LCGC North America
Five simple pictures that reveal problems with your GC analysis – and how to fix them!
A picture paints a thousand words. The art of gas chromatography (GC) troubleshooting often lies in being able to recognize a problem from the evidence presented in the chromatogram or baseline appearance. In this short series of two articles, we present 10 simple GC pictures that show you how to recognize the issues, deal with the causes, and prevent them from happening again. Part I presented the first five pictures (1). Here, in part II, we present the remaining five pictures.
Irreproducible Retention Times
There are two types of retention time variability that can be observed (see Figure 1). Firstly, this variability could be from injection to injection in a run of samples or multiple injections from the same vial. This effect may be caused by changes in carrier-gas linear velocity because of a defective electronic flow controller, leaking septum or column connection, or variable gas supply because of a leak in the gas supply tubing. Under these circumstances you would expect to see variable retention times for all analytes. To remedy the problem, ensure that the instrument settings are correct and occasionally verify that the instrument calculated flow or linear velocity matches the actual carrier gas flow by measuring all inlet flows using a flow meter. Other causes of this type of variable retention times are variations in the oven temperature (either from a faulty temperature controller or from insufficient thermal equilibration time) or a situation in which the inlet pressure is not high enough to sustain flow at high gradient temperatures. In constant-flow operating mode, the carrier pressure is ramped to obtain a constant flow through the temperature program (as the temperature is increased carrier-gas viscosity also increases, and therefore, requires a higher pressure to maintain constant flow); in constant pressure mode, the inlet pressure remains constant, which means that as the carrier-gas viscosity increases during the temperature program the linear velocity will decrease causing retention times to change. With mass-flow or flow-sensitive detectors, this phenomenon can also manifest itself as either a gradually increasing or decreasing baseline signal depending on the response characteristics of the detector. Therefore, it is often better to use constant flow mode to avoid these problems.
Figure 1: Irreproducible retention times.
Long-term variability in retention times may be attributed to changes in the column, such as stationary-phase degradation or a shortened column because of trimming, which when entered incorrectly into the GC data system causes the instrument to incorrectly calculate the linear velocity of the carrier gas. There are certain cases where reduction in column length is more significan-either with short columns (10 m or less) or for columns that are exposed to excessive contamination and need larger lengths removed more regularly. If the column is too short, retention times will be too short and vice versa.
Integration Issues
With the phenomenon shown in Figure 2, it is difficult to reliably integrate the peaks because there is a rising baseline. The first solution to avoid this problem would be to eliminate any type of baseline drift in the first place. Baseline drift can be eliminated by reducing column bleed by properly conditioning the column or by using constant flow or linear velocity mode (especially when using a temperature program); if you need to use constant pressure mode and are using a mass-flow sensitive ionizing detector, you may have the ability to use variable make-up flow to compensate for the change in linear velocity during the method. In the real world, rising baselines are often inevitable; therefore, spending some time familiarizing yourself with the advanced integration settings in your data system will allow you to properly integrate challenging chromatograms-these settings include threshold, slope sensitivity, baseline reset points, and different integration methods, such as valley-to-valley. Remember to always use the same integration method every time to provide consistent results.
Figure 2: Integration issues.
Step-Shaped Peaks
In a chromatogram with step-shaped peaks (Figure 3), the step shape may appear either before or after the peak. It may also be accompanied by tailing or a shoulder, and results from analyte thermal degradation in the inlet. To remedy this problem, reduce the inlet temperature in 20 °C steps until a normal peak shape is achieved. However, take care not to lower the inlet temperature to a point where it is too low to achieve proper volatilization of all of your analytes; if this happens you may begin to see irreproducible peak areas.
Figure 3: Step-shaped peaks.
Poor Peak Shape for Early Eluted Peaks
Poor peak shape for early eluted analytes when you are carrying out a splitless injection is caused by a sample solvent–column polarity mismatch or the wrong initial oven temperature (the oven temperature is too high) (Figure 4). In splitless injection, the sample transfer from the liner to the column is slow; therefore, to mitigate band-broadening effects the sample must be refocused at the head of the column. Two focusing mechanisms occur:
Figure 4: Poor peak shape for early eluted peaks.
Loss of Resolution from Loss of Efficiency
To diagnose resolution loss from decreased efficiency (Figure 5), carry out a plate count (your software will probably do this) on the peaks. If you are seeing a loss in efficiency, the plate count will decrease, in particular for early eluted peaks. There are several causes of this problem, including poor column cut and installation, which leads to analytes being held up at the entrance to the column and eventually to a broadened peak; loss or contamination of the stationary phase (this can be resolved by trimming the column); or the incorrect column length or diameter entered into the data system, which results in an incorrect carrier gas velocity and in turn affects efficiency (remember each carrier gas has an optimum linear velocity to produce chromatography with the best efficiency).
Figure 5: Loss of resolution from loss of efficiency.
There may be other causes of these problems, so for further troubleshooting tips visit the CHROMacademy GC Troubleshooter www.chromacademy.com/gc_troubleshooting.html
Reference
Get the full tutorial at www.CHROMacademy.com/Essentials (free until December 20).
AI and GenAI Applications to Help Optimize Purification and Yield of Antibodies From Plasma
October 31st 2024Deriving antibodies from plasma products involves several steps, typically starting from the collection of plasma and ending with the purification of the desired antibodies. These are: plasma collection; plasma pooling; fractionation; antibody purification; concentration and formulation; quality control; and packaging and storage. This process results in a purified antibody product that can be used for therapeutic purposes, diagnostic tests, or research. Each step is critical to ensure the safety, efficacy, and quality of the final product. Applications of AI/GenAI in many of these steps can significantly help in the optimization of purification and yield of the desired antibodies. Some specific use-cases are: selecting and optimizing plasma units for optimized plasma pooling; GenAI solution for enterprise search on internal knowledge portal; analysing and optimizing production batch profitability, inventory, yields; monitoring production batch key performance indicators for outlier identification; monitoring production equipment to predict maintenance events; and reducing quality control laboratory testing turnaround time.
2024 EAS Awardees Showcase Innovative Research in Analytical Science
November 20th 2024Scientists from the Massachusetts Institute of Technology, the University of Washington, and other leading institutions took the stage at the Eastern Analytical Symposium to accept awards and share insights into their research.