from the plating and mining industries; however, some POTWs report higher cyanide concentrations than derived from influent as a result of waste disinfection processes (3). These processes create unstable intermediates that are stabilized by base and degrade to cyanide during acid digestions used for total cyanide determinations.
This application demonstrates an accurate and sensitive method for direct determinations of total cyanide in municipal drinking water and wastewater effluent samples using ion-exclusion chromatography-pulsed amperometric detection (ICE-PAD) with a disposable platinum working electrode (4).
A Dionex ICS-3000 system with an electrochemical detector (ED) controlled by Chromeleon 6.8 Chromatography Data System software was used for this study. Samples were prepared in 100 mM NaOH and digested with a strong acid using the Lachat MICRO DIST™ sample preparation system (5).
Total cyanide was determined in acid-digested samples, including two municipal drinking water samples (base-treated), a certified wastewater sample, and a filtered (0.2 μm) municipal wastewater effluent sample (untreated and base-treated). Cyanide concentrations ranged from <LOD to 5.99 ± 0.09 μg/L (n = 6). Figure 1 compares the separation of cyanide in untreated and base-treated municipal wastewater effluent, and the base-treated effluent with 5 μg/L cyanide added. These results agree with previous reports of measurable total cyanide concentrations found when municipal wastewater samples were base-treated (3). Determinations of total cyanide in drinking water and wastewater samples were accurate and precise with recoveries ranging from 97.4–102% and peak area RSDs <2.9%.
Figure 1
ICE-PAD provides low detection limits and improved cyanide recoveries over ion-exchange chromatography by excluding chloride and providing an improved resolution of cyanide from sulfide. These data show the characteristic advantages of ICE combined with the sensitivity and specificity of PAD to directly determine total cyanide in municipal drinking and wastewater samples.
(1) U.S. Department of Health and Human Services, Public Health Service, Agency for Toxic Substances and Disease Registry (ATSDR). Toxicological Profile for Cyanide, PB2007-100674. ATSDR: Atlanta, GA, 2006, pp. 153–159.
(2) Water Quality Standards. Code of Federal Regulations, Part 131, Title 40, 2008; Fed Regist. 21, 451–463 (2008).
(3) S. A. Carr, R. B. Baird, and B. T. Lin, Water Res.,31(7), 1543–1548 (1997).
(4) Dionex Corporation. Determination of Total Cyanide in Wastewater and Drinking Water Samples by Ion-Exclusion Chromatography and Pulsed Amperometric Detection (ICE-PAD), Application Note 227; LPN 2265. Sunnyvale, CA, 2009.
(5) Hach Company. User Manual for MICRO DIST™ Operation and Applications, Catalog Number 01304, Loveland, CO, 2008.
Chromeleon, and IonPac are registered trademarks of Dionex Corporation.
MICRO-DIST is a trademark of Lachat Instruments, a Hach Company brand.
Dionex Corporation
1228 Titan Way, P.O. Box 3603, Sunnyvale, CA 94088
tel. (408)737-0700; fax (408)730-9403
Website: www.dionex.com
Separation of Ultra-Short and Long Chain PFAS Compounds Using a Positive Charge Surface Column
December 11th 2024A separation of ultra-short and long chain PFAS (C1-C18) is performed on a HALO®PCS Phenyl-Hexyl column along with a HALO®PFAS Delay column which demonstrates excellent retention for both hydrophilic and hydrophobic analytes.