The analysis of polar compounds in support of clinical and preclinical pharmacokinetic studies requires an analytical methodology capable of achieving ultra-low detection and quantification limits. The high sensitivity afforded by coupling HPLC with tandem mass spectrometry (MS-MS) has made it the technique of choice in this environment, but it is subject to the following limitations when reversed phase liquid chromatography (RPLC) is used:
The analysis of polar compounds in support of clinical and preclinical pharmacokinetic studies requires an analytical methodology capable of achieving ultra-low detection and quantification limits. The high sensitivity afforded by coupling HPLC with tandem mass spectrometry (MS-MS) has made it the technique of choice in this environment, but it is subject to the following limitations when reversed phase liquid chromatography (RPLC) is used:
1. Polar compound elution in a highly aqueous mobile phase: Gas phase ion generation is facilitated with more volatile solvents (1); therefore, the high water content used for RPLC retention of polar compounds decreases desolvation efficiency. In addition the highly aqueous mobile phase required for retention of polar compounds increases solvent surface tension, which decreases ESI spray stability (2).
2. Polar compound elution in the primary ion suppression region: Poorly retained compounds in RPLC increase the likelihood of analyte compounds eluting with the many endogenous compounds present in bioanalytical samples. These endogenous compounds compete for charge in the source and can suppress analyte ionization, reducing sensitivity and degrading LOD and LOQ. It is critical that a robust bioanalytical method retains analytes outside the primary ion suppression region (3).
Luna HILIC columns allow for improved sensitivity in the analysis of complex bioanalytical samples by creating more favorable conditions for polar compound retention and ionization.
On the Luna HILIC column, atenolol elutes in a high organic volatile mobile phase, which facilitates analyte desolvation and results in enhanced LC–MS-MS sensitivity. Polar metabolites are weakly retained under RPLC conditions and elute near the solvent front, often in the "primary ion suppression region." Under HILIC chromatographic conditions, where polar analytes are retained using high organic mobile phase, Luna HILIC strongly retains polar compounds that are otherwise weakly retained in RPLC.
The Luna HILIC column retains and elutes hydrophilic/polar compounds in highly organic mobile phase conditions, which improve analyte desolvation efficiency and ESI spray stability. The combination of these two improvements in the MS interface will often provide improved polar analyte detection and quantitation.
Figure 1
In HILIC mode, the analyte elution in the primary ion suppression region is avoided as a result of the increased retention of hydrophilic compounds. This increased retention for polar compounds regularly results in less interference from endogenous sample compounds, improved sensitivity and accuracy, and reduced LOD and LOQ. The Luna HILIC column meets the requirements of a robust bioanalytical method by eluting analytes outside the primary ion suppression region.
It has been shown here that Luna HILIC columns can improve LC–MS-MS sensitivity of polar compounds in biological matrices in support of clinical and preclinical pharmacokinetic studies where ultra-low detection and quantification limits are highly desirable.
Figure 2
1) Cech, N.B.; Enke, C.G. Mass Spectrom. Rev. 2001, 20, 362–387.
2) Kerbarle, P. J. Mass. Spectrom. 2000, 35, 804–817.
3) Matuszewski B.K.; Constanzer M.L.; Chavez-Eng C.M., Anal. Chem.
2003, 75, 3019–3030.
Phenomenex, Inc.
411 Madrid Ave.
Torrance, CA 90501
(310) 212-0555; (310) 328-7768
High pH and Temperature with HALO® ELEVATE C18, 1.5 mm ID Column
December 11th 2024Over the course of 5000 injections, the HALO®Elevate C18 1.5 mm ID column demonstrates stability and robustness in the stationary phase and the column hardware, with no significant changes to the retention time, peak shape, and back pressure.
Separation of Antidepressants Using HALO 90 Å PCS C18 Compared to C18
December 11th 2024A mix of four antidepressants is separated using the HALO®PCS (positive charged surface) C18 column demonstating the improved tailing factor and efficiency when compared to a traditional (uncharged) C18 stationary phase.
Advantage of Fused-Core® Particle Technology for Oligonucleotide Separations
December 11th 2024The HALO® OLIGO C18 2.7 µm column outperformed an FPP Oligo C18 column in oligonucleotide separations, achieving faster separations while maintaining efficiencies comparable to a 1.9 µm FPP with more than half the backpressure.