LCGC North America
It is often possible to achieve better sensitivity and lower limits of detection and quantitation using standard gas chromatography instrumentation. Paying attention to fundamental variables-like sample solvent, initial oven temperature hold, carrier gas operating mode, and thermal gradients- can lead to much improved method performance.
It is often possible to achieve better sensitivity and lower limits of detection and quantitation using standard gas chromatography (GC) equipment, such as a standard split–splitless injection port and a flame ionization detector (FID). Paying attention to fundamental variables can lead to much improved method performance, Here, we consider some important aspects necessary to optimize GC–FID performance.
By choosing a sample solvent that matches the stationary phase polarity as closely as possible, peak shape will be optimized with no tailing or broadening, which might otherwise reduce the signal-to-noise ratio. Here, we think of choosing nonpolar solvents such as n-hexane with a non-polar stationary phase such as 1% polydimethyl siloxane (DB1, RTX-1) or methanol with polar stationary phases such as waxes.
The inlet liner is designed to hold a certain volume of gas (typically 500 to 950 mL), and we need to avoid overfilling the liner with sample gas, or problems will occur such as poor quantitative reproducibility and carry over. By choosing a sample solvent with a lower expansion coefficient, and increasing the pressure within the liner during the injection, to constrain the volume of sample gas created, larger injection volumes can be made. Calculators, typically known as backflash or vapor volume calculators, can be used to determine the maximum sample volume that can be injected before liner overfilling occurs.
In splitless injection mode, the split vent valve is closed, and all of the contents of the inlet are directed into the GC column. After a predetermined length of time (the splitless time), the split vent value is opened, and the remaining inlet contents are vented to the atmosphere. This helps to prevent wide and badly tailing solvent peaks, increased baseline noise, and a rising baseline during the analysis. If the split vent is opened too early, there is a risk of analyte loss, and, if opened too late, the broad and tailing solvent peak may reduce the sensitivity of early eluting analytes; and noisier baselines will reduce the signal-to-noise ratio for all analytes. One must experimentally determine the optimum splitless time to achieve the best reproducibility and sensitivity from the splitless injection. For proper solvent focussing to occur, the initial oven temperature must be maintained around 20 °C below the boiling point of the sample solvent during the injection phase of the analysis.
On-column sample focusing during splitless injection takes a finite amount of time, and, during this time, the oven temperature must be held constant. However, if this initial hold time is too long, some degree of analyte dispersion may occur, reducing the peak efficiency, and, hence, the signal-to-noise ratio. The initial hold time should be carefully evaluated if optimum sensitivity is to be achieved.
Shorter columns (10–15 m) with narrow i.d. (0.18–0.25 mm) and thin films (<0.3 mm) will give the best peak efficiencies, and therefore the optimum signal-to-noise ratio. Further, less-polar stationary phase chemistries will show less inherent bleed, again improving the signal-to-noise ratio, although the choice of stationary phase chemistry will be sample dependent. One should choose the least polar column with the thinnest film that results in a satisfactory separation, in order to reduce column bleed to the minimum amount possible.
Ensure that the carrier gas operating mode is set to constant flow, rather than constant pressure, to ensure that the carrier flows at the same linear velocity during the whole temperature program. This approach prevents the carrier from slowing as the temperature increases, which can lead to broadening of later-eluted peaks and suboptimal detector response.
Optimize the fuel (hydrogen) to oxidizer (air) ratio of the FID to ensure the best response for your analytes; typically, start with a 10:1 ratio and adjust the fuel gas in steps of +/- 5 mL/min. Also, ensure that the make-up gas flow rate (nitrogen is recommended by many manufacturers as the most effective make-up gas) is optimized, and note that the make-up gas flow rate can have a pronounced effect on analyte sensitivity. Start with a 1:1 ratio of make-up gas to fuel gas (typically hydrogen), and adjust in +/- 5 mL/min. steps to investigate the optimum range.
Ballistic (short and fast) thermal gradients will produce the sharpest peaks and the best signal-to-noise ratios, but ensure the oven heater motor can follow the required temperature profile in a reproducible way. Of course, the use of ballistic thermal gradients will depend upon the separation requirements and the complexity of the sample.
AI and GenAI Applications to Help Optimize Purification and Yield of Antibodies From Plasma
October 31st 2024Deriving antibodies from plasma products involves several steps, typically starting from the collection of plasma and ending with the purification of the desired antibodies. These are: plasma collection; plasma pooling; fractionation; antibody purification; concentration and formulation; quality control; and packaging and storage. This process results in a purified antibody product that can be used for therapeutic purposes, diagnostic tests, or research. Each step is critical to ensure the safety, efficacy, and quality of the final product. Applications of AI/GenAI in many of these steps can significantly help in the optimization of purification and yield of the desired antibodies. Some specific use-cases are: selecting and optimizing plasma units for optimized plasma pooling; GenAI solution for enterprise search on internal knowledge portal; analysing and optimizing production batch profitability, inventory, yields; monitoring production batch key performance indicators for outlier identification; monitoring production equipment to predict maintenance events; and reducing quality control laboratory testing turnaround time.