The Application Notebook
Introduction
L. Lloyd, S. Ball and K. Mapp, Varian Inc., Church Stretton, Shropshire, UK.
Following the recent surge in popularity of flash chromatography as a rapid purification technique, two options now exist for small- to large-scale purifications of crude samples. HPLC is a high performance method using small particle media, which is suited to the purification of complex samples where high purity is required. Flash is a low-pressure technique using larger particle media and is suited to less complex samples or where lower purities are acceptable.
An analytical HPLC system was used for method development of a normal phase purification of δ-tocopherol. Flash chromatography was chosen as the purification technique, followed by subsequent screening of the collected fractions by HPLC to assess sample purity.
Tocopherols are a series of organic compounds consisting of various methylated phenols. They occur in alpha, beta, gamma and delta forms, determined by the number of methyl groups on the chromanol ring. Due to their limited UV absorption, tocopherols require an alternative to UV detection — ELSD.
The Varian 971-FP instrument uses an evaporative light scattering detector (ELSD) as an additional method of detection. Data are shown to demonstrate how a separation developed on a standard HPLC system can be scaled up for purification on the 971-FP flash instrument.
Figure 1: Structure of d-tocopherol.
Sample preparation
For method development, a solution containing 0.2 mg/mL of δ-tocopherol in heptane was used. Further 2 mg/mL and 10 mg/mL concentrations were also required for the purification stages.
Analytical HPLC
An analytical HPLC system consisting of a single isocratic pump, 7725i Rheodyne valve (with 200 µL sample loop) and Varian 385-LC ELS detector was used. The operating conditions were as follows:
Instrument: Isocratic HPLC with 385-LC ELSD
Column: SepTech ST60-10 Si, 10 µm 250 × 4.6 mm (pn: A8061250X046)
Sample conc.: 10 µL of 0.2 mg/mL solution of δ-tocopherol
Eluent: 90:10 heptane:ethyl acetate
Flow rate: 1 mL/min
Flash purification
Instrument: 971-FP with 385-LC ELSD
Column: SF15-12g Si 35 (pn: AX1371)
Sample conc.: 1000 µL of 10 mg/mL solution of δ-tocopherol
Eluent: 90:10 heptane/ethyl acetate
Flow rate: 6 mL/min
Fraction analysis
The fractions collected from the δ-tocopherol purification were analysed by HPLC using the SepTech ST60-10 Si column.
Figure 2 shows the chromatogram obtained from a 10 µL injection of a 0.2 mg/mL solution of δ-tocopherol onto the SepTech ST60-10 Silica HPLC column. This gave an analytical loading of 0.002 mg on-column.
Figure 2: Analysis of d-tocopherol on a SepTech ST60-10 Si column with ELS detection showing the analytical load, 0.002 mg.
A second injection, 100 µL of a 2 mg/mL solution, was made to demonstrate a preparative load (× 100) on the same column (Figure 3). This revealed some additional impurities in the sample that were not previously apparent.
Figure 3: Analysis of d-tocopherol on a SepTech ST60-10 Si column with ELS detection at increased sample load to show impurity profiles, 0.2 mg.
The flash purification was done on the 971-FP flash instrument, was done on by direct injection of the δ-tocopherol. Figure 4 shows the resulting chromatogram.
Figure 4: Flash purification of d-tocopherol on an SF15-12g Si 35 SuperFlash column with ELS detection. The coloured bands indicate the fractions collected.
During the flash purification run, four fractions were collected (as highlighted by the colored bands in Figure 4). 50 µL of each fraction were analysed using the HPLC analytical method. Figure 5 shows an overlay of their HPLC profiles.
Table 1
Table 1 summarizes the fraction qualities and shows δ-tocopherol was isolated with 100% purity.
Figure 5: HPLC analysis of the four fractions from the flash purification.
The results demonstrate that chromatographic separations can be transferred between HPLC and flash systems, saving both time and money.
A Varian 385-LC ELS detector can be used in combination with both techniques to give increased sensitivity for method development and fraction analysis.
These data represent typical results.
For further information, contact your local Varian sales office.
SepTech, SuperFlash, Varian and the Varian Logo are trademarks or registered trademarks of Varian, Inc. in the US and other countries. Rheodyne is a registered trademark of the IDEX Health & Science Group. ©2010 Varian Inc.
Figure 5: HPLC analysis of the four fractions from the flash purification.
Varian Inc.
Essex Road, Church Stretton, Shropshire SY6 6AX
tel. +44 1694 723581 fax +44 1694 722171
E-mail: flashpur@varianinc.com
Website: www.varianinc.com
A Guide to (U)HPLC Column Selection for Protein Analysis
April 16th 2025Analytical scientists are faced with the task of finding the right column from an almost unmanageable range of products. This paper focuses on columns that enable protein analysis under native conditions through size exclusion, hydrophobic interaction, and ion exchange chromatography. It will highlight the different column characteristics—pore size, particle size, base matrices, column dimensions, ligands—and which questions will help decide which columns to use.
The Benefits of Custom Bonded Silica
April 1st 2025Not all chromatography resins are created equal. Off-the-shelf chromatography resins might not always meet the rigorous purification requirements of biopharmaceutical manufacturing. Custom bonded silica from Grace can address a wide range of separation challenges, leading to real performance improvements. Discover more about the latest innovations in chromatography silica from Grace, including VYDAC® and DAVISIL®.
5 Things to Consider When Selecting a Chromatography Silica
April 1st 2025Particularly in the pharmaceutical industry, drug purity isn’t just a goal – it’s essential for achieving safety, stability and efficacy. However, purification is easier said than done, especially with challenging molecules like DNA and RNA “oligonucleotides,” due in large part to their diversity and the range of impurities that can be generated during production. Enter DAVISIL® chromatographic silica, with a wide range of pore diameters and particle sizes to meet your specific application, performance and sustainability requirements. Before you choose the chromatography resin for your next purification application, take a look at these 5 considerations.
Automating Protein Purification: Efficiency, Yield, and Reproducibility
March 27th 2025Recent advancements in automated protein purification stress the importance of efficiency, scalability, and yield consistency. This eBook compares different purification platforms, highlighting their impact on downstream applications and demonstrating how automation enhances throughput and process control.