Special Issues
Smaller particle size stationary phases have seen increased use in the modern analytical laboratory due to their ability to exhibit greater resolution while enabling the use of smaller column geometries. Shorter column lengths and narrower bores allow for faster run times and lower flow rates, decreasing laboratory costs in terms of solvent consumption and man-hours needed to run the analyses. This application note examines the effects of smaller particle sizes in various column geometries on size-exclusion chromatography (SEC) methods and the method variables affecting the characterization of antibody products.
Smaller particle size stationary phases have seen increased use in the modern analytical laboratory due to their ability to exhibit greater resolution while enabling the use of smaller column geometries. Shorter column lengths and narrower bores allow for faster run times and lower flow rates, decreasing laboratory costs in terms of solvent consumption and man-hours needed to run the analyses. This application note examines the effects of smaller particle sizes in various column geometries on size-exclusion chromatography (SEC) methods and the method variables affecting the characterization of antibody products.
This work focused on scaling-down a monoclonal antibody SEC method, starting with a typical analytical-size SEC column (YMC-Pack Diol 5 µm, 300 Å, 300 ??8.0 mm) and ending with a small particle, shorter length, narrower-bore column (YMC-pack Diol 2 µm, 300 Å, 150 ??4.6 mm). The purpose of this experiment was to detail the differing performance characteristics exhibited by each column as physical properties were altered, one variable at a time.
Sample: 1 mg/mL Avastin in PBS
Mobile Phase: 100 mM sodium phosphate with
200 mM sodium chloride at pH=7.0
Column Temp: 25 °C
Flowrate: 1.0 mL/min for 8.0 mm I.D columns
0.33 mL/min for 4.6 mm I.D. columns
Inj. Volume: 30 µL for 8.0 mm I.D. columns
10 µL for 4.6 mm I.D. columns
HPLC System: Agilent 1260
Changes in column geometry were addressed by scaling flow rate (1.0 mL/min to 0.33 mL/min) and injection volume (30 µL to 10 µL) accordingly as inner diameter was decreased from 8.0 mm to 4.6 mm. This allows for a method that is typically run at 1.0 mL/min for 15 min to be run at 0.33 mL/min for 10 min, cutting run time by 1/2 and solvent consumption by a factor of 6x. For resolution comparability see Table I.
As the results indicate, when applied to SEC analysis of monoclonal antibodies, smaller particle stationary phases in shorter columns can increase throughput, while still providing adequate resolution that rivals the standard 300 ? 8.0 mm, 5 µm columns.
941 Marcon Blvd., Suite 201, Allentown, PA 18109
tel. (610) 266-8650, fax (610) 266-8652
Website: ymcamerica.com
Efficient Calibration Curve Generation with Automated Dilution: Part 1
March 4th 2025This application note presents an automated method for generating calibration curves using an autosampler’s dilution function. It offers practical examples demonstrating a simple setup, enhanced productivity, accuracy and reliability, with a focus on eliminating manual dilution for greater efficiency in HPLC analyses.
Efficient Calibration Curve Generation with Automated Dilution: Part 2
March 4th 2025Part 2 delves into advanced applications using organic solvents with the HPLC autosampler dilution function. It details procedures for mixed standard solutions, highlighting automation’s role in reducing errors, saving labor and promoting efficient laboratory practices.