LCGC Europe
Describing an accurate analytical technique for the detection and quantification of water using IL-based capillary GC.
Developing versatile, rapid and accurate analytical techniques for the detection and quantification of water in a variety of materials remains an important and ubiquitous analytical problem. This article describes an effective and sensitive ionic liquid (IL)-based capillary gas chromatography (GC) method with a thermal conductivity detector (TCD) to determine the water content in liquid samples. The open tubular capillary columns, coated with specific ILs developed in this study, increased the sensitivity and significantly increased the ruggedness of this technique. The lower limit of detection (LOD) of this method is ~2 ng water. Samples containing high levels of water are also readily analysed without pretreatment. Conversely, organic solvents can be measured in water by the same approach using either a TCD or a flame ionization detector (FID). Standardization was performed with National Institute of Standards and Technology (NIST) reference materials and the accuracy was compared with another independent method (Karl Fischer titration). The developed method is highly sensitive, fast, and is not affected by interferences and side reactions common with existing Karl Fischer Titration (KFT) methods.
The determination of water content in solvents and consumer products, including foods, pharmaceuticals and industrial materials is of great importance. Analytical testing for the presence and concentration of water is one of the most frequent, important and ubiquitous measurements made in modern industrial society. Thus a versatile, simple and efficient analytical technique for the accurate quantification of water is imperative. The essentially universal presence of water requires accurate, easy and sensitive techniques to quantify it. Various techniques such as gravimetry (1), Karl Fischer Titration [KFT] (2,3), gas chromatography (GC) (4–13), near infra red spectroscopy [NIRS] (4,14–16), solvatochromic sensing (17,19), fluorine nuclear magnetic resonance spectroscopy (F-NMRS) (18), isotope ratio mass spectrometry [IRMS] (19) and others have been reported in the literature, but only few methods are widely accepted and used.
Currently, the most commonly used method for water analysis is the KFT, which was first reported in 1935 (2). In this titrametric method, I2 is reduced to HI in the presence of water (9). There are four components in the Karl Fischer reagent consisting of: iodine, sulphur dioxide, a suitable base (RN), (originally pyridine was used, but now imidazole is more common); and a suitable solvent (methanol, ethanol, diethylene glycol monomethyl ether, etc) (20).
The accepted mechanism of this two-step reaction is (20,21):
The end point is determined potentiometrically. Two types of KFT methods are used: coulometric titration and volumetric titration. The former is used to detect trace amounts of water, ranging from 10 µg to 99 µg (1 ppm – 5%), and it requires ~5 g or more of sample (22). Volumetric titration is used to detect higher water quantities greater than 1 mg (10 ppm – 100%) and the amount of sample required varies from 0.1 mg to 500 mg (23). Therefore, prior knowledge of the approximate amount of water present in the sample is required in choosing the correct KF method of analysis.
Although this is a well established method, interference of side reactions (9), reagent instability (21), sample insolubility (24) and pH issues (4) prevent it from being accepted as a universal method. Variations on the basic KFT methodologies have been developed in an attempt to overcome these problems (3,9,24). However, many issues still remain, not the least of which is that the reagents degrade with time and there is residual water in all KFT reagents. Another applied method for water detection is based on GC. Early attempts using GC were mainly based on packed (molecular sieve) columns, involving both direct detection by a thermal conductivity detector (TCD) (7–9,11,13) and indirect detection (i.e., reacting water with calcium carbide to convert to acetylene) with a flame ionization detection (FID) (5). Peak asymmetry (4), poor sensitivity, poor efficiency, strong adsorption of water and many solvents by the stationary phase (7), overlapping of the water peak by other larger peaks, and the inability to detect higher amounts of water (4) limited its application in many cases. Attempts to eliminate peak asymmetry (4,25) using wide-diameter open tubular columns [OTCs] (6) and capillary columns (10) showed some improvement. Additionally, most conventional capillary column GC stationary phases are degraded by water.
A truly useful, broadly effective capillary GC method for water should meet several criteria including the following:
(1) Water should not alter or degrade the stationary phase, thereby altering retention times and peak shapes.
(2) There must be a considerable difference in the retention of water and most/all organic solvents especially when the solvent peak is very large relative to the water peak.
(3) The water peaks should show good efficiency and symmetry.
(4) The water and solvent chromatogram should have sufficient separation space for an appropriate, baseline separated internal standard (IS).
In this study, we examine an IL-based capillary GC method with TCD for the direct determination of water content in liquid samples. The unique nature of ILs including: high thermal stabilities (26), variable polarities (27,28) and exceptional stability to water and oxygen make them excellent choices as stationary phases for this methodology. As will be shown, open tubular capillary (OTC) columns coated with specific IL stationary phases (developed for water analysis) can increase the sensitivity, while greatly enhancing the applicability and reliability of this technique. Furthermore, analysis times can be decreased to 6 min or even less than 3 min in many cases, and samples with virtually any concentration of water can be analysed. The efficacy of this approach is demonstrated with 50 different solvent samples.
