This article presents a novel approach for toxicology and pharmacology laboratories that combines multi-compound screening followed by multi-compound determination in a single method. The multi-compound assay developed allows 90 molecules including benzodiazepines, cocaine and related stimulants, amphetamines, and opioids to be measured. The multi-targeted screening (MTS) method was developed using a spectra library containing over 1200 compounds. The approach was evaluated in a routine clinical toxicology laboratory to detect and quantify compounds in unknown samples.
Photo Credit: CA-SSIS/Shutterstock.com
Tiphaine Robin1,2, 1Shimadzu Corporation, Marne-la-Vallée, France, 2Department of Pharmacology and Toxicology, Limoges University Hospital, France
This article presents a novel approach for toxicology and pharmacology laboratories that combines multi-compound screening followed by multi-compound determination in a single method. The multi-compound assay developed allows 90 molecules including benzodiazepines, cocaine and related stimulants, amphetamines, and opioids to be measured. The multi-targeted screening (MTS) method was developed using a spectra library containing over 1200 compounds. The approach was evaluated in a routine clinical toxicology laboratory to detect and quantify compounds in unknown samples.
Screening has been the first analysis to be performed in toxicology when the nature or the presence of a drug is totally unknown, which is often the case in clinical and forensic toxicology (1). Usually, a screening precedes more specific analyses, allowing the quantitation of detected compounds. It is now possible to combine both tasks and to benefit from methods that allow both identification and quantitation.
In addition to reducing the possibility of false defect reporting without compromising accuracy, precision, or limits of detection, the actual liquid chromatography tandem mass spectrometry (LC–MS/MS) instruments could combine the sensitivity of multiple reaction monitoring (MRM) detection with the identification power of a full scan product ion spectrum.
Simple and rapid preparation procedures are also necessary. In this article, a generic sample preparation method based on QuEChERS salts was optimized. QuEChERS (an acronym for “quick, easy, cheap, effective, rugged, and safe”) are notably used in food safety and pesticide analysis. However, a few studies have used this approach in clinical toxicology (2,3,4,5).
The global strategy of this study was: i) to build an exhaustive spectra library of the most pertinent compounds; ii) to develop a generic sample preparation method based on QuEChERS salts; and iii) to test the identification and quantitation performance of an MS acquisition mode combining threshold triggered product ion scan and MRM detection.
Experimental
The chromatographic system consisted of two Shimadzu LC-30 AD pumps (Nexera X2), a CTO 20AC oven, and a SIL-30 AC-MP autosampler (all Shimadzu). Chromatographic separation was performed using a 100 × 2.1 mm, 2.7âµm Raptor biphenyl column (Restek). A Shimadzu LCMS-8060 triple quadrupole mass spectrometer was used.
Spectra Library: Each library spectrum was acquired by authentic standard flow injection at collision energies of 10–55 V. Compounds that ionized efficiently in more than one adduct state were saved resulting in 1507 library entries from 1207 compounds (1278 positive mode, 229 negative mode). Spectral library information was registered for three collision energies corresponding to CE ±10, ±35, and ±55 V. Optimized MRM transitions were also determined for all compounds together with retention time and peak area data ratio of response between quantifier and qualifier transitions. Retention time (RT) analysis included internal standard compounds for relative RT calculation. Compound metadata included: structure, retention time, CAS number, formula, synonyms, compound class or properties, ChemSpider URL and ID number, InChI, and InChIKey.
Sample Preparation Method: A 200âµL measure of acetonitrile was added to 100 µL of whole blood containing internal standard compounds. Samples were vortexed for ~5 s and then incubated at -20 °C for 10 min. A 40-mg measure of QuEChERS salts was added and samples were then vortexed for ~5 s. Samples were then centrifuged (16,000 g, 10 min) and 50-µL of supernatant was removed and added to 150-µL mobile phase A.
MS Acquisition: Two predefined MRM transitions for each compound were monitored. Once the MRM signals were above a set threshold intensity of 10,000 counts, a series of alternating product ion scans were acquired at three collision energies (10, 35, and 55 V) with an automatic exclusion for 3 s after two consecutive MS/MS scans. Product ion spectrum data from each collision energy were combined to create a fragment-rich single spectrum for library searching.
