Metal-free columns should be considered when troubleshooting ion suppression. Chelating compounds that can form metal salt with stainless steel column housings are examined.
Most HPLC columns are made with stainless steel 316, which has surface oxides, and Fe(++), which have a high affinity for the exposed anionic oxygens present on phosphorylated compounds, and will readily interact causing adsorption and loss of sample. Additionally, these interactions can result in the formation of metal salts that can cause ion suppression when using LC–MS detection, potentially reducing analyte signal or, in extreme cases, cause no signal at all.
In this article, we show two cases of ion suppression and loss of sample due to adsorption, and recovery of signal strength by using our newly developed metal-free column housing. Our design utilizes the typical stainless-steel hardware so that we can preserve high-pressure capability, but the internal surfaces of the tubing and end fittings, as well as the inlet and outlet frits, are all coated with PEEK. This removes the largest contribution of metal surfaces within the HPLC sample flow path, significantly improving the analysis for certain compounds.
Figures 1 and 2 show organophosphorus pesticides or herbicides, and nucleoside triphosphates, respectively. All running conditions are shown within the figures. In each case, identical conditions were used on two columns packed with the same material; one with our metal-free housing, and the other with standard stainless steel.
The most dramatic example of ion suppression can be seen in Figure 1, particularly for glyphosate. While glufosinate does show signs of signal suppression and chelating, glyphosate is completely missing on the metal column, with LC–MS detection. Both peaks showed a much-improved LC–MS signal and analyte recovery when using a metal-free vs. a metal column. Peak shape was also improved.
Figure 1: Complete ion suppression of glyphosate signal
The triphosphates in Figure 2, while not completely missing as in the case of glyphosate, did show significant evidence of ion suppression and chelating for all nucleosides tested. In all cases, peak heights were dramatically reduced, and peak shape was negatively impacted. Under identical conditions, but in metal-free housing, all peaks showed an increase in symmetry and signal strength.
Figure 2: Triphosphates on metal-free vs. metal column
Often when troubleshooting issues with poor peak shape, low recovery and signal suppression, contributions of interaction with the column housing can easily be overlooked. Here we show that, particularly for certain compounds, there can be considerable interaction of the analytes with metal surfaces within the column. Therefore, the use of a metal-free column should be considered when troubleshooting issues related to poor peak shape, loss of sample, and ion suppression for potentially chelating compounds, especially when peak shape cannot be improved through modification of the mobile phase. It is also worth noting that we now offer metal-free columns in 15 different stationary phases, improving the range of available separation conditions. Please visit our website for more information.
Imtakt USA
2892 NW Upshur St., Portland, OR 97210
tel. (888) 456-HPLC, (215) 665-8902. Fax: (501) 646-3497
Website: ImtaktUSA.com
A Guide To Finding the Ideal Syringe and Needle
January 20th 2025Hamilton has produced a series of reference guides to assist science professionals in finding the best-suited products and configurations for their applications. The Syringe and Needle Reference Guide provides detailed information on Hamilton Company’s full portfolio of syringes and needles. Everything from cleaning and preventative maintenance to individual part numbers are available for review. It also includes selection charts to help you choose between syringe terminations like cemented needles and luer tips.
Oasis or Sand Dune? Isolation of Psychedelic Compounds
January 20th 2025Magic mushrooms, once taboo, have recently experienced a renaissance. This new awakening is partially due to new findings that indicate the effects of psilocybin, and its dephosphorylated cousin psilocin may produce long lasting results for patients who might be struggling with anxiety, depression, alcohol and drug abuse, and post-traumatic stress disorder. Hamilton Company has developed a methodology for the isolation and identification of 5 common psychedelic compounds used in the potential treatment of disease. The PRP-1 HPLC column resin remains stable in the harsh alkaline conditions ideal for better separations.
GPCRs – The Biological Traffic Modulator: Chromatographic Analysis of Peptides in the GI Tract
January 20th 2025The G protein-coupled receptors (GPCR) are found throughout the entire body and have shown significance in the development of new therapeutic treatments. Isolation of seven classic GRPC peptides initiating in the GI tract highlights the benefits of using the polymer-based PRP-3, a reversed-phase resin. The covalent bonds found in the PRP-3 exhibit advantageous interactions between the biological π bonds found in the peptides and the available aromatic benzyl rings of the resin.