Apparatus
The analysis was performed using a 6890N gas chromatograph (Agilent Technologies Inc., Wilmington, Delaware, USA), equipped with a 7683B series autoinjector, TCD and Chemstation plus software (Rev. B.01.03). An Agilent Technologies 10 µL syringe (5181-1267), was used with the autosampler, while a 10 µL syringe (Hamilton, Reno, Nevada, USA) was used for all manual injections. The fused silica capillary columns were coated with IL stationary phases synthesized as previously reported (26,29). The columns were 30 m long, 0.25 mm internal diameter (i.d.) and 0.2 µm film thickness. KFT was performed using Aquastar V1B volumetric titrator (EM Science, a division of EM industries, Inc., Cherry Hill, New Jersey, USA).
Materials
The water reference material: 8509, moisture in methanol (MeOH, 97 ± 13 ppm water) was obtained from the National Institute of Standards and Technology (NIST, Gaithersburg, Maryland, USA). The 4 Å molecular sieves, hydranal water standard 1.0, tetrahydrofuran (THF) and both the ISs, acetone and acetonitrile, were purchased from Sigma-Aldrich (St Louis, Missouri, USA). Aquastar combititrant 2 and Aquastar methanol were purchased from EMD chemicals (Gibbstowm, New Jersey, USA). The high purity water was obtained by filtering the deionized water with Synergy 185 (Millipore, Billeria, Massachusetts, USA). The testing solvents were from Sigma-Aldrich, Mallinckrodt, EMD, Fisher, Omni solvent and Acros organics (see Supporting Information for more details).
Sample Preparation
The accurate quantification of water was achieved using one of two ISs, either acetone or acetonitrile. Two different ISs were available in case one co-eluted with the analyte solvent under the conditions of the experiment. ISs were dried to less than 10 mg/kg by storing over 30% w/v 3 Å molecular sieves for seven days prior to use. A typical sample preparation involved drying a 2 mL autosampler vial overnight at 130 °C followed by the addition of approximately 1 mL of solvent. The mass of the solvent was recorded using an analytical balance. 5.0 mg of IS was added prior to analysis.
Methods
For all analyses, helium was used as a carrier gas at 1 mL/min. Inlet and detector temperatures were set at 250 °C. Splitless injections were made with a purge flow of 20 mL/min. Similar results are obtained using split mode with 20:1 split ratio. Oven temperatures varied from 40 °C to 120 °C (See Supplemental Table S1, which is available on the online version of the article on www.chromatographyonline.com). For initial screening, the columns were conditioned at 120 °C for 2 h. The sample of interest was injected in the splitless mode (0.2–5 µL) to examine the separation between water and the bulk solvent. This preliminary experiment helped to determine which IS should be used.
Water quantification was achieved by integration of the IS peak and the water peak. The concentration of IS in mg/kg was multiplied by the ratio of water peak to IS peak and the result was divided by the response factor (RF). The resulting number represents the concentration of water in the sample in mg/kg. Each solvent was prepared in triplicate and each individual sample was injected in triplicate for a total of n = 9 individual integrations for each solvent analysed.
Detection Limit and Quantification Limit
The limit of detection (LOD) and the limit of quantification (LOQ) were calculated according to the guidelines of the US Food and Drug Administration (FDA) (30,31), using the following equations:
where,σ = the standard deviation (SD) of the response and S = the slope of the calibration plot.
The σ is normally obtained from the SD of the blank sample (31). As it is impossible to obtain a sample without water, the first point of the calibration plot, where there is no added water is used as the blank sample and its SD was used as σ. The slope was obtained from the regression analysis of the plot; peak area ratio of water and IS versus the amount of water.
Figure 1
Figure 1 shows the structures of the three ILs that were used as stationary phases in this study. Generally, IL stationary phases containing trifluoromethylsulphonate (TfO– ) anions resulted in more symmetric water peak shapes than those that contained PF6– , BF4– or bis[(trifluoromethyl)sulphonyl]imide (NTf2– ), with the same cation (data not shown). In addition to the IL-based GC columns, a commercially available polyethylene glycol (PEG) column was also evaluated for comparison.