Applications of the Method: Whole blood samples were spiked with a panel of 90 commonly detected compounds reported in toxicological screening and normally measured in separate assays, namely, benzodiazepines, cocaine and related stimulants, amphetamines, and opioids; in addition, antipsychotic compounds were also screened. Calibration curves were prepared in the range 5 µg/L to 2000 µg/L for benzodiazepines, and 5 µg/L to 500 µg/L for cocaine and related stimulants, amphetamines, and opioids. The method performed well in identification of the 90 compounds, regardless of concentration levels: the similarity index (that is, similarity between the spectrum of a detected compound and its spectrum in the library) was above 75–80 in a majority of cases. After determining the concentration of the 90 compounds, performances for quantitation were recorded as accuracy in an 85–115% interval and r2 values above 0.99. Figures 2 and 3 illustrate these performances for two isobaric compounds: hydromorphone and morphine.
To test the viability of this approach and to quantify and identify targets in the different test panels, the MRM-triggered product ion spectrum acquisition method was applied to a series of patient blood samples and compared to a validated LC–MS/MS method using two MRMs for each target compound (2,3). The study concerned blood samples of 24 patients from a diverse range of contexts including overdoses, routine drug testing, driving under the influence of drugs (DUID), and samples provided by patients undergoing psychiatric treatment. For a number of patient samples more than one target compound was measured, resulting in 54 sample points in the regression analysis. Regression analysis showed a close agreement between both methods (slope = 0.9996; regression coefficient r2 > 0.99) (Figure 4).
Discussion and Conclusion
A method was developed for clinical toxicology and forensic analysis using QuEChERS sample preparation, a single LC analysis, and an MRM triggered product ion spectrum acquisition method. The method was designed to meet the needs of a routine clinical toxicology and forensic laboratory by delivering a single approach to the analysis of a diverse range of target panels, and also reducing false defect reporting.
There are two clear advantages of acquiring MRM-triggered product ion spectrum data: (i) a higher confidence in compound identification as a result of library searching, for example, in situations where reference ion ratios can be outside method tolerances or if concentrations measured are below the lower limit of quantitation (LLOQ); (ii) quantitative data is identical to a conventional 2 MRM method.
This was implemented successfully to enable screening and quantitation by MTS for a broadly targeted range of drugs of abuse included in a spectra library containing more than 1200 compounds of interest. Further studies are ongoing to test the capacity of the tool to target this number of compounds in one run.
References
Tiphaine Robin obtained a State Doctorate in pharmacy in 2017 (University of Limoges, France). She is currently a PhD student in analytical chemistry. The main objective of her thesis is to develop innovative LC–MS screening procedures for applications in clinical and forensic toxicology. These works are under the supervision of both the Department of Pharmacology and Toxicology of the Limoges University Hospital (Franck Saint-Marcoux and Souleiman El Balkhi) and Shimadzu Corporation (Stephane Moreau and Alban Huteau, Shimadzu France, Marne-la-Vallée, France). The Mass Spectrometry Business Unit (Overseas) of Shimadzu Corporation, with Neil Loftus and Alan Barnes, is also greatly involved in these developments.
E-mail:tro@shimadzu.fr
Website:www.shimadzu.eu
Measuring Procyanidin Concentration in Wines Using UHPLC
January 24th 2025Researchers from the University of Bordeaux (Villenave d'Ornon, France) report the development and validation of a rapid and quantitative analytical method measuring crown procyanidin concentration in red and white wines using ultra-high performance liquid chromatography (UHPLC) coupled with a ultra-high performance liquid chromatography (Q-TOF) mass spectrometer.
The Next Frontier for Mass Spectrometry: Maximizing Ion Utilization
January 20th 2025In this podcast, Daniel DeBord, CTO of MOBILion Systems, describes a new high resolution mass spectrometry approach that promises to increase speed and sensitivity in omics applications. MOBILion recently introduced the PAMAF mode of operation, which stands for parallel accumulation with mobility aligned fragmentation. It substantially increases the fraction of ions used for mass spectrometry analysis by replacing the functionality of the quadrupole with high resolution ion mobility. Listen to learn more about this exciting new development.
The Complexity of Oligonucleotide Separations
January 9th 2025Peter Pellegrinelli, Applications Specialist at Advanced Materials Technology (AMT) explains the complexity of oligonucleotide separations due to the unique chemical properties of these molecules. Issues such as varying length, sequence complexity, and hydrophilic-hydrophobic characteristics make efficient separations difficult. Separation scientists are addressing these challenges by modifying mobile phase compositions, using varying ion-pairing reagents, and exploring alternative separation modes like HILIC and ion-exchange chromatography. Due to these complexities, AMT has introduced the HALO® OLIGO column, which offers high-resolution, fast separations through its innovative Fused-Core® technology and high pH stability. Alongside explaining the new column, Peter looks to the future of these separations and what is next to come.
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.