Several new materials and columns have been introduced in recent years for size-exclusion separations of proteins. How do I know which one to choose, and which separation conditions will be the best for my protein separation?
Several new materials and columns have been introduced in recent years for size-exclusion separations of proteins. How do I know which one to choose, and which separation conditions will be the best for my protein separation?
In Part 1 of this series (1), we focused on reversed-phase separations of proteins. In recent years, many new materials and columns have been introduced that provide potential for substantially better separations compared with those from one or two decades ago. Although some things have stayed the same, much of the old conventional wisdom has been overturned with the development of better stationary-phase chemistries and new research that has provided deeper insights into why we observe some phenomena (for example, low recovery of proteins from reversed-phase materials under some conditions). This research has also led to new guidance for operating conditions that improve the likelihood of obtaining acceptable chromatographic results.
Over the past few years, we have seen tremendous expansion in commercially available offerings for size-based separations of proteins as well. These separations are most commonly referred to as sizeâexclusion chromatography (SEC), and we will use that term here. As with reversedâphase separations of proteins, the upside to having more commercially available columns to choose from is that we can more precisely tailor our column choices to the needs of our applications. However, the downside to more options is that we have to choose which one is the most suitable, and in some cases, this can be a challenging task in itself. On the other hand, recent research studies have added considerable insights to the existing knowledge base to support this decision-making process. Even if we don’t fully understand why SEC materials behave the way they do in every situation (for example, see reference 2), we are in a much better position today to make good choices about columns and operating conditions than we were five years ago.
For this instalment of “LC Troubleshooting”, I have asked two of my collaborators in the biomolecule application space, and genuine experts in SEC separations of proteins, to join me in sharing some of the details that we have found to be particularly important to successful SEC separations.
Dwight Stoll
Basics of SEC Separations
From a theoretical point of view, SEC is arguably the simplest of all chromatographic separation modes. In reversed-phase mode and other separation modes, we spend a lot of time thinking and talking about retention (that is, retention factors greater than zero are very important!), which is a function of differences between the strength of intermolecular interactions between analytes, mobile phase, and stationary phase. It is differences between the way one analyte interacts with the mobile and stationary phases compared to another analyte that give rise to differences in retention (that is, selectivity) and ultimately resolution of two analytes. In this way, resolution in reversed-phase and similar separation modes (sorptive modes) is inherently chemically driven. SEC, on the other hand, is completely different, at least in the ideal case. Here, resolution has a physical basis, rather than a chemical one, and in the ideal case, there is no retention of the analyte by the stationary phase (that is, retention factors are zero or apparently negative). Instead, separation arises from differences in the physical limitations that analytes of certain sizes experience preventing them from exploring the entire pore network of porous particles used in SEC columns. Very small analytes in a sample will be able to explore most of the pore network. On the other hand, larger analytes that are too big to explore all of the pores will travel through the column with a higher velocity, and be observed flowing from the column earlier than the small analytes. From the point of view of the large analytes, the mobile phase volume inside the column is effectively smaller. Under ideal circumstances (that is, no retention as a result of intermolecular interactions), very small analytes will be eluted at what we would normally refer to as the dead time (tm) in reversed-phase separations. The mobile-phase volume associated with this time (that is, tm × F) is referred to as the inclusion volume (corresponding to the total porosity of the column). Larger analytes will elute at earlier times, before the inclusion volume.
Decision 1 - Choosing the Column
Before we dive into the details here, we want to be clear about our intent for this instalment. A tremendous amount of very good information on the following topics has been published in recent years. Our discussion here is limited to a survey of highlights of that work. Readers interested in the details behind our discussion are strongly encouraged to engage the literature cited here to learn more.
Particle Size and Column Length
Before the advances in column technology for SEC in recent years, most SEC columns in use were relatively large-typically 7.8 mm in diameter, and 150 to 300 mm in length. The long column lengths were required because of the large particles that were used, most of which did not have high mechanical strength and had to be used at relatively low pressures. The recent trend in column technology for SEC has been focused on the development of columns with smaller particles (<3 µm), in shorter columns (the standard now is 15 cm), and in smaller diameters (typically 4.6 mm). This trend has been supported by the development of particle chemistries that are both sufficiently mechanically stable to be used at the higher pressures that accompany the smaller particle sizes, and sufficiently inert toward biomolecules, to produce separations based mostly on molecular size. The move to smaller particle diameters also provides opportunities to improve separation speed by using higher flow rates through these columns. Whereas with larger particles, using high flow rates tends to result in decreases in efficiency (that is, plate number) and resolution, the price paid for doing so with smaller particles is not as severe.
