Tips & Tricks GPC/SEC: How to Care for your GPC/SEC Instrument
Columns are at the heart of a gel permeation/size-exclusion chromatography (GPC/SEC) system. Software is used to evaluate the data, to understand and present the results, and to act as an interface between the analyst and the system. But what about GPC/SEC pumps, injection systems, and detectors? They should run uninterrupted 24/7, without the analyst having to think about them or to deal with them, but these components also need care. Following some simple rules can give better results in less time.
(Photo Credit: Adam Gault/Getty Images)
Modern gel permeation/size-exclusion chromatography GPC/SEC pumps, injection systems, and detectors are (in general) very robust and stable against lots of different solvents/mobile phases. Some precautions are required, however, in daily operation, even when using the best components. Nearly all of these precautions are associated with the mobile phase that is applied.
Mobile Phase Selection
All solvents should be of the highest quality (HPLC-grade). Even if these reagents are more costly, the difference in purity is marked. It is very important that the mobile phase is free of particles and dust because this might otherwise cause blockages in the system or the columns. It is good practice before using a new solvent to verify that all parts of the GPC/SEC system are compatible with the solvent and, when using aqueous systems, can be used at the pH value applied. This is especially important if components are used that have originally been designed for high performance liquid chromatography (HPLC). Unfortunately, GPC/SEC analysis often requires “exotic” organic GPC/SEC mobile phases that might cause problems with seals or other parts of the components. If the application allows, the preferred choice of solvent should be a non-corrosive solvent. For example, if an application can be run in tetrahydrofuran (THF) or in chloroform, from an instrument point of view the choice should be THF. Many GPC/SEC applications require the addition of salt, acidic, or basic additives. The concentration of these additives should always be kept as low as possible. In the case of dimethylacetamide (DMAc) or dimethylformamide (DMF), lithium (Li) salts are often used as additives. Because of the increased solubility of LiBr compared to LiCl, bromide is recommended in many application notes. However, it is more corrosive than LiCl and therefore potentially more dangerous for stainless steel instrumentation. In aqueous systems, the addition of an additive preventing algae growth is recommended. Adding 0.05% NaN
3
or acetonitrile will suppress the growth of microorganisms that could otherwise block tubing or columns. Seal wash options should be installed if mobile phases with salts are used. The best seal wash liquid is in most cases the pure solvent. Seal wash liquids should also be exchanged regularly and if the seal wash seals and holders are installed it is important that they are not running dry. If very high salt concentrations or extreme pH values are required it might be a good investment to use stainless-steel free instrumentation, which is now commercially available in many flavours and for nearly all detection options.
Preventive Maintenance
Preventive maintenance should be performed regularly to avoid unexpected instrument downtime, to ensure highest data quality, and to avoid costly repairs as a result of secondary damages as a consequence of not replacing worn parts. In every GPC/SEC system there are at least some in-line filters, pump seals, and injection rotor seals that require frequent exchange. In addition, check valves, pistons, tubing, and lamps should be checked regularly and replaced if required. The cycle for replacing of worn parts depends again on the application. As a rule of thumb we can say that one preventive routine maintenance a year is sufficient for a system with salt-free applications that is running regularly. If the application is more demanding and corrosive solutions are used, a reduced cycle of 6 months (or even less) is recommended.
Idle Mode, Weekend, and Vacation
Users sometimes have to decide if it is worth powering down the GPC/SEC system or not. Reducing the flow-rate is a good option that helps to save mobile phase, but still allows the instrument to be started up again very quickly. For a short shutdown period of a few days or less, it is also possible to run the system in recycle mode. In this case, the effluent from the detector is redirected into the solvent reservoir. However, running a GPC/SEC system should only be done if no injections are performed and the mobile phase is free from salt and additives (with the exception of algae prevention additives). Even in the case of running the instrument in recycle mode, the mobile phase should be exchanged regularly. If an instrument is not needed for several weeks or months it can be powered off completely. Special care is required for instrumentation and columns when mobile-phases with salts or other additives are used. As long as there are salts or additives in the system, a low flow-rate should always be applied (at least 0.05–0.1 mL/min) to prevent corrosion of the instrument or the columns. If the pump is going to be completely turned off, the salt solution should first be replaced by pure eluent. For this, at least 3–5 column volumes of pure mobile phase, if not more, should be used. If the instrument is not used for a long time the separation columns can be detached and stored tightly plugged with their original plugs in the refrigerator (without freezing them). In the case of corrosive mobile phases (chloroform), the instrument can be switched to a different mobile phase (isopropanol) prior to shutdown. Restarting the GPC/SEC system then requires a switch back to the original mobile phase. For new sample runs the GPC/SEC mobile phases and buffers should be prepared freshly on the day required. Starting an analysis with solutions that are several days (or even weeks) old and have been run in recycle mode will most probably end with low quality data with drifting and wavy baselines. If the columns were reinstalled it is good practice to first apply a low flow-rate (to ensure that there is no air from storage trapped) and to flush the columns with at least 3–5 column volumes before attaching them to the detectors. It is also worth checking the calibration with a checkout sample. If in doubt a new calibration should be performed.
Summary
Daniela Held
studied polymer chemistry in Mainz, Germany, and works in the PSS software and instrument department. She is also responsible for education and customer training. E-mail:
Dheld@pss-polymer.com
Website:
www.pss-polymer.com
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.