Special Issues
Semiquantitative elemental analysis by ICP-MS is a powerful tool for quick screening of unknown samples for a wide range of elements.
Recent advances in collision–reaction cell technology have all but eliminated interferences from quantitative inductively coupled plasma mass spectrometry (ICP–MS) for most matrix types. However, in general, this technology requires specific information about the matrix and target analytes to set up the correct collision–reaction chemistry to eliminate the interferences. Additionally, the conditions required to eliminate one interference in one matrix frequently are not effective for all analytes in all matrices. For this reason, multiple sets of collision–reaction conditions typically are used. This results in significant deviations from predicted relative response factors for elements acquired under varying conditions. Making useful semiquantitative measurements requires that all elements be acquired under the same conditions. As a result, it has not been possible to use collision–reaction cell technology to reduce interferences in semiquantitative analysis in the same way as in full quantification. However, polyatomic interferences can be eliminated using carefully controlled kinetic energy discrimination (KED) in helium collision mode (Agilent 7500ce Octopole Reaction System), which can allow good semiquantification in most matrices for most analyte isotopes. Because helium is a nonreactive gas, no new interferences are formed in the cell, and plasma-formed polyatomic interferences are eliminated generically as a function of their larger ionic radii (Figure 1).
Figure 1
When only a single set of collision–reaction cell conditions is used, accurate relative response factors are maintained. Because helium collision mode eliminates only polyatomic interferences, interference-free isotopes of elements suffering from isobaric interferences still must be used. For the most part, the only significantly affected isotope is 40Ca. In this case, 44Ca can be used. KED eliminates the transmission of the larger polyatomic ions to the quadrupole by placing an energy barrier between the octopole and quadrupole (Figure 2).
Figure 2
Because the larger polyatomic ions undergo more energy-depleting collisions with He in the cell than do the smaller atomic ions, they lack sufficient energy to cross the barrier. Figure 3 depicts the effects of KED on ion energy. Only the high-energy atomic ions exceed the stopping potential and are transmitted.
Figure 3
Figures 4–7 illustrate the effective removal of multiple polyatomic interferences on several interference-prone elements. In this case, a standard containing 5 ppb of the elements of interest was measured in a matrix containing increasing concentrations of sulfuric acid, hydrochloric acid, and butanol.
Figure 4
Figure 5
Figure 6
Figure 7
This matrix was designed to create sulfur-, carbon-, chlorine-, and oxygen-based interferences. Measurements were taken in the "normal" mode (no gas in the collision cell) and in helium collision mode. Instrument conditions are listed in Table I.
Table I: Instrument conditions used for acquisitions depicted in Figures 4â7
The instrument was tuned for typical robust plasma conditions, which are essentially the same as the default parameters. Helium collision mode was tuned for reasonable transmission of low mass ions (helium flow = 3–4 mL/min and QP focus = –10 V) and efficient energy discrimination of polyatomic species (3 V gradient between octopole bias and quadrupole bias). Under these conditions, matrix effects such as ionization suppression and the formation of interferences in the plasma are minimized. Extract 1 is set at 0 V to maximize the system stability and ensure the narrowest possible ion energy distribution. Low ion energy distribution is essential to efficient energy discrimination (Figure 3). The following acquisition parameters were used (Figure 8):
Figure 8
Integration time: .........0.1 s
Mass range: ..............5–240
Replicates .................1
Peak pattern ..............Peak hopping, single slice per mass
The "empty masses" — for example, between 210 and 230 — and masses for elements including C, O, N, F, and the noble gases were skipped to reduce total acquisition time. Total acquisition time is less than 60 s for one replicate. Additional slices per mass can be acquired if desired, but at increased total acquisition time. A single slice is sufficient for semiquantitative screening mode.
A single calibration standard at 100 ppb, made up in 1% nitric acid– 0.5% hydrochloric acid, was used to update the semiquantitative response factor database for a range of elements across the mass range. Noncalibrated elements are updated by interpolating between calibrated isotopes, which the software does automatically. Any number of calibration elements can be used, but increasing the number of calibration elements will improve semiquantitative accuracy. Internal standardization was applied using a typical suite of internal standard elements distributed across the mass range.
