Delivering samples to the analytical column in “clean” mobile phase is important for robust methods and high-quality results.
Dwight R. Stoll, LC Troubleshooting Editor
Delivering samples to the analytical column in “clean” mobile phase is important for robust methods and high-quality results.
Editors’ Note from John Dolan: With this instalment of “LC Troubleshooting,” you’ll begin to see evidence of a transition. I’m retiring as editor of this column in October 2017 and handing the reins over to Dwight Stoll. This is Dwight’s first article during the transition period-you’ll see more of him during the next few months. Many of you will recall that Dwight was the recipient of LCGC’s 2011 Emerging Leader in Chromatography award. He earned his Ph.D. from Peter Carr’s group at the University of Minnesota and is now tenured at Gustavus Adolphus College. I’ve had the privilege of knowing Dwight since his graduate school years and of working with him on an LC column selectivity project. He’s one of the world’s leaders in two-dimensional LC (2D-LC), so I’m sure we’ll all gain some insights into the successful implementation of this exciting technique. Join me in welcoming Dwight onto the LCGC team.
Sometimes the instrumentation we use in liquid chromatography (LC) works remarkably well in spite of our sometimes justified neglect of best practices designed to protect the instrument’s more-delicate components. One of the ways I balanced the checkbook during my graduate school days was doing instrument repair on nights and weekends. I vividly recall receiving an LC pump from a laboratory in Kansas that looked like it had sat out in the middle of a farm field through a Minnesota winter-the electronic boards were full of dust and the front area of the pump where the pump head sits was literally packed with dirt. To my amazement it still pumped liquid, albeit not with the accuracy we would expect.
For this instalment of “LC Troubleshooting” I decided to focus on filters and filtering as they relate to proper, reliable operation of high performance liquid chromatography (HPLC) instruments. John Dolan has written extensively and often on these topics in the past (1–4), and yet many of the problems I discuss with LC practitioners in the field-and with students in my own laboratory-are in some way related to filters and filtering. Technologies, practices, and the demands of our laboratory settings change, too, and so this month I’d like to discuss what has stayed the same and what has changed, while providing some guidance for our everyday work.
In preparation for writing this column, I informally surveyed several people, because I was interested to know what folks are actually doing in their laboratories. I consider all of these people experts in LC, with an average of about 15 years of experience, and they represent a broad cross-section of our community, ranging from pharmaceutical and chemical companies to instrument manufacturers. What was strikingly evident from this survey was that people handle the topic of filters and filtering in diverse ways, even though instrument manuals and books on LC may suggest only one way. Our reality today is that the practice of LC is strongly influenced by a number of drivers that the manuals don’t consider, including minimizing the number of laboratory operations to minimize variation in analytical results, and optimizing the costs of analytical results from a holistic point of view (in other words, analyzing the benefit of component X to the method relative to the cost of the component).
The topic of filters and filtering is very important in LC because a number of components in a LC system are susceptible to premature failure if particulates are present in the mobileâphase flow path. For example, dust or other particulates in the mobile phase pulled from a solvent bottle can cause pump check valves to fail partially (they become leaky) or entirely (the pump produces no flow at all). Perhaps the most obvious failure mode involves particulates from injected samples or instrument components accumulating at the column inlet, resulting in a plugged column. In the following paragraphs I will discuss some of the most important points in an instrument setup and method related to filters and filtration. In each case I will describe best practices from the point of view of the instrument and will also discuss cases in which this advice may not be strictly followed for good reason. We will see that many of these considerations boil down to thinking about risk and benefit in terms of the potential benefit derived from deviating from best practices and the risks associated with doing so.
