Starch is used for a variety of industrial and nutritional purposes. Its functional properties are influenced by the ratio and molar masses of its macromolecular constituents, which vary with source, crop year, and climate. Starch contains large homopolymers of amylose (AMY) and amylopectin (AMP).
Linear AMY consists of long chains of (1⇨4)-α-D-glucose linkages, while the higher molar mass AMP is a branched structure containing a mixture of (1⇨4)-α- and (1⇨6)-α-D-glucose linked residues. The goal of this work was to apply AF4–MALS–RI to separate AMY and AMP to calculate a mass ratio, to determine the molar mass distributions, the average molecular weights (Mw), and the mean-square radius (Rz) of the AMP component. We applied the technique to starches with AMY:AMP ratios covering a wide range.
An Eclipse AF4 system (Wyatt Technology) was equipped with a short (18 cm) channel, a 350 μm spacer, and a regenerated cellulose (10 kDa cutoff) membrane. Detection was accomplished with DAWN multi-angle light scattering (MALS) and Optilab refractive index (RI) detectors (both Wyatt Technology). The channel flow was maintained at 1.0 mL/min and the cross-flow was varied linearly from 1.0 to 0.1 mL/min for 10 min, then abruptly switched to 0.0 mL/min.
Figure 1: AF4âMALSâRI results for five native starches of varying AMY:AMP ratio: AF4âRI fractograms with molar mass distributions overlaid. (Cross-flow [Vx] = 1.0 to 0.1 mL/min in 10 min, then Vx = 0.0 mL/min).
Integration of RI peak areas enabled calculation of the AMY:AMP ratios, in excellent agreement with the nominal values. The values for Mw and Rz fall within the generally accepted limits found in the literature. Conformational plots for the AMP component verify its branched nature.
Figure 2: Conformation plot (log Rz versus log Mw) for the amylopectin component of five starches (Slopes 0.39â0.41 indicative of branching).
Starch characterization by size-exclusion methods can be limited because of shear degradation and/or column adsorption of the higher mass fractions of AMP. The open-channel separation of AF4 obviates these limitations and enables recovery of intact AMP. Resolution of AMY and AMP was achieved by varying the cross-flow. Coupled with MALS and RI detection, the mass ratios of AMY:AMP were accurately determined and the molar masses of these macromolecules were estimated without the need for calibration standards.
This note was graciously submitted by Rick White and Eija Chiaramonte, Global Analytical Sciences—Personal Health, The Procter & Gamble Company, Mason, Ohio, USA.
Wyatt Technology Corporation
6300 Hollister Avenue, Santa Barbara, California 93117, USA
Tel: +1 (805) 681 9009 fax: +1 (805) 681 0123
Website: www.wyatt.com
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.