Thermo Scientific Application Note
Lipika Basumallick and Jeffrey S. Rohrer, Thermo Fisher Scientific, Sunnyvale, California, USA.
Coffee carbohydrates constitute the major part (at least 50% of the dry weight) of raw coffee beans, and are good tracers for assessing the authenticity of instant coffee.
Currently, the Association of Official Analytical Chemists (AOAC) Method 995.13—which is based on high-performance anion-exchange chromatography with pulsed amperometric detection (HPAE-PAD)—is used to determine carbohydrates in instant coffee.
This study first tested Method 995.13 on a Thermo Scientific Dionex ICS-3000 system. A few modifications to the official method are proposed to achieve separation of two pairs of sugars that are otherwise difficult to resolve. A fast method using the Thermo Scientific Dionex CarboPac SA10 column was also tested.
Both methods described provide good sensitivity and consistent response. The fast method is recommended when rapid separation is desired, keeping in mind that two pairs of sugars are not resolved. In applications where all 11 common coffee carbohydrates need to be resolved, Method 995.13 (with minor modifications) is recommended.
The experimental setup and the sample preparation procedures are described in Dionex Application Note 280 (now part of Thermo Fisher Scientific, Inc.).
Figure 1(a) shows the separation of the carbohydrates present in a mix of standards. Rhamnose-arabinose and sucrose-xylose are not completely resolved. The official method suggests excluding rhamnose from the mixed standard solution and performing 2–3 injections of the standard solution, or increasing the re-equilibrium time in order to achieve separation of the above-mentioned pairs. As an alternative, the column temperature may be lowered (15 °C) to achieve separation of all 11 carbohydrates [Figure 1(b)], albeit with increased run time.
Figure 1
The first set of coeluting peaks (rhamnose and arabinose) was resolved by eluting with 10 mM hydroxide for the first 6 min, then switching to deionized water [Figure 1(c)]. Only the mobile phase was modified; all other chromatography conditions were the same as in Method 995.13.
Using the fast method, the mixture of coffee carbohydrate standards separated on a Dionex CarboPac SA10 column [Figure 1(d)] in 8 min. Note that two pairs of sugars coelute.
For both methods, precisions ranged from 0.13–1.8% for retention time and 1.05–5.4% for peak area. Average recoveries were 70–127%.
Both methods are sensitive, accurate, reliable and differ primarily in their total analysis time and peak resolutions for coffee carbohydrate determinations.
Scan to receive complete application note.
Thermo Fisher Scientific Inc. (formerly Dionex Corp.)
1228 Titan Way, P.O. Box 3603, Sunnyvale, California, USA
tel: (408) 737-0700 fax: (408) 730-9403
Website: www.thermoscientific.com/dionex
RAFA 2024 Highlights: Cutting-Edge Chromatography Techniques for Food Safety and Food Analysis
November 18th 2024An illuminating session focusing on progress in analytical techniques used in food analysis took place on Wednesday 6 November 2024 at RAFA 2024 in Prague, The Czech Republic, including a talk on the analysis of 1000 toxins in 10 minutes.
Profiling Volatile Organic Compounds in Whisky with GC×GC–MS
November 1st 2024Researchers from Austria, Greece, and Italy conducted a study to analyze volatile organic compounds (VOCs) present in Irish and Scotch whiskys using solid-phase microextraction (SPME) Arrow with comprehensive two-dimensional gas chromatography coupled to mass spectrometry (GC×GC–MS) to examine the organoleptic characteristics that influence the taste of spirits.