Figure 2 shows the GC separations of small amounts of water impurities from two of the organic solvents that were analysed. Note that the water peak can be eluted before or after the large organic solvent peak, depending on the relative elution order of the two on the IL-based stationary phases [Figures 2(a–c)]. In most cases the "separation window" between the water and solvent peaks is sufficiently large [Figure 2(a)] that a thermal gradient can be used to further narrow the peak width of water and reduce analysis times [Figure 2(b)]. A comparison of the separation of water from methylene chloride on an IL column versus a commercial PEG type column (see Experimental) is shown in Figure 3. Note that the IL-based column [Figure 3(a)] produced better peak shapes and selectivity even though the separation conditions were optimized for the PEG column. In all cases (for all solvent samples) the IL-based separations were substantially better in terms of selectivity and efficiency.
Figure 2
Figure 3
Table 1 compares the amount of water measured in ten solvents and a NIST methanol standard using four different capillary GC methods and the KFT. The samples were chosen so that the KFT analysis could be done without the use of special reagents (For example, as are needed to quantify water in aldehydes and ketones) (3). Also, most of these samples could be analysed on the commercial PEG column. As can be seen from the data, the IL-based columns usually produced the most precise and accurate results (as indicated from the relative standard deviations (RSDs) and NIST standard respectively, Table 1). It should be noted that no more than one microlitre of sample was used for any capillary GC determinations. Conversely, the KFTs used between 5–15 g of sample depending on the water content of the solvents (larger samples were required for samples containing the least water) (Table 1). Figure 4 shows a standard addition plot for the determination of water in SRM 8509. As can be seen, standard addition is a viable alternative to determine water content if a RF is not available. High levels of accuracy and precision are possible when using a calibrated syringe.
Table 1: Comparison of the ionic liquid GC results with the best results from the Karl Fischer Titration (KFT) and the use of a commercial PEG GC column.
The LODs and LOQs for the IL columns were better than those found for the commercial PEG column, even when analysing samples and using conditions that are necessary for favourable separations on the PEG column. The limit of detection (LOD) of coulometric KFT is 10 µg, and it requires at least 5 g of sample. The IL-based GC method required only 0.2 µL of sample to obtain a much lower detection limit (~2.0 ng or ~5000× greater sensitivity).
Figure 4
Table 2 lists the water content of 50 different solvents as determined with the three IL columns evaluated in this study. Figure 5 indicates the number of these solvents that could be successfully analysed on each column. To be considered a successful separation, the water (analyte) peak had to be separated from both the solvent and at least one of the ISs (i.e., acetone or acetonitrile). All three IL columns could be used to quantify water in 38 out of the 50 solvents (Figure 5). The HAIM-PEG-TFO overall was the most successful as it could be used to quantify water in 47 of the solvent samples, including two samples that were not amenable to separation on the other tested columns.
Table 2
The TTP TfO– column was used to quantify water in 42 solvents, including pyridine and 1,2-dichlorobenzene, which were not possible using the other two columns that were investigated. The DMIM-PEG TfO2– column often gave the best separation window between the water peak and solvent peaks, compared with the other columns. Hence all the separations that are possible on this column can be done at even higher temperatures that those used in this report (Table 3) or using temperature gradients [Figure 2(b)] if desired. In most cases this means that many analyses times can be less than 3 min (results not shown).
Figure 5
As capillary columns containing IL stationary phases are effective in the measurement of small amounts of water in organic solvents, it would not be surprising if they also can be used to determine trace organic solvents in water. Such a determination is shown in Figure 6. Note that in Figure 6(a), TCD is used to show the elution of the water peak relative to trace organic solvents. In Figure 6(b) even lower concentrations of the organic contaminants are seen when using a FID. Of course, water cannot be seen with this detector. It should be noted that the injection of water samples is not recommended on ordinary commercial columns that are not based on ionic liquids. Virtually all traditional commercial columns show appreciable degradation and continuously changing chromatograms when analysing water samples. Analysing water for organic solvents tends to be much more damaging for these columns than analysing organic solvents for trace amounts of water. For example three successive commercial PEG columns had to be used for this study while the separations and conditions of all IL columns used remained unchanged throughout the study (after more than 1600 injections).
Figure 6
A rapid, easy IL-based GC method has been developed for the quantification of water in extremely diverse solvent samples. LODs using this technique are superior when compared with KFT. Furthermore, the IL-GC methodology requires less sample and is free from other complications associated with KFT. The IL-GC method can be used regardless of the chemical nature of the solvent and produces no additional waste products. When compared to other GC methods using commercially available PEG columns, the IL-based columns possessed superior selectivity for water in all solvents tested. Furthermore, they show no degradation or chromatographic changes with time. Typical analysis times ranged from less than 3 min to 7 min. It is expected that this approach could greatly simplify the analysis of water in a variety of applications. The extension of this approach to gases and solid samples is under investigation and will be reported subsequently.
This work was financially supported by the Robert A. Welch Foundation (Y-0026). We would like to thank Professor Zoltan A. Schelly; University of Texas at Arlington, for providing the KFT device.
