A guide to solving recovery problems in solid-phase extraction (SPE)
A guide to solving recovery problems in solid-phase extraction (SPE)
Problems with quantitation in solid-phase extraction (SPE) may be associated with an inability to recover target analytes or internal standards from the extraction sorbent. These problems may be the result of poor choices in the chemistry of the protocol, or issues with the processing of the method.
The choice of sorbent chemistry is very important, and, to achieve the optimum extraction selectivity, one needs to choose a sorbent based on the analyte and matrix component chemistry. It is very useful to know the LogP/LogD values and pKa values wherever possible. In general terms, the specificity of the sorbent–analyte interaction increases in the following order:
The strength of electrostatic interactions between analytes and sorbent, some 15 times stronger than simple hydrophobic interactions, means that sorbents containing ionâexchange moieties are likely to give the most selective extraction under most circumstances.
Of course, although the ability of the sorbent to retain the analyte is of great importance, one also needs to consider the requirement to selectively elute the analyte, which again requires knowledge of the analyte chemistry and the interactions with the sorbent. For the avoidance of doubt: in solid-phase extraction, the ability to be selective invariably equates to the extent of analyte recovery.
To ensure optimum analyte retention, one should consider both the sample diluent and the sorbent equilibration solvent. If an analyte is loaded in a solvent that is too highly eluting, then analyte breakthrough may occur or final elution volumes may be larger than necessary. For hydrophobic sorbents (and analytes), use a sample diluent with as little organic solvent as possible to ensure sample solubility. If the sorbent requires a conditioning (wetting) step, which is typically 100% organic solvent, ensure that an equilibration step is used to lower the elutropic strength of the solvent within the sorbent bed prior to the sample loading step.
If ion-exchange sorbents are used, the pH of the sample diluent should be adjusted to ensure target analytes are in their ionized form before loading. Similarly, the pH of the equilibration solvent should be adjusted to ensure that any weak ion-exchange ligands are in the charged form (the pKa of weak ionâexchange sorbents can be obtained from the manufacturer‘s literature), to promote efficient retention of the target analytes. Further, one should consider loading the sample at a lower flow rate than is usual, because of the point-to-point nature of the electrostatic interactions in ion exchange, which require some degree of molecular orientation to occur, and therefore are inherently slower than other retention mechanisms. It may be necessary to include a “soak” time, in which the flow is stopped and the sample remains stationary within the sorbent bed to maximize analyte retention and prevent poor analyte recovery. Soak times of 30 s to several minutes can be necessary to ensure optimum recovery, and can be used with all types of sorbents.
Wash solvents, used to selectively elute interferences from the sorbent while retaining the analyte, should be strong enough to obtain the cleanest extract possible. However, care needs to be taken to avoid analyte breakthrough. Optimization of wash solvent strength should be performed by titrating the elution strength in discrete steps over several tubes or wells containing the same nominal analyte amount, plotting the solvent strength against analyte recovered, choosing the highest solvent strength that does not contain any analyte response. Remember here also that the organic solvent type and composition can be used in tandem with solution pH and ionic strength in order to achieve a “strong” wash solvent that removes a large amount of sample matrix components but does not cause analyte breakthrough. The “mixed-mode” phases are particularly useful in this respect because the dominant retention mechanism can be “flipped” between hydrophobic and hydrophilic/electrostatic using pH and solvent strength, resulting in improved sample cleanliness and recovery.
Similarly, the strength of the elution solvent should be optimized by evaluating successively stronger solvents, until one observes an unacceptable amount of interferences or background within the final eluate. Again, solvent strength, pH, and ionic strength should all be considered for optimization of elution strength, and soak steps may also be required to optimize recovery.
Get the full tutorial at www.CHROMacademy.com/Essentials (free until 20 July).
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.