A guide to solving recovery problems in solid-phase extraction (SPE)
A guide to solving recovery problems in solid-phase extraction (SPE)
Problems with quantitation in solid-phase extraction (SPE) may be associated with an inability to recover target analytes or internal standards from the extraction sorbent. These problems may be the result of poor choices in the chemistry of the protocol, or issues with the processing of the method.
The choice of sorbent chemistry is very important, and, to achieve the optimum extraction selectivity, one needs to choose a sorbent based on the analyte and matrix component chemistry. It is very useful to know the LogP/LogD values and pKa values wherever possible. In general terms, the specificity of the sorbent–analyte interaction increases in the following order:
The strength of electrostatic interactions between analytes and sorbent, some 15 times stronger than simple hydrophobic interactions, means that sorbents containing ionâexchange moieties are likely to give the most selective extraction under most circumstances.
Of course, although the ability of the sorbent to retain the analyte is of great importance, one also needs to consider the requirement to selectively elute the analyte, which again requires knowledge of the analyte chemistry and the interactions with the sorbent. For the avoidance of doubt: in solid-phase extraction, the ability to be selective invariably equates to the extent of analyte recovery.
To ensure optimum analyte retention, one should consider both the sample diluent and the sorbent equilibration solvent. If an analyte is loaded in a solvent that is too highly eluting, then analyte breakthrough may occur or final elution volumes may be larger than necessary. For hydrophobic sorbents (and analytes), use a sample diluent with as little organic solvent as possible to ensure sample solubility. If the sorbent requires a conditioning (wetting) step, which is typically 100% organic solvent, ensure that an equilibration step is used to lower the elutropic strength of the solvent within the sorbent bed prior to the sample loading step.
If ion-exchange sorbents are used, the pH of the sample diluent should be adjusted to ensure target analytes are in their ionized form before loading. Similarly, the pH of the equilibration solvent should be adjusted to ensure that any weak ion-exchange ligands are in the charged form (the pKa of weak ionâexchange sorbents can be obtained from the manufacturer‘s literature), to promote efficient retention of the target analytes. Further, one should consider loading the sample at a lower flow rate than is usual, because of the point-to-point nature of the electrostatic interactions in ion exchange, which require some degree of molecular orientation to occur, and therefore are inherently slower than other retention mechanisms. It may be necessary to include a “soak” time, in which the flow is stopped and the sample remains stationary within the sorbent bed to maximize analyte retention and prevent poor analyte recovery. Soak times of 30 s to several minutes can be necessary to ensure optimum recovery, and can be used with all types of sorbents.
Wash solvents, used to selectively elute interferences from the sorbent while retaining the analyte, should be strong enough to obtain the cleanest extract possible. However, care needs to be taken to avoid analyte breakthrough. Optimization of wash solvent strength should be performed by titrating the elution strength in discrete steps over several tubes or wells containing the same nominal analyte amount, plotting the solvent strength against analyte recovered, choosing the highest solvent strength that does not contain any analyte response. Remember here also that the organic solvent type and composition can be used in tandem with solution pH and ionic strength in order to achieve a “strong” wash solvent that removes a large amount of sample matrix components but does not cause analyte breakthrough. The “mixed-mode” phases are particularly useful in this respect because the dominant retention mechanism can be “flipped” between hydrophobic and hydrophilic/electrostatic using pH and solvent strength, resulting in improved sample cleanliness and recovery.
Similarly, the strength of the elution solvent should be optimized by evaluating successively stronger solvents, until one observes an unacceptable amount of interferences or background within the final eluate. Again, solvent strength, pH, and ionic strength should all be considered for optimization of elution strength, and soak steps may also be required to optimize recovery.
Get the full tutorial at www.CHROMacademy.com/Essentials (free until 20 July).
Liquid Chromatography to Analyze Vitamin D Proteins in Psoriasis Patients
January 21st 2025Can a protein involved in delivering Vitamin D to target tissues have an altered serum profile in psoriasis patients with cardiovascular disease? Researchers used liquid chromatography (LC) to help find out.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ions used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Oasis or Sand Dune? Isolation of Psychedelic Compounds
January 20th 2025Magic mushrooms, once taboo, have recently experienced a renaissance. This new awakening is partially due to new findings that indicate the effects of psilocybin, and its dephosphorylated cousin psilocin may produce long lasting results for patients who might be struggling with anxiety, depression, alcohol and drug abuse, and post-traumatic stress disorder. Hamilton Company has developed a methodology for the isolation and identification of 5 common psychedelic compounds used in the potential treatment of disease. The PRP-1 HPLC column resin remains stable in the harsh alkaline conditions ideal for better separations.