Just as medical practitioners are able to discern worrying features from a variety of medical physics devices (electrocardiogram, electroencephalogram, ultrasound, for example), we need to develop the skill to identify worrying symptoms from our HPLC instrument output.
Just as medical practitioners are able to discern worrying features from a variety of medical physics devices (electrocardiogram, electroencephalogram, and ultrasound, for example), we need to develop the skill to identify worrying symptoms from our HPLC instrument output. Medical professionals learn an innate ability to identify critical symptoms (signals) from the noise or random variation in the instrument output, and we need to develop these same skills in order to avoid production of data which is not fit for purpose or instrument failure.
One of the most useful diagnostics in high-performance liquid chromatography (HPLC) is the nature of baseline produced by the detector while the eluent is flowing. While there can be many baseline characteristics such as drift, irregular, or more regulation cycling (pulsations), baseline noise is perhaps the most commonly encountered, and can arise from a variety of different sources. One needs to be aware of what constitutes “normal” baseline as opposed to unusual levels of baseline, depending upon the instrument configuration. Of course, the business imperative is not only to spot problems, but to quickly and efficiently deal with them, and that is the subject of this post.
The S/N of the HPLC output is usually measured as the ratio of the detector signal to the inherent background signal variation and is a useful measure of the “normal” noise within the system. The inherent or background noise is typically measured over a pre-defined portion of the baseline, and most data systems will be capable of making this measurement and reporting the result.
Figure 1: Signal (S) to noise (N) measurement of 5:1.
When inherent or background noise within the system is unusually high, this can affect system performance and will usually result in an increase in the limit of quantitation and issues with reproducible integration. This is why, as chromatographers, we get so worked up about noise levels that are higher than expected.
The smallest detectable signal is usually estimated to be equivalent to three times the height of the average baseline noise. This would give an S/N ratio of 3:1 for the “Limit of Detection” of the detector. If the amount of analyte injected is less than this, then the signal ceases to be distinguishable from noise. For quantitative analysis a S/N ratio of 10:1 is recommended for the “Limit of Quantitation.”
The magnitude of the analyte signal cannot be used in isolation when calculating detector sensitivity, the sensitivity of detection is usually defined in terms of S/N ratio, a measurement of the ratio of the analyte signal to the variation in baseline. S/N measurements are usually carried out by the data system.
Figure 2: Determination of S/N ratio and the independence of absolute signal intensity on signal to noise ratio.
One needs to begin by establishing, preferably for each method and set of instrument conditions, the S/N when the method (or instrument) is performing well and perhaps even set a system suitability performance criterion (usually a range or lower acceptable limit) for the determination.
Of course, the seasoned chromatographer will typically know by glancing at the baseline, whether the inherent noise is “usual,” and this comes only through experience. One should also take care to assess the noise at a reasonable screen magnification or signal attenuation, as any baseline can be made to appear noisy with the correct level of magnification!
Figure 3: Developing a “chromatographer’s-eye” for what is typical (normal) noise and what is unusual. Care should be taken to select a sensible on-screen attenuation for the signal.
However, once again the data system may be able to help us out by reporting what is known as the Peak-to-Peak noise, which may be expressed as absorbance units. This measurement is of the variation in the normal baseline portion, rather than a ratio to the height of a signal and can be very useful at establishing acceptable limits for the background noise. Most HPLC detectors will run a noise test evaluation as part of their initialization routine or can perform a longer test using ASTM criteria with HPLC grade water flowing through the flow cell. Specifications for acceptable noise levels will be given in your manufacturers literature.
Figure 4: Manufacturers test for short term noise in a UV detector (Courtesy of Agilent, Santa Clara, CA).
Although typically associated with detector phenomenon, there are many contributors to the noise within an HPLC, and here we will examine just some of the main culprits. Noise can be both random and period depending upon the nature of the underlying cause of the problem and this difference can, in itself, give us some clues to the nature of the issue.
Detector noise
Electronic and Stray Light Noise
Every detector has associated background electronic and stray light noise, which kept to a minimum through shielding of the optical bench with insulating materials. The in-built tests within the detector firmware should keep this to a minimum, and the instrument will flag if any of these tests are failed on start-up. This is a good reason to cycle the power on the detector every now and then if your laboratory practice is to leave the detector powered on when not in use.
