The increasing use of pesticide testing coupled with reductions in maximum permissible residue levels of pesticides in food have driven demand for fast, sensitive, and cost-effective analytical methods for high-throughput screening of multiclass pesticides in food. Detection of 510 pesticides at low parts-per-billion levels can be achieved within minutes using orbital trap technology. The high resolving power of these systems enables accurate mass confirmation of all compounds, including isobaric pesticides. This article will provide an overview of current legislation and illustrate how mass spectrometry instrumentation can enable fast and accurate pesticide screening.
Growing food safety concerns and the globalization of agricultural trade have triggered the enforcement of stringent pesticide regulations worldwide. These different national regulations result in considerable diversity in the level of food protection on a global basis. For example, in 2007, the U.S. Environmental Protection Agency (EPA) completed a 10-year reassessment of 9721 pesticide tolerances to meet more stringent safety standards and recommended the revocation or modification of thousands of uses of pesticides in food (1). China published national standard GB 2763-2005 in 2005, which established 478 maximum residue levels (MRLs) for 136 pesticides (2). Japan's Positive List System, introduced in 2006, established MRLs for hundreds of agricultural chemicals, including approximately 400 pesticides in food, and set a uniform limit of 10 ppb to chemicals for which MRLs have not been determined (3). These regulations have fueled the need for faster, more sensitive and cost-effective analytical methods for high-throughput screening of multiclass pesticide residues in food.
Pesticides in food traditionally have been monitored and quantified using gas chromatography (GC) coupled with either selective detectors (for example, electron capture) or mass spectrometry (MS). Gas chromatography (GC)–MS continues to be widely used in pesticide analysis because it is highly selective, provides confirmation of multiple classes of pesticides in a single analytical run, and is relatively inexpensive and easy to operate. However, GC–MS cannot detect polar, thermally unstable, or low volatility compounds without derivatization.
Recent improvements in liquid chromatography (LC) throughput and MS detection capabilities have led to a surge in the use of LC–MS-based techniques for screening, confirmation, and quantitation of ultratrace levels of multiclass pesticide residues, including those that are not GC-amenable. LC–triple-quadrupole tandem MS (LC–MS-MS) enables highly selective and sensitive quantitation and confirmation of hundreds of target pesticides in a single run, but this approach requires extensive compound-dependent parameter optimization and cannot be used to screen for untargeted pesticides.
Full-scan approaches using high performance time-of-flight (TOF) or single-stage orbital trap mass spectrometers coupled to ultrahigh pressure LC (UHPLC) facilitate rapid and sensitive screening and detection of LC-amenable pesticide residues present in a sample. The superior resolving power of the single-stage orbital trap mass spectrometer up to 100,000 FWHM compared to that of TOF instruments (10,000–20,000 FWHM) ensures the high mass accuracy required for complex sample analysis (4). Full-scan UHPLC–single-stage orbital trap MS can be used to screen a virtually limitless number of pesticides and, unlike MS-MS methods, does not require compound-dependent parameter optimization. High-resolution LC–MS instrumentation, however, can be cost-prohibitive for many routine monitoring laboratories.
Newly designed LC–MS single-stage orbital trap mass spectrometers facilitate reliable mass measurements of complex samples and enable confident discrimination of coeluted isobaric compounds in complex samples at a lower cost (5,6). In addition, a wide in-scan dynamic range of 3–4 orders of magnitude facilitates the detection of trace levels of compounds in the presence of highly abundant matrix interferences. High scan speeds and polarity switching ensure full compatibility with both UHPLC and high-throughput methods. Cost-effective and easy to operate, these systems are an ideal tool for compliance monitoring in regulatory laboratories.
This study demonstrates rapid screening and accurate mass confirmation of 510 pesticides at low ppb levels using a UHPLC system (Accela, Thermo Fisher Scientific, Madison, Wisconsin) coupled to a high-resolution benchtop single-stage Orbitrap mass spectrometer (Exactive, Thermo Fisher Scientific) and an autosampler (PAL, CTC Analytics, Carrboro, North Carolina). Detection of the 510 pesticides at low parts-per-billion levels was achieved within 12 min. The high resolving power of the single-stage Orbitrap platform enabled precise mass confirmation of all compounds, including isobaric pesticides. The system is ideally suited for routine, comprehensive screening of targeted and nontargeted pesticides at or below the 0.01 mg/kg (10 ppb) default limit set by EU and Japanese legislation.
Table I: LC parameters
Pesticide standards were obtained from the U.S. Food and Drug Administration (FDA). A stock solution of a mixture of 510 pesticides was prepared at a concentration of 3 mg/L. Calibration solutions with concentrations of 1–250 ppb were prepared by serial dilution of the stock solution in 50:50 (v/v) acetonitrile–water. Spiked spinach samples were prepared for analysis using a modified QuEChERS (quick, easy, cheap, effective, rugged and safe) sample preparation procedure (6). Malathion-d6 was used as an internal standard for calibration.
