Tips to improve signal-to-noise ratio when using standard gas chromatography (GC) equipment.
Tips to improve signal-to-noise ratio when using standard gas chromatography (GC) equipment
It is often possible to achieve better sensitivity and lower limits of detection and quantitation using standard gas chromatography (GC) equipment, such as a standard split–splitless injection port and a flame ionization detector (FID). Paying attention to some of the fundamental variables can lead to much improved method performance, and here we consider some important aspects necessary to optimize GC–FID performance.
Sample Solvent: By choosing a sample solvent that matches as closely as possible the stationary phase polarity, the peak shape will be optimized with no tailing or broadening, which might otherwise reduce the signal-to-noise ratio. Here we think of choosing nonpolar solvents such as n-hexane with nonpolar stationary phase such as 1% polydimethyl siloxane (DB1, RTX-1), or methanol with polar stationary phases such as waxes.
High Head-Pressure (Pressure Pulsed) Injection: The inlet liner is designed to hold a certain volume of gas (typically 500 to 950 mL), and we need to avoid overfilling the liner with sample gas, or problems will occur with poor quantitative reproducibility and carryover. By choosing a sample solvent with a lower expansion coefficient, and increasing the pressure within the liner during the injection, in order to constrain the volume of sample gas created, larger injection volumes can be made. Calculators, typically known as backflash or vapour volume calculators, can be used to determine the maximum sample volume that can be injected before liner overfilling occurs.
Optimize Splitless Time: In splitless injection mode, the split vent valve is closed and all of the contents of the inlet are directed into the GC column. After a predetermined length of time (the splitless time), the split vent value is opened, and the remaining inlet contents are vented to the atmosphere. This helps to prevent wide and badly tailing solvent peaks, increased baseline noise, and a rising baseline during the analysis. If the split vent is opened too early, there is a risk of analyte loss, and, if opened too early, the broad and tailing solvent peak may reduce the sensitivity of early eluting analytes; and noisier baselines will reduce the signal-to-noise ratio for all analytes. One must experimentally determine the optimum splitless time in order to achieve the best reproducibility and sensitivity from the splitless injection. In order for proper solvent focusing to occur, the initial oven temperature must be maintained at around 20 °C below the boiling point of the sample solvent during the injection phase of the analysis.
Initial Oven Temperature Hold: On column sample focusing during splitless injection takes a finite amount of time, and, during this time, the oven temperature must be held constant. However, if this initial hold time is too long, some degree of analyte dispersion may occur, reducing the peak efficiency, and, hence, the signal-toânoise ratio. The initial hold time should be carefully evaluated if optimum sensitivity is to be achieved.
GC Column Choice: Shorter columns (10–15 m) with narrow internal diameters (i.d.) (0.18–0.25 mm) and thin films (< 0.3 mm) will give the best peak efficiencies, and therefore the optimum signal-to-noise ratio. Further, less polar stationary phase chemistries will show less inherent bleed, again improving the signal-to-noise ratio, although the choice of stationary phase chemistry will be sample dependant. One should chose the least polar column with the thinnest film, which results in a satisfactory separation, in order to reduce column bleed to the minimum amount possible.
Carrier Gas Operating Mode: Ensure that the carrier gas operating mode is set to constant flow, as opposed to constant pressure, to ensure that the carrier flows at the same linear velocity during the whole temperature program. This will avoid the carrier slowing as the temperature increases, which can lead to broadening of later eluting peaks, and suboptimal detector response.
FID Optimization: Optimize the fuel (hydrogen) to oxidizer (air) ratio of the FID to ensure the best response for your analytes (typically start with a 10:1 ratio and adjust the fuel gas in steps of +/- 5 mL/min). Also, ensure that the make-up gas flow rate (nitrogen is recommended as the most effective make-up gas by many manufacturers) is optimized, and note that the make-up gas flow rate can have a pronounced effect on analyte sensitivity. Start with a 1:1 ratio of make-up gas to fuel gas (typically hydrogen), and adjust in +/- 5 mL/min steps to investigate the optimum range.
Thermal Gradients: Ballistic (short and fast) thermal gradients will produce the sharpest peaks and the best signalâtoânoise ratios, but ensure the oven heater motor can follow the required temperature profile in a reproducible way. Of course, the use of ballistic thermal gradients will depend upon the separation requirements and the complexity of the sample.
Get the full tutorial at www.CHROMacademy.com/Essentials (free until 20 March).
Measuring Procyanidin Concentration in Wines Using UHPLC
January 24th 2025Researchers from the University of Bordeaux (Villenave d'Ornon, France) report the development and validation of a rapid and quantitative analytical method measuring crown procyanidin concentration in red and white wines using ultra-high performance liquid chromatography (UHPLC) coupled with a ultra-high performance liquid chromatography (Q-TOF) mass spectrometer.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ions used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.