Filtering to Protect the Pump
From the point of view of an LC pump, ideal solvents are absolutely free of particulate matter. Particulates can interfere with the sealing interface between the check valve ball and seat, causing it to leak or to fail entirely, resulting in inconsistent or no flow from the pump. Particulates can also accelerate wear on pump pistons and seals, which can result in leaks and flow-rate inaccuracy. These problems usually manifest themselves as noisy or irregular baselines, unexpected fluctuations in the operating pressure reported by the pump, and variations in retention time. Most sources of advice on this topic recommend that solvents should be filtered to 0.45 µm before use. Modern LC pumps come equipped with filters that are fitted to the inlet end of solvent lines that pull solvents from their reservoirs, but these filters have much larger porosity and are not designed to retain particles at the 0.5-µm level. Off-line filtration, however, adds cost (both consumables and operator time) and is a potential source of both chemical and particulate (yes, it happens) contamination of the solvent. Reputable manufacturers of solvents sold specifically for use in LC filter their solvents at the point of packaging (typically to 0.2 µm, and some to 0.1 µm), so one way to avoid in-house filtering altogether is to purchase and use premixed solvents. There was a time when choices for these solvents were limited to water and neat organic solvents, but these days most commonly used mobile phases can be purchased from major manufacturers as premixed solutions. For example, 0.1% formic acid in water, 0.1% formic acid in acetonitrile, and 0.1% ammonium acetate in water are all readily available. There is also an incredible array of premade buffers available for HPLC, including phosphate buffers, buffers with ionâpairing reagents added, and so on. The convenience associated with using these premixed solvents comes at a significant increase in purchase price over pure solvents, however, so users must decide if the increased cost is worth the benefits they offer. Some laboratories find that the increased cost of premixed solvents is offset by savings in labour and reduced downtime that would otherwise result from errors in mobile-phase preparation.
Mobile-Phase Additives
The majority of LC methods in use today involve some type of mobileâphase additive, including buffering agents and ion-pairing reagents. If we choose not to use the type of premade solvents described above, then we have to make a choice-do we filter the solvent after adding the additives, or not? At this point there is an opportunity to divide the question over two scenarios-those that involve liquid additives (for example, formic acid, ammonium hydroxide, and phosphoric acid), and those that involve solid additives (for example, potassium phosphate, sodium perchlorate, and sodium chloride). Among the experts I surveyed, very few reported that their laboratories filter solvents involving only liquid additives. I would report the same for my laboratory. The responses from the experts about solvents involving solid additives are mixed. A majority, but not all, of their laboratories routinely filter solvents containing solid additives. In my laboratory we often base our decision on the amount of solid additive used. If the amount of additive is small and it is highly soluble in the solvent-say 10 mM ammonium acetate in water-we don’t filter, but we always filter solutions containing high concentrations of additives. The compelling reasons to filter in these cases include increased risk of particulate contaminants present in the solid additive, particulates present as a result of the incomplete dissolution of the additive, and, in the case of
buffer–organic solvent mixtures, precipitation of the additive upon addition of the organic solvent.
Why or Why Not?
In the face of these compelling reasons to filter, especially when solid additives are used, one might reasonably ask-why not? Here is a short list of reasons, the significance of which often depends on context.
Microbial Contamination
Finally, the topic of microbial growth in solvent reservoirs is important. If microbes begin growing in the solvent container feeding the pump, there is potential for the instrument to be contaminated both physically and chemically. Here, too, some themes emerged from the experts I surveyed. The following best practices will help minimize problems with microbial growth, which is limited to aqueous solutions:
Filtering to Protect the Column
The size of the interstitial spaces between particles of packed columns typically used in LC is a small fraction of the diameter of the particles themselves. Thinking of it this way, it becomes evident that the packed bed itself can act as an effective filter; clearly we want to employ the column for separation, not as a filter. The frits on the ends of the columns serve two purposes then: They retain the particles inside the column, and they prevent particulates carried by the mobile phase from entering and partially blocking flow through the particle bed. Accumulation of particulates on the frit at the inlet side of the column can also lead to plugging of the frit, rendering the column unusable because of the increased pressure drop over the frit. In the preceding section I discussed the mobile phase as it is drawn into the pump as a source of particulates. But even if the mobile phase entering the pump is perfectly particulateâfree, there are several potential sources of particulates entering the flow stream between the pump and the column. Most prominent among these are shedding of pump-seal material (chunks of polymer); shedding of valve rotor seals (again, chunks of polymer), including the sample injection valve and column switching valves; and the sample itself. The topic of particulates in the sample is one that clearly crosses over into the area of sample preparation, and one could easily write an entire column on this topic alone. Here we will focus on the first two potential sources, the instrumentârelated sources of particulates.
