Methods: Vials and closures were exposed to acetonitrile for 2 h and subsequently analysed by LC-UV, LC–MS, and GC–MS to characterize the interferences.
L. Pereira, T. Edge, L. Shick, and M. Slade, National Scientific, Part of Thermo Fisher Scientific
Purpose: The evaluation of a new vial and closure for reducing interferences.
Methods: Vials and closures were exposed to acetonitrile for 2 h and subsequently analysed by LC-UV, LC–MS, and GC–MS to characterize the interferences. Comparisons were made between precleaned MS certified vials and closures, a control, and a competitor vial and closure set.
Improvements in chromatographic techniques, instrumentation, and sample handling continue to push the limits of detection in trace analysis. As such, the cleanliness of the total workflow process becomes even more important to reduce the potential for interferences and contamination that can ultimately reduce the sensitivity of the assay. The selection of the correct autosampler vial and closure is thus an important consideration.
Vials that are not effectively cleaned can introduce particulate matter that can cause blockages and accumulation of foreign material at the head of the separation column affecting chromatographic performance. Additionally, residual organic compounds that might survive the glass forming process or that leach from the closure when exposed to the sample solvent can reduce the analysis sensitivity.
The work presented in this poster evaluates the performance of the new MS certified vial and an ultra high pure bonded PTFE/ silicone closure compared with a control and a competitor vial and closure set.
The control (or blank) was generated by rinsing a vial (with no cap) twice with injection solvent (acetonitrile). 1 mL of acetonitrile was measured into the vial and vial with no closure was placed in the autosampler tray for analysis.
Figure 1 illustrates the UV traces for a control, a vial with PTFE/silicone closure and a vial + closure with pre-slit, incubated in the inverted position at room temperature. The three traces are identical, demonstrating that in LC-UV it is not feasible to see the interferences using this detector technology.
Figure 1: LC-UV chromatograms of a blank, a vial with PTFE/silicone closure, and a vial with pre-slit closure, incubated at room temperature in the inverted position.
Figure 2 shows the LC–MS analysis of a control, a competitor vial + closure, and an MS certified vial + closure. The LC–MS chromatograms for the control and the MS certified are very similar, suggesting that there is minimal interference from the vial + closure. However, the total ion chromatogram (TIC) obtained for the competitor vial shows a substantial amount of contamination. Combining the MS spectra of the latter part of the chromatogram reveals that the majority of the peaks in the positive ESI spectra are 74 units apart, suggesting the presence of polydimethylsiloxane (C2H6OSi)n.
Figure 2: LCâMS +ve ESI chromatograms obtained for different sample types. The MS spectra are obtained by summing the spectra over the last portion of the chromatogram.
Figures 3 show the results of the GC–MS analysis of the MS certified vials compared to the control. Both sets of chromatograms and spectra clearly demonstrate that there is minimal difference between the MS certified vials and the control.
Figure 3: GCâMS data for the MS certified vials. Comparison to control TIC and spectrum across the full time span, demonstrate that there is minimal difference between the two.
National Scientific, Part of Thermo Fisher Scientific
3970 Johns Creek Court, Suite 500, Suwanee, GA 30040
tel. (800) 332-3331; Email: chromatography@thermofisher.com
Website: www.nationalscientific.com
Efficient Calibration Curve Generation with Automated Dilution: Part 1
March 4th 2025This application note presents an automated method for generating calibration curves using an autosampler’s dilution function. It offers practical examples demonstrating a simple setup, enhanced productivity, accuracy and reliability, with a focus on eliminating manual dilution for greater efficiency in HPLC analyses.
Efficient Calibration Curve Generation with Automated Dilution: Part 2
March 4th 2025Part 2 delves into advanced applications using organic solvents with the HPLC autosampler dilution function. It details procedures for mixed standard solutions, highlighting automation’s role in reducing errors, saving labor and promoting efficient laboratory practices.