Tables II and III show the results of a semiquantitative screen of two standard reference materials, NIST 1640 water and LGC 6177 landfill leachate. No attempt was made to matrix-match, and a single generic set of plasma conditions was used with the helium collision mode only. In all cases, for every certified element, the measured response was very close to the certified concentration from as low as 2 ppb for Rb in NIST 1640 to over 1700 ppm for Na in the LGC 6177 landfill leachate.
Table II: Results of semiquantitative analysis of NIST 1640 standard reference water using He semiquantitative mode (ppb)
Semiquantitative analysis has always been a powerful tool available to ICP-MS analysts for quickly estimating the concentration of unknown, uncalibrated elements in a variety of simple matrices. However, in complex matrices, polyatomic interferences could render the results for many elements useless. Collision–reaction cell technology, which requires more than one set of conditions for all masses, cannot be used because it would result in deviation from the standard relative response tables upon which semiquantitative analysis is based. Helium collision mode coupled with kinetic energy discrimination can overcome these limitations. By effectively removing polyatomic interferences, rapid semiquantitative screening of a wide range of sample types for most analyte elements is possible.
Table III: Results of LGC 6177, standard reference landfill leachate using He semiquantitative mode (ppb)
Glenn Woods is ICP-MS Applications Engineer, and Ed McCurdy is ICP-MS Product Specialist, with Agilent Technologies, Manchester, UK.
Steven Wilbur is Applications Chemist with Agilent Technologies, Lopez Island, Washington.
Determination of 3-MCPD and Glycidol in oil by ISO 18363-1, AOCS Cd 29c-13, DGF C-VI 18 (10)
January 28th 2025Fully automated method for 3-MCPD and Glycidol determination in edible oil by GC-MS, based on the widely used methods ISO 18363-1, AOCS Cd 29c-13, and DGF C-VI 18 (10). The automated GC-MS determination of 3-MCPD and glycidol in edible oils with evaporation step and GC column backflush ensures low LODs by eliminating excess derivatization reagent for improved method stability and system ruggedness.
Determination of 3-MCPD, 2-MCPD and Glycidol in oil and fat by ISO 18363-4 Zwagerman/Overman
January 28th 2025Fully automated method for 3-MCPD, 2-MCPD and Glycidol determination in Edible Oil and Fat based on ISO 18363-4 - Zwagerman/Overman with validation data. A recent upgrade to PTV injection has further improved the quality and robustness of results. Fatty acid esters of 3- and 2-monochloropropanediol (3-MCPD-e, 2-MCPD-e) and glycidol (Gly-e) are process contaminants that are formed, for example, when edible oils and fats are refined. After ester cleavage during digestion in the human body they pose a relevant health risk and therefore need to be determined in edible oils and fats and in fat containing food.
Automated Analysis of MOSH/MOAH in food and packaging extracts by LC-GC-FID
January 28th 2025The AppNote describes the fully automated determination of MOSH/MOAH in Food and Packaging extracts following DIN EN 16995. Industrial production, processing and transportation invariably put food at risk of contamination with MOSH/MOAH. To ensure a reasonable cost benefit balance, high laboratory productivity and good quality of results, leading contract laboratories increasingly strive to automate their processes. An example is the determination of MOSH/MOAH using a LC-GC-FID Coupling Platform. Depending on the sample matrix, additional automated sample preparation by aluminum oxide clean-up, epoxidation, and/or saponification is necessary prior to analysis. The dedicated evaluation software integrates the complex MOSH/MOAH chromatograms accurately and reproducibly.
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
A Guide To Finding the Ideal Syringe and Needle
January 20th 2025Hamilton has produced a series of reference guides to assist science professionals in finding the best-suited products and configurations for their applications. The Syringe and Needle Reference Guide provides detailed information on Hamilton Company’s full portfolio of syringes and needles. Everything from cleaning and preventative maintenance to individual part numbers are available for review. It also includes selection charts to help you choose between syringe terminations like cemented needles and luer tips.