The Application Notebook
The efficiency of ion transfer between the quadrupole mass analyzers and the Q2 collision cell in a triple quadrupole mass spectrometry (MS) is critical for instrument robustness and sensitivity. Q2 is a radio frequency (RF) only non-mass filtering quadrupole containing only an inert collision gas such as Ar, He, or N2 to provide collision induced dissociation of the precursor ion selected in Q1.
The efficiency of ion transfer between the quadrupole mass analyzers and the Q2 collision cell in a triple quadrupole mass spectrometry (MS) is critical for instrument robustness and sensitivity. Q2 is a radio frequency (RF) only non-mass filtering quadrupole containing only an inert collision gas such as Ar, He, or N2 to provide collision induced dissociation of the precursor ion selected in Q1.
The traditional lens approach to MS involves a collision cell around Q2 with narrow apertures in and out to limit gas flow. The electrostatic lenses are used to refocus ions in and out. Lens-based design TQ MS is common but it requires complex tuning and is prone to contamination so it requires periodic cleaning and is subject to undesired noding effects. The lens-free TQ MS (Figure 1) has a lens-free Q0 ion guide and a collision area defined with seals around and outside of the Q2 ion guide path (1). The RF-only ion guides and no lenses are used to guide ions in and out of Q2. This approach provides high transmission, and has a simple and robust instrument design, also allowing more instrument up-time, without cleaning and re-tuning downtime. The lens-free approach is used in the Bruker SCION™ TQ GC–MS-MS platform.
Figure 1: Geometry details of the quadrupole interface models.
Two computational models are considered in this study. The first model is the lens-free quadrupole interface marked Q1Q2 and Q2Q3 in Figure 2a. Gas cell boundaries are marked with dashed lines. Ion transfer is achieved through close proximity RF-only ion guides: Q1 post-filters, Q2 RF-only input and output sections, and Q3 pre-filters (2).
Figure 2: Details of ion trajectories in and out of Q2 for the lens-free (a) and three-lens (b) quadrupole interfaces with and without gas in Q2.
A second model is the three-lens quadrupole interface marked Q1LLLQ2 and Q2LLLQ3 in Figure 2b. Gas is contained in Q2 with narrow apertures and three ion lenses are located at each end for refocusing the ion beam. The model also includes Q1 post-filters and Q3 pre-filters.
Ion trajectory simulations were undertaken using software SIMION 8 (3) and LUA programming.
As observed in Figure 2a, smooth ion transfer occurred between RF-only quadrupole sections (pre- and post-filters). There were no losses due to collisions when Q2 gas is on. Collisions occurred with the RF-only filters focus ion beam via "collisional cooling", reducing the ion loss in transmission. The three-lens based ion guide, on the other hand, showed significant ion transmission loss when the Q2 gas is on (Figure 3b), demonstrating that the lens cannot effectively focus the ions in the presence of the collision gas.
Figure 3: Transmission efficiency in and out of Q2 for the lens-free and the three-lens interfaces as a function of Q2 pressure (collision energy 25 eV and m/z 264), collision energy (pressure 0.266 Pa and m/z 264) and mass (collision energy 25 eV and pressure 0.266 Pa).
Ion transmission studies were also carried out by varying the collision gas pressure and collision energy and masses, and the results are summarized in Figure 3:
Another interesting observation of this study is that when Q3 resolution is reduced, which is a common setting in triple quad MRM scanning, the lens-free interface maintains nearly 100% efficiency, but the efficiency of the three-lens interface is still significantly limited due to collision losses in the lens regions (Figure 4).
Figure 4: Transmission efficiency Q2...Q3 when Q3 is set to lower resolution.
A lens-free, RF-only interface between a quadrupole and collision cell can significantly improve the ion transmission in the presence of the collision gas, compared to the traditional lens based ion guides. Also unlike the lens-based ion guides, the high ion transmission efficiency of lens-free ion guides is not affected by the collision gas pressure, energy, and the mass. Additional benefit of the lens-free pre- and post-filters is that the tuning is much simpler and required less frequently, resulting in a more robust system for long operation.
(1) US Pat 6576897.
(2) "Lenses or No Lenses?" poster presented at the 59th ASMS Conference on Mass Spectrometry and Allied Topics, Denver, Colorado, 2011, ThP04 83.
(3) SIS Inc., Ringoes, NJ 08551; www.simion.com.
For research use only. Not for use in diagnostic procedures.
Bruker Daltonics Inc.
Chemical and Applied Market,
3500 West Warren Ave, Fremont, CA, 94538
Website: www.bruker.com
The Benefits of Custom Bonded Silica
April 1st 2025Not all chromatography resins are created equal. Off-the-shelf chromatography resins might not always meet the rigorous purification requirements of biopharmaceutical manufacturing. Custom bonded silica from Grace can address a wide range of separation challenges, leading to real performance improvements. Discover more about the latest innovations in chromatography silica from Grace, including VYDAC® and DAVISIL®.
5 Things to Consider When Selecting a Chromatography Silica
April 1st 2025Particularly in the pharmaceutical industry, drug purity isn’t just a goal – it’s essential for achieving safety, stability and efficacy. However, purification is easier said than done, especially with challenging molecules like DNA and RNA “oligonucleotides,” due in large part to their diversity and the range of impurities that can be generated during production. Enter DAVISIL® chromatographic silica, with a wide range of pore diameters and particle sizes to meet your specific application, performance and sustainability requirements. Before you choose the chromatography resin for your next purification application, take a look at these 5 considerations.
Automating Protein Purification: Efficiency, Yield, and Reproducibility
March 27th 2025Recent advancements in automated protein purification stress the importance of efficiency, scalability, and yield consistency. This eBook compares different purification platforms, highlighting their impact on downstream applications and demonstrating how automation enhances throughput and process control.
MilliporeSigma: Ultrapure Water for Sensitive LC-MS Analysis of Pesticides
March 25th 2025The aim of the study was to illustrate the efficiency of Milli-Q® water purification systems in eliminating pesticides from tap water, thereby producing and delivering reliable and consistent-quality ultrapure water suitable for pesticides analysis