Over-expression of recombinant proteins is commonly used for the production of protein reagents in industry and academia. Problems often occur relating to the stress put on the cells to deal with this huge increase in synthesis. Cellular proteins that are part of the protein synthesis machinery are often up-regulated under such conditions. Large quantities of the recombinant protein can be bound to these cellular proteins, making purification difficult.
Wyatt Technology Corporation
Over-expression of recombinant proteins is commonly used for the production of protein reagents in industry and academia. Problems often occur relating to the stress put on the cells to deal with this huge increase in synthesis. Cellular proteins that are part of the protein synthesis machinery are often up-regulated under such conditions. Large quantities of the recombinant protein can be bound to these cellular proteins, making purification difficult.
Scale-up purification of a 61kDa recombinant protein was found to contain a significant contaminant (Figure 1), not found in the smaller scale trials. The 28kDa protein was identified by Edman degradation as rotamase, an enzyme that assists in protein folding. Co-purification of the rotamase suggests it is bound to the protein of interest, thus preventing further purification with techniques such as size exclusion chromatography (SEC).
Figure 1
Analytical size exclusion chromatography, followed by on-line direct mass detection by multi-angle light scattering (MALS), showed that a significant amount of the recombinant protein was not bound to the rotamase.
The sample was separated with a YMC-pack Diol-200 column, and detected by Wyatt Technology's 18-angle DAWN photometer and the Optilab-DSP Interferometric Refractometer.
Three peaks were isolated which contained 93kDa, 68kDa, and 28kDa species (Figure 2). The DAWN was able to identify the peaks in the following way: a non-covalent complex of one rotamase molecule and one recombinant protein molecule (peak 1), the free recombinant protein (peak 2), and the free rotamase (peak 3).
Figure 2
Estimating the molar masses of the proteins using columns calibrated to a set of protein standards (BioRad) resulted in significantly different values for the three peaks: 146kDa, 98kDa, and 46kDa, respectively. Thus, the information provided by MALS system—an absolute system because it is not calibrated with standards—contributed to the correct interpretation of the chromatogram, provided invaluable insight for development of the scale-up purification protocol, and into the protein-protein interactions.
We are most grateful to Nestor B. Nestor and George A. Karam, Pfizer Central Research, Groton, CT, for their efforts in preparing this application note.
Wyatt Technology Corporation
6300 Hollister Avenue, Santa Barbara, CA 93117
tel. (805)681-9009, fax (805)681-0123
Website: www.wyatt.com
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.
Axcend Focus LC® Cartridge Options
December 19th 2024Welcome to our comprehensive selection of High-Performance Liquid Chromatography (HPLC) cartridges. Designed for precision and reliability, our cartridges cater to diverse analytical needs, ensuring optimal performance in your laboratory processes.