Using Focused Gradients on a Combined Analytical/Preparative HPLC System to Optimize the Purification Scale-up Process

Article

The Application Notebook

The Application NotebookThe Application Notebook-06-01-2013
Volume 0
Issue 0

Agilent Technologies Inc.

This application note describes how a focused gradient is used to optimize the scale-up process from a 4.6 × 250 mm analytical column to a 50 × 243 mm preparative HPLC purification column.

Direct scale-up from analytical flow rates of 1.5 mL/min to preparative flow rates of 177 mL/min on a combined system poses several challenges: solvent delivery must be accurate at both low and high flow rates, and detection must span a wide dynamic range to cope with vastly different concentrations (1–5).

Experimental Conditions (6)

Agilent SD-1 Binary Purification System with Agilent Prep-C18 Scalar, 4.6 × 250 mm, 10 µm analytical column and Agilent Load & Lock preparative column, 50 × 243 mm, packed with 300 g of Agilent Prep-C18, 10 µm, at 1500 psi.

Formulas from the literature were used to scale-up the generic gradient profile from the analytical to preparative column (3,4). Taking dwell volume and column dead volume into consideration, the elution points of five standard compounds were determined using a linear elution gradient from 2 to 98% B within 16 min. The elution points of two adjacent peaks gave the start- and end-points of the focused gradients each with a slope of 2% B/min.

Figure 1: Generic and focused gradient applied to the standard calibration mixture and overlaid with a sample.

Results

5 µL of sample were analyzed on the analytical column using the generic gradient profile. The two target compounds eluted within window #3, hence focused gradient #3 was used for purification. A resolution of 1.11 for the two compounds was still obtained with an injection volume of 40 µL equivalent to 18.7 mg of sample. For purification on the preparative column, the scale-up formula was used to calculate a final injection volume of 4500 µL (2.11 mg). The recovery level of the collected fractions was 85 and 89%, respectively.

Conclusion

The Agilent SD-1 Purification System is a combined analytical/preparative HPLC system that provides for straightforward scale-up from a 4.6 × 250 mm i.d. analytical column to a 50 × 243 mm i.d. self-packed Agilent Load & Lock column. A focused gradient facilitated increase of column load and reduced runtime and solvent consumption.

References

(1) J.W. Dolan, LCGCNorth Am. 24(5), 458–466 (2006).

(2) J.W. Dolan and L.R. Snyder, J. of Chromatography A 99, 21–34 (1998).

(3) D. Guillarme, D. Nguyen, S. Rudaz, and J.L. Veuthey, Eur.J. Pharm. Biopharm. 66, 475–482 (2007).

(4) D. Guillarme, D. Nguyen, S. Rudaz, and J.L. Veuthey, Eur.J. Pharm. Biopharm. 68, 430–440 (2008).

(5) L.R. Snyder, J.J. Kirkland, and J.W. Dolan, Introduction to Modern Liquid Chromatography (3rd Edition, Wiley, Hoboken, New Jersey, 2010), pp. 403–498.

(6) P. Penduff, Scale-up with the Agilent SD-1 Purification System Analytical and Preparative Runs on a Single System, Agilent publication number 5991-2012EN, 2013.

Agilent Technologies Inc.

5301 Stevens Creek Blvd., Santa Clara, CA 95051

tel. (800) 227-9770 (Directory)

Website: www.agilent.com/chem/purifyyourway

Related Content