Table II: MS parameters
The LC and MS experimental parameters are detailed in Table I and Table II. Data acquisition was performed using proprietary software (Xcalibur, Thermo Fisher Scientific) as was data processing (Pathfinder, Thermo Fisher Scientific).
Experimental results demonstrate that UHPLC improves chromatographic resolution, speed, and sensitivity, and when coupled to MS, it facilitates rapid, high-throughput analysis of challenging samples. Using UHPLC single-stage orbital trap MS, a mixture of 510 pesticides representing a broad spectrum of chemical classes was separated and detected within 12 min. High resolution (50,000) and high mass accuracy (< 5 ppm without internal calibration for most compounds) enabled identification of all analytes. Separation of isobaric pesticides was achieved only at the high resolving powers provided by single-stage orbital trap MS, as demonstrated in Figure 1.
Figure 1
Excellent linearity in detector response was observed over the range of 1–250 ppb, with correlation coefficients greater than 0.99 for the majority of pesticides. For the concentration range studied (1–250 ppb), limits of quantitation (LOQs) were estimated from triplicate injections (CV < 15%) of standard solutions at concentration levels corresponding to a signal-to-noise ratio of 10. LOQs ranged from 1–50 ppb and for 499 pesticides LOQs were at or below 10 ppb, which is the MRL imposed by EU and Japanese regulations.
To evaluate the applicability of the technique for complex food samples, UHPLC–single-stage orbital trap MS was used to screen for pesticides extracted from a spiked spinach matrix. An extraction procedure based on fast and efficient QuEChERS methodology was used to facilitate rapid high-throughput multiresidue analysis. Extracted ion chromatograms and calibration curves for six pesticides extracted from the spiked spinach matrix are depicted in Figure 2.
Figure 2
The detection and quantitation capabilities of the method were assessed using the EPA method detection limit (MDL) procedure (7). For all pesticides, limits of detection (LODs) and LOQs were lower than 1 ppb.
Strict worldwide pesticide regulations have created an increasing need for rapid, sensitive, reliable and cost-effective analytical techniques capable of facilitating high-throughput screening of multiclass pesticides in food. In response, UHPLC–single-stage orbital trap MS has emerged as a viable method for fast multiresidue screening of pesticides at low parts-per-billion levels. The high mass resolution and accuracy of the technique enable dependable identification of all compounds at or below the default limits set by legislation. The technique is ideally suited for the routine monitoring of targeted and nontargeted pesticides by regulatory laboratories.
Allen Zhang, James S. Chang, Christine Gu, and Mark Sanders are with Thermo Scientific, Madison, Wisconsin.
(1) Environmental Protection Agency, Accomplishments under the Food Quality Protection Act (FQPA), http://www.epa.gov/pesticides/regulating/laws/fqpa/fqpa_accomplishments.htm
(2) National Standard GB 2763-2005, "Maximum residue limits for pesticides in food", Standardization Administration of the People's Republic of China, 2005.
(3) Japanese Ministry of Health, Labor and Welfare, Department of Food Safety, Introduction of the Positive List System for Agricultural Chemical Residues in Foods, http://www.mhlw.go.jp/english/topics/foodsafety/positivelist060228/introduction.html
(4) "High Resolution and Precise Mass Accuracy: A Perfect Combination for Food and Feed Analysis in Complex Matrices," Thermo Scientific Application Note 30163, 2008.
(5) "The Thermo Scientific Exactive Benchtop LC/MS Orbitrap Mass Spectrometer," Thermo Scientific Application Note 30162, 2008.
(6) AOAC Official Method 2007.01. "Pesticide Residues in Foods by Acetonitrile Extraction and Partitioning with Magnesium Sulfate. Gas Chromatography/Mass Spectrometry and Liquid Chromatography/Tandem Mass Spectrometry."
New Study Uses MSPE with GC–MS to Analyze PFCAs in Water
January 20th 2025Scientists from the China University of Sciences combined magnetic solid-phase extraction (MSPE) with gas chromatography–mass spectrometry (GC–MS) to analyze perfluoro carboxylic acids (PFCAs) in different water environments.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ion used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
A Guide To Finding the Ideal Syringe and Needle
January 20th 2025Hamilton has produced a series of reference guides to assist science professionals in finding the best-suited products and configurations for their applications. The Syringe and Needle Reference Guide provides detailed information on Hamilton Company’s full portfolio of syringes and needles. Everything from cleaning and preventative maintenance to individual part numbers are available for review. It also includes selection charts to help you choose between syringe terminations like cemented needles and luer tips.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Oasis or Sand Dune? Isolation of Psychedelic Compounds
January 20th 2025Magic mushrooms, once taboo, have recently experienced a renaissance. This new awakening is partially due to new findings that indicate the effects of psilocybin, and its dephosphorylated cousin psilocin may produce long lasting results for patients who might be struggling with anxiety, depression, alcohol and drug abuse, and post-traumatic stress disorder. Hamilton Company has developed a methodology for the isolation and identification of 5 common psychedelic compounds used in the potential treatment of disease. The PRP-1 HPLC column resin remains stable in the harsh alkaline conditions ideal for better separations.