Identifying Unexpected Environmental Contaminants with High-resolution, Accurate Mass LC–MS–MS

Article

The Application Notebook

The Application NotebookThe Application Notebook-07-02-2010
Volume 0
Issue 0
Pages: 6

This paper describes the use of the AB SCIEX TripleTOF 5600 LC–MS–MS system for the screening and identification of unexpected contaminants in lake water, demonstrating the value of high-resolution, accurate-mass data and retrospective data processing for the identification of nontargeted compounds.

Andre Schreiber, AB SCIEX, Foster City, California, USA.

This paper describes the use of the AB SCIEX TripleTOF 5600 LC–MS–MS system for the screening and identification of unexpected contaminants in lake water, demonstrating the value of high-resolution, accurate-mass data and retrospective data processing for the identification of nontargeted compounds.

Experimental

The AB SCIEX TripleTOF 5600 LC–MS–MS system (AB SCIEX, Concord, Ontario, Canada) equipped with TurboV source and an ESI probe was used. Full-scan MS spectra were acquired over the mass range of 100–1000 Da with an accumulation time of 100 ms. Information dependent acquisition (IDA) was used to trigger full scan MS–MS experiments using standardized settings of collision energy (CE) = 35 V and collision energy spread (CES) = ±15 V. The MS–MS accumulation time was set to 25 ms.

Figure 1: Total ion chromatogram (TIC), high resolution extracted ion chromatogram (XIC) with XIC width of ±10 mDa and accurate mass MS and MS–MS to calculate the empirical formula and characterize product ions.

Discussion and Results

High-resolution, accurate-mass, full-scan MS chromatograms allow retrospective data mining to identify nontarget and unexpected analytes. As previously presented statistical data analysis is an important tool to find the unexpected in complex samples.1 A large set of environmental water samples were compared by principal components analysis (PCA) using MarkerView software. The scores plot of PCA help to visualize differences between samples and the loadings plot can be used to find the nontarget contaminants responsible for these differences. PCA was used to identify a mass signal of 192 Da to be characteristic for many water samples originating in urban, but also wilderness areas. Using accurate mass and isotopic pattern, Formula Finder software empirically calculated the formula of the molecular ion of C12H18NO+. Formula Finder software also uses accurate mass MS–MS information to characterize the structure based on formula assignment for each major product ion. Finally, the MS–MS spectrum was searched against a comprehensive LC–MS–MS library with PeakView software (AB SCIEX) and the FIT value of 96% identified the detected compound clearly as the mosquito repellent DEET. With this information, DEET can be added to the list of pesticides commonly monitored.

Figure 2:

Reference

A. Schreiber and N. Pace, LC.GC Chromatography Online (March 2010).

If you would like more information on this application note or any of the information contained within please contact AB SCIEX

AB SCIEX

353 Hatch Drive, Foster City, California 94404, USA

tel. +1 650 638 5800 or +1 800 952 4716

fax +1 650 638 5884

Email: ServiceSales@absciex.com

Website: www.absciex.com