Daniel W. Armstrong is the Robert A. Welch Professor of Chemistry & Biochemistry at the University of Texas at Arlington. He developed comprehensive solvation and characterization models for ionic liquids (ILs) and pioneered their use in analytical chemistry. He also specializes in the development of modern chiral separations and pseudophase separations. He has over 500 publications and has received over 20 awards for his research and teaching.
Ross Woods is a graduate research assistant at the University of Texas at Arlington.
Ying Zhang is a graduate research assistant at the University of Texas at Arlington. She will graduate in December 2011.
Dilani A. Jayawardhana lives on Pioneer Dr., Irving, TX. She received her Ph.D. in Analytical Chemistry from the University of Texas at Arlington in 2010.
Chunlei Wang is a Research Investigator I at Bristol Meyers Squibb and Company, Plainsboro, NJ, U.S.A. He received his Ph.D. in Analytical Chemistry from the University of Texas at Arlington in 2009.
1. G.M. de Jong et al., Clin. Chim. Acta, 166, 187–194 (1987).
2. K. Fischer, Angew. Chem., 48, 394–396 (1935).
3. E. Scholz, Anal. Chem., 57, 2965–2971 (1985).
4. H.G Streim, E.A. Boyce and J.R. Smith, Anal. Chem., 33, 85–89 (1961).
5. H.S. Knight and F.T. Weiss, Anal. Chem., 34, 749–751 (1962).
6. E.R. Quiram, Anal. Chem., 35, 593–595 (1963).
7. J.M. Hogan, R.A. Engel and H.F. Stevenson, Anal. Chem., 42, 249–252 (1970).
8. J.C. MacDonald and C.A. Brady, Anal. Chem., 47, 947–948 (1975).
9. T.E. Houston, Met. Finish., 95, 36–38 (1997).
10. J. Jalbert, R. Gilbert, and P. Tetreault, Anal. Chem., 71, 3283–3291 (1999).
11. R. Nussbaum et al., Chromatographia, 51, 119–121 (2000).
12. C. Lee and S. Chang, Atmos. Environ., 36, 1883–1894 (2002).
13. W.K. O'Keefe, F.T.T. Ng and G.L. Rempel, J. Chromatogr. A, 1182, 113–118 (2008).
14. P.F. Vornheder and W.J. Brabbs, Anal. Chem., 42, 1454–1456 (1970).
15. M. Li and G.E. Pacey, Talanta, 44, 1949–1958 (1997).
16. X. Zhou et al., Anal. Chem., 70, 390–394 (1998).
17. C. Pinheiro et al., Sens. Actuators B, β14, 978–983 (2006).
18. H. Sun, B. Wang and S.G. DiMagno, Org. Lett., 10, 4413–4416 (2008).
19. E. Jamin et al., J. Agric. Food Chem., 51, 5202–5206 (2003).
20. EMD Chemicals AQUASTAR Tech Notes Fischer Titration Basics.
21. E. Scholz, In Karl Fisher Titration. Determination of water, Springer-Verlag: New York, pp 138 (1984).
22. http://www.jmscience.com/AQ-300-Karl-Fischer-Coulometric-Titrator.htm.
23. http://www.jmscience.com/AQ_2200/AQ-2200.htm.
24. H.D. Isengard and U. Striffler, Fresenius. J. Anal. Chem., 342, 287–291 (1992).
25. H.S. Knight, Anal. Chem., 30, 2030–2032 (1958).
26. K. Huang et al., Anal. Bioanal. Chem., 389, 2265–2275 (2007).
27. P. Sun and D.W. Armstrong, Anal. Chim. Acta, 661, 1–16 (2010).
28. X. Han and D.W. Armstrong, Acc. Chem. Res., 40, 1079–1086 (2007).
29. P.S. Sharma et al., Chem. Mater., 20, 4182–4184 (2008).
3.0. Center for Drug Evaluation and Research (CDER), US Food and Drug Administration (FDA), 1–30 (1994).
31. Center for Drug Evaluation and Research (CDER), US Food and Drug Administration (FDA), 1–10 (1996).
Study Examines Impact of Zwitterionic Liquid Structures on Volatile Carboxylic Acid Separation in GC
March 28th 2025Iowa State University researchers evaluated imidazolium-based ZILs with sulfonate and triflimide anions to understand the influence of ZILs’ chemical structures on polar analyte separation.
Quantifying Microplastics in Meconium Samples Using Pyrolysis–GC-MS
March 26th 2025Using pyrolysis-gas chromatography and mass spectrometry, scientists from Fudan University and the Putuo District Center for Disease Control and Prevention detected and quantified microplastics in newborn stool samples.