These five tips could improve your for gas chromatography methods.
These five tips for gas chromatography (GC) could improve your methods.
Hydrogen can provide increased analysis speed by increasing carrier gas linear velocity without losing too much efficiency, allowing for shorter run times and thereby increasing throughput. Lower temperature separations are another benefit; at faster elution times, it may not be necessary to increase the column temperature ramp rate. It may even be possible to lower the maximum temperature needed for the analysis, which will reduce the reequilibration time too. The use of lower temperatures leads to less column bleed, which in turn can mean a longer column life. Hydrogen is readily available through the electrolysis of water and, with gas generators, it can be generated safely on demand. Hydrogen gas is already used in the laboratory for a variety of purposes; it is the fuel gas for the most commonly used GC detector, the flame ionization detector.
Analytical sensitivity and detection limits are ultimately a function of the signal-to-noise ratio (S/N). A decrease in noise increases sensitivity. All GC columns bleed, although polar phases and thicker films are more prone to bleed. This normal degradation of the stationary phase polymer results in increased baseline noise. Degradation is accelerated when excess oxygen is present and at higher temperatures; hence the elevated baseline that is seen as the temperature rises to the column's upper limit. Columns designated for gas chromatography–mass spectrometry (GC–MS) have been designed to exhibit reduced bleed and high inertness at elevated temperatures, which ultimately increases S/N. Many detectors are sensitive to contamination from bleed products and require less maintenance when using low-bleed columns. For GC–MS applications, low-bleed stationary phases reduce column contribution to background noise, resulting in improved mass spectral purity and more-accurate library identification.
Manufacturers have developed new ways of deactivating GC columns in recent years, which provides improved sensitivity because they exhibit low bleed and low silanol activity. Decreased silanol activity is particularly pertinent when analyzing bases or polar compounds, or for certain specialist applications, such as pesticide, food, environmental, or drug analyses, all of which require ever decreasing detection limits. Active silanol species in the column can interact with bases or polar compounds and cause peak tailing that impacts sensitivity (via reduced peak heights relative to baseline noise) and makes integration (and hence quantitation) more challenging and less reproducible. The benefit of highly deactivated columns can only be fully realized in conjunction with a system that has an inert flow path (that is, inlet liner and packing, column, ion source, and so forth). Any advantages from low silanol activity in only the column will be mitigated if peak tailing occurs in the inlet. Most manufacturers provide highly inert consumables, including liners and packing, as well as deactivated inlets.
Splitless injection is used for low-concentration samples to provide optimum sensitivity. Sample vapor transfer from the inlet is much slower compared with split injection, resulting in band broadening; therefore, sample vapors must be trapped at the head of the column by using a low initial oven temperature. If some sensitivity can be sacrificed, it may be better to use a little bit of split, for the following reasons:
A split of 1:10 is good for balancing sensitivity with the benefits of split injection.
Gas-saver mode can be used during split injection to change the split ratio at a specified time after sample injection, which reduces carrier gas consumption. For example, if the split flow rate is reduced from 150 mL/min to 25 mL/min 1 min after sample injection (ensure all the sample and solvent vapors have been transferred to column prior to using the gas saver), the reduced split flow rate can be maintained throughout the analytical run until the next analysis. Under these conditions, that is a 79% gas saving per analysis: analysis time: 20 min; split ratio: 100:1; gas saver mode: split ratio 15 after 1 min; column temperature: 100 °C; column: 30 m × 0.25 mm, 0.25 µm.
University of Rouen-Normandy Scientists Explore Eco-Friendly Sampling Approach for GC-HRMS
April 17th 2025Root exudates—substances secreted by living plant roots—are challenging to sample, as they are typically extracted using artificial devices and can vary widely in both quantity and composition across plant species.
Miniaturized GC–MS Method for BVOC Analysis of Spanish Trees
April 16th 2025University of Valladolid scientists used a miniaturized method for analyzing biogenic volatile organic compounds (BVOCs) emitted by tree species, using headspace solid-phase microextraction coupled with gas chromatography and quadrupole time-of-flight mass spectrometry (HS-SPME-GC–QTOF-MS) has been developed.
Fundamentals of Benchtop GC–MS Data Analysis and Terminology
April 5th 2025In this installment, we will review the fundamental terminology and data analysis principles in benchtop GC–MS. We will compare the three modes of analysis—full scan, extracted ion chromatograms, and selected ion monitoring—and see how each is used for quantitative and quantitative analysis.
Quantifying Microplastics in Meconium Samples Using Pyrolysis–GC-MS
March 26th 2025Using pyrolysis-gas chromatography and mass spectrometry, scientists from Fudan University and the Putuo District Center for Disease Control and Prevention detected and quantified microplastics in newborn stool samples.
The Role of SPME Combined with GC–MS for PFAS Analysis
Published: March 25th 2025 | Updated: March 25th 2025Emanuela Gionfriddo and Madison Williams from University at Buffalo, the State University of New York, NY, USA discuss the important role that solid-phase microextraction (SPME) techniques with gas chromatography mass spectrometry (GC–MS) can play in the analysis of per- and polyfluoroalkyl substances (PFAS).