Although we must be careful with generalizations, it is useful to think a bit about what the trend towards the use of small particles can do for us, in a practical sense. In rough terms the plate height scales with the particle diameter. So, upon moving from a 5-µm particle to a 2-µm particle, the plate height should decrease by about a factor of two (3). There are two main ways we can capitalize on this improvement in plate height-we can either improve resolution while using a column of the same length, or we can decrease analysis time while maintaining resolution. In the first case, if we use two columns of the same length-one with 5-µm particles and one with 2-µm particles-the plate number for the 2-µm particle should be approximately double that of the column with 5-µm particles. Since resolution scales with the square root of plate number, we should expect the resolution to improve by about 40%. In the second case, the plate number is directly proportional to column length, and inversely proportional to plate height. If the plate height decreases by a factor of two with the smaller particles, then we can decrease the column length by a factor of two, while maintaining the same plate number and resolution. If the same flow rate is used in both cases, we should expect this to immediately result in a 50% decrease in analysis time. This is a simple but useful view of these scenarios. There are a number of other factors to think about when considering the move to smaller particles, including the pressure limitations of the column and particles, and specifications of the instrument. More detailed discussions of the theory relevant to these considerations can be found elsewhere (4,5).
Average Pore Size and Distribution
As described above, the velocity of a particular molecule through a SEC column depends on the extent to which it can explore the pores of the particles. For particles with a well-defined pore size distribution, there is a range of molecular sizes for which a particular particle will be effective for size-based separations. The calibration curve shown in Figure 1 shows the selectivity (that is, difference in elution volume for a given change in molecular weight) for particles with different average pore diameters. We see that with small-pore columns there is good selectivity for small molecules, but the largest molecules will effectively be coeluted. On the other hand, the very large pore materials effectively separate the largest molecules, but the smallest molecules are coeluted. This type of plot can be used to decide which pore size will be most effective for the application at hand. For protein characterization, typical pore sizes between 150 and 500 Å are used. For common therapeutic proteins (MW ≈15–80 kDa), a pore size of 150–200 Å works well, while a 200–300 Å pore size is usually used for monoclonal antibodies (mAbs, MW ≈ 150 kDa). For very large proteins (MW > 200 kDa, for example, pegloticase or PEGylated proteins), typically the 500–1000 Å materials offer the most appropriate selectivity.
The pore size distribution has an impact on the slope of the calibration curves. The wider the pore size distribution, the steeper the curve is. Therefore, with a wide pore size distribution, the selectivity will be lower but the range of the analytes that can be separated will be broader. A narrow pore size distribution provides higher selectivity between species with slight differences in size, but only a limited size range of analytes can be separated.
The challenge in practice is that the only data that is readily available from column manufacturers is the nominal pore size. Unfortunately there is not broad agreement about how exactly to report pore size, and most of these measurements are based on gas adsorption/desorption measurements and may not be very meaningful for protein analyes. Thus, from the point of view of users of these columns, it is practically useful to experimentally determine the calibration curve by injecting a mixture of standard proteins in order to have a good sense for the selectivity that can be expected for a given protein sample.
Decision 2 - Choosing the Mobile Phase
After choosing the column, the next most important decision involves choosing exactly what will go into the mobile phase. As described above, one of the basic tenets of SEC separations is that conditions should be chosen so that retention (in a chemical sense) is minimized. If achieved, this approach ensures that the elution volume is an indicator of molecular size (as in a calibration curve of the type shown in Figure 1) and nothing else. At first glance, this seems like it should be straightforward-we should just choose a stationary phase that does not interact strongly through specific types of interactions with the analyte, and choose a mobile phase in which the analyte has a high solubility and that is able to minimize analyte–stationary phase interactions. But, if we’ve learned anything from 50 years of liquid chromatography, one of the big lessons has been that apparently tiny changes in the chemistry or structure of stationary phase or analyte can lead to big changes in retention. Indeed, we often exploit these interactions to great effect in reversed-phase separations when developing a new method. However, implementing this approach also means that achieving the “no retention” condition in SEC separations of proteins can be quite difficult in practice. There is a rich literature describing studies that have explored the use of different mobile phase modifiers and conditions to minimize stationary phase–analyte interactions.
It has been our experience that many of the specific effects of different mobile-phase conditions are protein or stationary-phase specific (or both), and thus some amount of exploration of variables is a necessary part of method development when starting work with a new molecule. However, based on our experience and the literature available to date, we can provide some suggestions for starting conditions:
And What About the Instrument?