In-Line Filters Versus Guard Columns
In the context of a LC instrument, an in-line filter is usually one that is placed in the mobile-phase flow path at some point between the pump outlet and the column inlet. These filters have evolved over the years to the point that the most high performing ones can be used at very high pressures, have small throughpore sizes (down to 0.2 µm), and have internal volumes in the microlitre range so as to limit contributions to peak broadening. The sole purpose of these filters is to collect particulates in the mobile phase as a means of preventing damage to downstream components from these particulates. In my laboratory we do not regularly use in-line filters immediately after the pump, but we are very disciplined about using them between valves with polymeric rotor seals (for example, autosampler and column-switching valves) and the analytical column. We find that these seals shed a lot of particles as they wear, and that capturing them with an in-line filter generally increases the lifetimes of our analytical columns and small diameter capillary connection tubing. I would say that only about 50% of the LC instruments I see in a year in different laboratories have an in-line filter installed ahead of the analytical column, which is very surprising to me. Replacement filters cost anywhere from a few dollars to $25 depending on their features, but it is easy to justify replacing a few of these if we can significantly increase the lifetimes of our analytical columns and improve instrument robustness in general. It is important to note that some applications involving very small peak volumes (for example, small particle sizes and short columns) will be more susceptible to peak broadening inside the in-line filter. In these cases one has to think carefully about which is more important-minimizing dispersion, or protecting the column from particulates.
Guard columns are essentially very short versions of the analytical column (often with the same packing material and diameter, but lengths on the order of 5 or 10 mm) that are screwed into the column inlet. One can argue that guard columns can also serve as particulate filters to protect the analytical column, but this approach is a much more expensive way of doing so (on the order of 10–20% of the column purchase price per guard column). Instead, guard columns are more appropriately used for chemical protection of the analytical column. For example, sample constituents that are very highly retained by the analytical column, or perhaps will not even be eluted under the conditions of the experiment, can be trapped by the guard column. The contaminated guard column can then be replaced with a new one after the analysis of a number of samples at a fraction of the price of a new analytical column. The experts I surveyed reported more frequent use of in-line filters than guard columns, citing the cost of guard columns as being generally too high to justify their use in a uniform way.
Closing Thoughts
In this instalment of “LC Troubleshooting”, I made generalizations where possible about when and why to filter. Of course there are special situations where these generalizations are not applicable. For example, certain types of detectors, such as light-scattering detectors used for polymer analysis, are very sensitive to particulates in the mobile phase and may require more-thorough and moreâconsistent filtering of mobile-phase solvents. LC users should think carefully about special requirements associated with their particular analyses and adjust their approaches to the use of filters and filtering accordingly.
References
Dwight Stoll and John Dolan are the co-editors of LC Troubleshooting, and Stoll is the author of this instalment. Direct correspondence about this column via e-mail to LCGCedit@ubm.com.
Dwight Stoll is Associate Professor and Co-Chair of Chemistry at Gustavus Adolphus College in St. Peter, Minnesota, USA. He has authored or coauthored 48 peerâreviewed publications in separation science and more than 90 conference presentations. His primary research focus is on the development of twoâdimensional liquid chromatography (2D-LC) for both targeted and untargeted analyses. Within this area he has made contributions on the topics of stationaryâphase characterization, new 2D-LC methodologies and instrumentation, and fundamental aspects including reequilibration in gradient elution reversed-phase LC and analyte focusing. He is the 2009 recipient of the John B. Phillips Award for contributions to multidimensional gas chromatography, the 2011 recipient of LCGC’s Emerging Leader in Chromatography Award, and the 2015 recipient of the American Chemical Society Division of Analytical Chemistry Award for Young Investigators in Separation Science.
Best of the Week: Food Analysis, Chemical Migration in Plastic Bottles, STEM Researcher of the Year
December 20th 2024Top articles published this week include the launch of our “From Lab to Table” content series, a Q&A interview about using liquid chromatography–high-resolution mass spectrometry (LC–HRMS) to assess chemical hazards in plastic bottles, and a piece recognizing Brett Paull for being named Tasmanian STEM Researcher of the Year.
Using LC-MS/MS to Measure Testosterone in Dried Blood Spots
December 19th 2024Testosterone measurements are typically performed using serum or plasma, but this presents several logistical challenges, especially for sample collection, storage, and transport. In a recently published article, Yehudah Gruenstein of the University of Miami explored key insights gained from dried blood spot assay validation for testosterone measurement.
Determination of Pharmaceuticals by Capillary HPLC-MS/MS (Dec 2024)
December 19th 2024This application note demonstrates the use of a compact portable capillary liquid chromatograph, the Axcend Focus LC, coupled to an Agilent Ultivo triple quadrupole mass spectrometer for quantitative analysis of pharmaceutical drugs in model aqueous samples.