Simulated Moving Bed technology (SMB) is a continuous chromatography technique widely used in the pharmaceutical/biotechnology industry. The continuous chromatographic separation of sodium sulphate (Na2 SO4) from a mixture of glucose and mannose was established by using the Orochem Zuccheroâ„¢ columns and the SMB technology. The main advantages of SMB over batch chromatography includes better yields, higher purity, and decreased solvent usage, which together make the SMB technology economically viable and ideal for desalting applications.
Anil R. Oroskar and Rakesh Nair, Orochem Technologies, Inc.
Simulated Moving Bed technology (SMB) is a continuous chromatography technique widely used in the pharmaceutical/biotechnology industry. The continuous chromatographic separation of sodium sulphate (Na2 SO4) from a mixture of glucose and mannose was established by using the Orochem Zucchero™ columns and the SMB technology. The main advantages of SMB over batch chromatography includes better yields, higher purity, and decreased solvent usage, which together make the SMB technology economically viable and ideal for desalting applications.
Figure 1
SMB system
Orochem SMB system equipped with eight Zucchero -Na+ columns.
Columns: Zucchero Na+ , 100 x 20 mm
Eluent: DI Water
Temperature: 65°C
Feed rate: 2 mL/min (10% Na2 SO4 / 5% Glucose + Mannose)
Figure 2
Analytical
Equipment: HP 1090 HPLC system
Column: Zucchero, 300 x 7.2 mm
Eluent: 0.009N H2SO4 in water
Detection: RI
Inj volume: 15 µL
The chromatograms show that a good separation of salt and the sugar was achieved by using the SMB system. The measured recovery of this process was 95% and the measured purity was 99.9%. This system will be a useful tool for companies in pharmaceutical and biotechnology industries.