Wyatt TechnologyReversed-phase chromatography represents one of the most popular applications of high performance liquid chromatography (HPLC), with particular importance for protein characterization. Because elution depends on the hydrophobicity of the sample, it is generally impossible to identify the separated products on the basis of their elution time (volume). Frequently, each eluted fraction must be isolated further and analyzed with other techniques to gain some understanding of the molecular behaviour of the protein.
Wyatt Technology
Reversed-phase chromatography represents one of the most popular applications of high performance liquid chromatography (HPLC), with particular importance for protein characterization. Because elution depends on the hydrophobicity of the sample, it is generally impossible to identify the separated products on the basis of their elution time (volume). Frequently, each eluted fraction must be isolated further and analyzed with other techniques to gain some understanding of the molecular behaviour of the protein.
Adding a miniDAWN® or DAWN® multi-angle light scattering (MALS) detector to one’s HPLC system, however, simplifies protein identification significantly. It allows one to measure absolute molar masses directly, irrespective of retention time, and to observe the properties of the protein in solution.
The experimental system used to collect the data shown here consisted of a miniDAWN connected downstream of an HPLC quaternary pump, degasser, autoinjector, and UV diode-array detector. A Vydac Protein C-4 column (150 mm × 4.6 mm, 300 Ã ) was used with a flow rate of 0.7 mL/min. ASTRA® software collected and analyzed the data to determine absolute molar mass and size.
Figure 1 shows a chromatogram of basic Fibroblast Growth Factor (bFGF) with a particular degree of oxidation produced by 0.2 equivalents of DTNB. The light scattering signal for each of the first two peaks is approximately twice the UV signal, while for the third peak the two signals are equal. This shows at a glance that the three peaks correspond, from left to right, to two types of dimers and a single monomer of bFGF; calculations show the molar masses to be 33 kDa, 34 kDa, and 17 kDa, respectively.Figure 2 presents a superposition of molar mass versus elution volume from four different separations, each based on a different degree of oxidation. The presence of dimers, trimers, tetramers, etc. is clearly evident, yet there is no consistency in the order of the elution volume.
MALS measurements made with the miniDAWN permit the absolute identification of each eluting peak. In addition, the root-mean-square radius of an eluting molecule may often be determined. For example, the rms radius of the bFGF pentamer was shown to be about 13 nm, indicative of a rod-like structure for this 85 kDa multimer.
Simply adding a miniDAWN to reversed-phase chromatography permits the absolute determination of the molar mass of each eluting fraction, the detection and identification of different multimeric forms, and the size and conformation of the separated molecules in solution.
Wyatt Technology
6330 Hollister Avenue, Santa Barbara, California 93117, USA
Tel. +1 (805) 681 9009
Website: www.wyatt.com • E-mail: info@wyatt.com
Altering Capillary Gas Chromatography Systems Using Silicon Pneumatic Microvalves
May 5th 2025Many multi-column gas chromatography systems use two-position multi-port switching valves, which can suffer from delays in valve switching. Shimadzu researchers aimed to create a new sampling and switching module for these systems.
Studying Cyclodextrins with UHPLC-MS/MS
May 5th 2025Saba Aslani from the University of Texas at Arlington spoke to LCGC International about a collaborative project with Northwestern University, the University of Hong Kong, and BioTools, Inc., investigating mirror-image cyclodextrins using ultra-high performance liquid chromatography–tandem mass spectrometry (UHPLC–MS/MS) and vibrational circular dichroism (VCD).