HPLC Column Maintenance: Tips for Extending HPLC Column Lifetime

Article

LCGC Europe

LCGC EuropeLCGC Europe-11-01-2018
Volume 31
Issue 11
Pages: 642

Tips on how to extend high performance liquid chromatography (HPLC) column lifetimes.

Tips on how to extend high performance liquid chromatography (HPLC) column lifetimes.

While the relative cost of high performance liquid chromatography (HPLC) columns has been reduced over time, extending column lifetime remains an important consideration for most laboratories. The following tips should help to protect your columns and significantly extend the useful lifetime of most phases.

Always read the manufacturer’s literature with respect to the recommended pressure, eluent pH, and temperature operating ranges for the column, and stick to these ranges

Note that higher operating temperatures often go hand in hand with reduced pH operating ranges. At low pH, the main symptom of column degradation is typically loss of efficiency (peak broadening) and at high pH, peak tailing and an increase in column back pressure.

Avoid mechanical shock of the column bed (dropping the column) and ramp the pressure and flow slowly (1 mL/min/min is ideal) each time eluent flow is initiated

Most modern HPLC equipment is capable of achieving this flow–pressure ramp automatically through secondary instrument settings. Bed voiding as a result of pressure shock often manifests itself via split or very badly tailing peaks. Columns may be reversed for analysis if a replacement is not readily available; however, the efficiency of the column is likely to reduce much more quickly as the bed will ultimately slump in the opposite direction, leading to the same chromatographic symptoms.

 

If columns have dried out, initiate the flow very slowly (0.1 mL/min/min) using an eluent containing at least 50% acetonitrile

If a “standard” (non “aq” or nonpolar embedded phase type) reversed phase column is suspected of phase collapse (shortening retention times, poor efficiency) from use with 100% aqueous mobile phases, the column should be reactivated at high flow with 100% acetonitrile at 60 °C (take care to not precipitate any solid buffers from the eluent remaining within the column). In both of these cases, between 50 and 100 column volumes may be required to properly re-equilibrate or re-activate the phase.

Columns should be properly washed after each use

A recommended washing routine may be:

  • Current content to 90% acetonitrile at 10% organic per 2 column volumes and hold for 10 column volumes (again take care to avoid precipitation of solid buffers by ramping the acetonitrile concentration slowly, as recommended here).

  • 10% acetonitrile per 2 column volumes to 50:50 acetonitrile–water and hold for 10 column volumes.

  • Remove from the system, end cap, and store. End capping of the column is very important to avoid the phase drying out.

One might use an older HPLC system as a column “wash station”, which can save significant amounts of operating time on “live” instruments.

If samples are likely to contain particulate matter, choose a qood quality inline filter with the appropriate mesh size; 0.45 µm for traditional columns and 0.2 µm for UHPLC columns is typical

If the sample matrix or diluent is likely to harm the sorbent (as a result of pH for example) or is particularly chemically dirty or intractable, a guard column may be used, and the phase should be matched with that of the analytical column. Take great care when selecting the dimensions of the guard column and connecting to the analytical column to ensure that the efficiency of the separation is not compromised.

 

If column or frit contamination is suspected as a result of peak splitting or loss of efficiency, it is possible to reverse the direction of the column for back flushing purposes, and the column washing procedure mentioned above is a good “recipe” for this purpose

One should uncouple the column from the detector to avoid fouling and note that reversal for flushing should only be used as a last resort and that the original efficiency of the column may not be achieved.

Remember that column volume may be estimated using π × r2 × L × 0.6 (the approximate interstitial porosity of silica used for HPLC column packing materials)

So, for a 150 × 4.6 mm column, this would approximate to:

3.142 × (2.3)2 × 150 × 0.6 = 1496 µL or 1.5 mL

Get the full tutorial at www.CHROMacademy.com/Essentials.

Recent Videos
Related Content