With recent advances in technology and the use of ultra high pressure chromatography systems becoming more commonplace, it is of use to compare reproducibility of retention time and peak area of UHPLC vs HPLC.
With recent advances in technology and the use of ultra high pressure chromatography systems becoming more commonplace, it is of use to compare reproducibility of retention time and peak area of UHPLC vs HPLC.
Table I: Reproducibility of retention times and peak area (n515)
The X-LC utilized in this experiment was a JASCO X-LC system consisting of a 3185PU pump, capable of pumping 15,000 PSI, 3080DG degasser, 3067CO column, 3070UV UV/Vis detector, 3159AS autosampler allowing fast cycle time and low carryover with the ability to operate at very high pressures, column X-PressPak C18S (2.1 mm × 50 mm), 2 micron particle size and EZChrom Elite chromatography data system, vers.3.2.1.
A standard mixture contains acetaminophen, caffeine, aspirin, and benzoic acid.
The chromatogram in Figure 1 is a series of 15 injections overlaid. Excellent reproducibility is obtained on the X-LC system as well as a shortened analysis time of one min.
Table I lists the retention time and peak area Relative Standard Deviation (RSD) of retention time. The RSD is in a range between 0.09–0.26% and that of peak area in a range between 0.38–0.86%. These results demonstrate that UHPLC and the use of X-LC hardware as being extremely reproducible and reliable, and with confidence one can replace any HPLC analysis with this technique.
Figure 1
JASCO, Inc.