These three tips provide a clear starting point developing any HPLC method, to achieve optimum resolution in the minimum amount of time.
The aim for any high performance liquid chromatography (HPLC) method is to achieve optimum resolution in the minimum amount of time. Such optimization will allow the best data to be gathered for the analysis, and keep running costs and analysis time low. To achieve this, chromatographers need to be able to rapidly design methods based on an understanding of the fundamentals of HPLC. These three top tips should provide a great starting point for any method.
1. Consider Analyte Chemistry
The interaction of an analyte with the HPLC stationary and mobile phase results in retention; without this interaction, there will be no separation. Therefore, going back to some chemistry basics in regard to your analytes will govern many of the choices you make in relation to the entire HPLC method, including column selection and mobile phase design. Of particular importance is the polarity of your analytes; purely nonpolar analytes are ideal for retention and separation under reversed-phase HPLC conditions on hydrophobic stationary phases (such as C18, C8, and phenyl). Polar or ionizable analytes will require a more polar stationary phase; for example, cyano or amino, or even a different mode of chromatography such as hydrophilic interaction liquid chromatography (HILIC), normal phase, or mixed mode to achieve separation. For ionizable analytes, which can be difficult to retain under reversed-phase HPLC conditions when in their ionized form, it is essential to correctly adjust the pH of the mobile phase to render these analytes neutral-in order to obtain retention and avoid peak shape problems (such as tailing with bases). Knowledge of analyte pKa can be incredibly useful in designing a mobile phase. The mobile phase should be two pH units below or above this value (depending if the analyte is an acid or a base, respectively) to render the analyte neutral. It can be useful to screen a range of pH values and assess the effect on retention and resolution.
2. Optimize Selectivity
Of all the chromatographic parameters that govern resolution, selectivity has the greatest impact in HPLC (Figure 1). Each parameter (selectivity, resolution, efficiency) can be altered via specific method parameters, such as mobile phase composition, temperature, or column dimensions. However, column stationary phase has the biggest impact on selectivity. Therefore, the second tip for HPLC method development is to screen columns with different phase chemistries. This approach has several advantages, including quickly finding the optimum stationary phase for your separation, as well as the fact that orthogonal stationary phase chemistries may reveal underlying impurity peaks that may not have been picked up if only similar column chemistries had been tried.
Figure 1: Impact of chromatographic parameters on resolution.
Other method parameters that can be used to affect selectivity further are organic solvent, mobile phase pH, solvent additives, and temperature.
3. Use a Scouting Gradient
A scouting gradient is a linear gradient from 5–10% B to 100% B over a set time (20 min is standard). The elution strength is held at 100% B for a few minutes, to make certain that all sample components have been eluted. For reversed-phase gradient elution, water (or buffer if there are ionizable analytes) and acetonitrile are typically the eluents of choice. The use of volatile buffers (such as ammonium formate) will result in a method that is compatible with mass spectrometry for use in LC–MS.
The chromatogram obtained can reveal a lot about the mobile phase composition, column chemistry, and mode of analysis required for the separation. If compounds are eluted over more than 25% of the gradient time after the end of the gradient, then a stronger B solvent or a less retentive column is required for the analysis. Furthermore, equation 1 (where tg is the scouting gradient time) can be used to determine if an isocratic separation is possible; this is always preferred, as it increases sample throughput (no re-equilibration time) and simplifies analysis.
âtg < 0.25tg [1]
The mobile phase composition at which the first peak is eluted should be used as the initial mobile phase composition when developing a gradient.
New Study Uses MSPE with GC–MS to Analyze PFCAs in Water
January 20th 2025Scientists from the China University of Sciences combined magnetic solid-phase extraction (MSPE) with gas chromatography–mass spectrometry (GC–MS) to analyze perfluoro carboxylic acids (PFCAs) in different water environments.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ion used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
A Guide To Finding the Ideal Syringe and Needle
January 20th 2025Hamilton has produced a series of reference guides to assist science professionals in finding the best-suited products and configurations for their applications. The Syringe and Needle Reference Guide provides detailed information on Hamilton Company’s full portfolio of syringes and needles. Everything from cleaning and preventative maintenance to individual part numbers are available for review. It also includes selection charts to help you choose between syringe terminations like cemented needles and luer tips.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Oasis or Sand Dune? Isolation of Psychedelic Compounds
January 20th 2025Magic mushrooms, once taboo, have recently experienced a renaissance. This new awakening is partially due to new findings that indicate the effects of psilocybin, and its dephosphorylated cousin psilocin may produce long lasting results for patients who might be struggling with anxiety, depression, alcohol and drug abuse, and post-traumatic stress disorder. Hamilton Company has developed a methodology for the isolation and identification of 5 common psychedelic compounds used in the potential treatment of disease. The PRP-1 HPLC column resin remains stable in the harsh alkaline conditions ideal for better separations.