There are at least two major issues we could discuss here-the impact of system dispersion on the performance of high-quality SEC separations, and the impact of instrument construction and the use of bioinert, biocompatible materials. The latter topic is complex and we will reserve that discussion for a later date. On the topic of system dispersion, we have to recognize that SEC separations are particularly prone to the negative effects of peak dispersion outside of the column (that is, extracolumn dispersion) because, again, the peaks are eluted with no retention or even before the inclusion volume. In separation modes where retention is desirable, the effects of extracolumn dispersion are less severe for peaks that are more retained, and in the case of gradient elution in many cases nearly all precolumn dispersion can be eliminated. Not so in SEC, because no peaks are retained, and all separations are isocratic.
As discussed above, until relatively recently most SEC columns in use were large in diameter (~7.8 mm) and long (300 mm). This resulted in separations where the peak volumes (that is, the peak width in time units, times the flow rate) were large enough in comparison to the injector-toâdetector volumes of LC systems they were connected to. However, with the improved plate heights and smaller volumes of state-of-the-art columns, the peak volumes are small enough that extracolumn dispersion has become a very important issue again (11). Figure 3 shows a comparison of the detected peak for a monoclonal antibody monomer obtained on three different LC systems with different levels of extracolumn peak dispersion. Given that resolution is often very valuable in SEC separations, this comparison makes it clear that one should seriously consider the effect of extracolumn dispersion on the observed chromatography, particularly when using modern SEC columns with small volumes and small particles.
When working with a state-of-theâart 150 mm × 4.6 mm SEC column, for a small analyte that is eluted near the inclusion volume, only 25–60% of the intrinsic column efficiency can be attained on conventional high performance liquid chromatography (HPLC) systems. The situation is even worse with a partially excluded analyte. Optimized ultrahigh-pressure liquid chromatography (UHPLC) systems having very low extracolumn volumes (typically Vec< 10 μL) have to be used to properly operate these columns. Therefore connector tubing volume and detector cell volume must be as low as possible. As most SEC separations are performed at ambient temperature, the mobile-phase preheater unit can also be removed to further gain in apparent efficiency. Another interesting finding is that conventional HPLC systems also have a big impact on the apparent elution time of proteins-and therefore on mass-calibration curve-when working with 150 mm × 4.6 mm columns. Under these conditions the resulting calibration data will not be reliable, except if corrected for extracolumn residence time.
Summary
Developing effective and high performing SEC separations for proteins requires attention to all facets of the method, including choices around stationary phase, particle size, and column dimensions, mobileâphase conditions, and instrument effects on chromatographic efficiency and resolution. Several research groups are continually contributing to our understanding of the effects of all of these decisions on separation performance. Although we certainly are very far from a complete understanding, we are in a better position than ever before to leverage the information we do have to develop the best methods possible today.
With this instalment of “LC Troubleshooting”, I am approaching my first full year of writing monthly columns that address some of the pain points we experience as practitioners of liquid chromatography. As I have said many times already here, some new problems emerge as technology changes and we adapt to the new behaviours of instruments and columns, but there are also many problems that nominally remain the same over time. I will continue working to bring a mix of discussions of old and new topics to the column, but I am also particularly interested to hear what you, as a regular consumer of the column, have to say about topics you would like to see addressed here. Are there topics that are emerging challenges that you have not seen addressed in the past? Are there “old” topics that you would like to see addressed in more depth? I’d love to hear your topic suggestions! Please send them along to LCGCedit@ubm.com
References
Szabolcs Fekete holds a Ph.D. degree in analytical chemistry from the Technical University of Budapest, Hungary. He worked at the Chemical Works of Gedeon Richter Plc at the analytical R&D department for 10 years. Since 2011, he has been working at the University of Geneva in Switzerland. He has contributed to more than 100 journal articles and authored book chapters and edited handbooks. His main interests include liquid chromatography (reversed-phase, IEX, SEC, HIC, SFC, HILIC), column technology, method development, pharmaceutical, and protein analysis.
Davy Guillarme holds a Ph.D. degree in analytical chemistry from the University of Lyon, France. He is now senior lecturer at the University of Geneva in Switzerland. He has authored more than 200 journal articles related to pharmaceutical analysis. His expertise includes HPLC, UHPLC, HILIC, LC−MS, SFC, and analysis of proteins and mAbs. He is an associate editor of Journal of Chromatography B and editorial advisory board member of several journals including LCGC North America, Journal of Chromatography A, Journal of Separation Science, and others.
Dwight R. Stoll is the editor of “LC Troubleshooting”. Stoll is a professor and co-chair of chemistry at Gustavus Adolphus College in St. Peter, Minnesota, USA. His primary research focus is on the development of 2D-LC for both targeted and untargeted analyses. He has authored or coauthored more than 50 peerâreviewed publications and three book chapters in separation science and more than 100 conference presentations. He is also a member of LCGC’s editorial advisory board. Direct correspondence to: LCGCedit@ubm.com
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.