Detector Wavelength and Acquisition Settings
Lower wavelength settings (<220nm) will inherently show an increase in noise that is associated with solvents (MeOH absorbs up to 201 nm) and buffers that lower the amount of light falling onto the photodiode array. The noise of the detector is inversely proportional to the amount of light falling on the photodiodes, and therefore any other factors which decrease the amount of light falling onto the photodiode will also increase the noise within the output. Typically, this can include an ageing lamp or dirty flow-cell windows.
These problems may be overcome in the following ways:
Figure 5: Baseline noise using a methanol/water eluent at 278 nm and 210 nm (1).
Most diode array UV detectors have a slit width setting that describes the width of the slit that focuses light onto the photodiode after passing through the flow cell. As the slit width is increased, the light becomes more diffuse and each wavelength falls over a number of photodiodes (corresponding to the slit width setting). This reduces the noise of the baseline while increasing the signal intensity and so gives greater analytical sensitivity when quantifying analytes. However, the spectral resolution necessary for qualitative analysis (peak track via library matching, for example) is lost, and so it needs a much narrower slit width setting so that the light is less diffuse and each wavelength falls on a smaller number of diodes. When slit width is reduced, the noise level increases, and one needs to set width according the analytical requirements.
Figure 6: The relationship between slit width and baseline noise in diode array detectors.
The detector acquisition rate and data bunching may also have a profound effect on signal noise according to the following relationship:
Where n is the number of data points. Simply put, the more measurements that are taken, the better that we model the random variations associated with instrument response change and the noise becomes better resolved. One needs to carefully balance the S/N ratio with the actual peak to peak noise generated, which is usually achieved using the instrument response or frequency. This needs to be further balanced with the number of data points acquired across each peak which should be in the region of 20–25 or higher for UV data.
Figure 7: Relationship between baseline noise and sampling frequency (data acquisition rate) in UV detection.
As alluded to earlier, the deuterium lamp within the UV has a finite lifetime, and as such can lead to not only an increase of baseline noise, but also to spikes as the lamp arcs against the metal casing onto which the filament is mounted. This gives a baseline appearance similar to that shown in Figure 8, and one needs to consider lamp replacement to judge if the spiking is due to the age of the lamp. Note that other causes of baseline spikes (such as a poorly shielded electrical supply or power board within the detector) are also possible.
Figure 8: Baseline spikes caused by UV detector lamp arcing.
Spikes can be distinguished from ‘real’ peaks as they will have no Gaussian shape when zoomed in.
Other sources of noise
Improper mixing
HPLC UV detectors can be very susceptible to improper mixing of mobile phases which are formed online via either binary or quaternary pumping systems. The absorbance changes due to poor mixing are usually associated with a sinusoidal pulse (covered in a future blog) however at lower pumping volumes or when using additives such as TFA, the disturbances can resemble noise rather than a discernible sinusoidal pulse.
Figure 9: Comparison of UV absorbance signals (214 nm) obtained with different mixers in use. The same pump was used as in Figure 5, but now with a column (30 mm x 2.1 mm i.d. Agilent SB-C18) in use, and gradient elution. Solvent A is 0.1% trifluoroacetic acid in water, solvent B is 0.1% trifluoroacetic acid in acetonitrile, and the gradient runs from 2 to 40% B in 4 min (1).
Figure 9 shows the differences made to the baseline appearance of a 0.1% TFA gradient with various mixer volumes.
Even though modern HPLC instruments are designed with high-efficiency mixers, the addition of a post-market static mixer or simple in-line filter can help to drastically reduce the baseline noise under the circumstances described above. One should note that the additional extra column volume contributed by the fitting of in-line mixers may have a detrimental effect of peak width, especially in UHPLC systems, and one needs to balance any increase in peak width with the reduction in baseline noise when assessing the S/N characteristics of the method.
One should further note that very similar noise characteristics can also be obtained from failing quaternary pump gradient proportioning vales.
Lack of Degassing
If the mobile phase is poorly degassed, small bubbles may outgas from the mobile phase as the eluent enters the flow cell of the detector, due to the pressure change that occurs. This “frothing” of the mobile phase can cause significant baseline noise, and while detector flow cells are designed with a back-pressure restriction in order to reduce the pressure reduction on entering the flow cell, this can be ineffective when the eluent is not properly degassed.
Figure 10: Baseline noise caused by eluent out-gassing the flow cell of a UV detector.
One needs to ensure that the eluent is properly degassed using on-line degassing modules and in some cases, by degassing using vacuum methods prior to mounting onto the HPLC system.
Dewetting
If the HPLC column contains residual packing solvents or has been used with immiscible solvents (for example in normal phase mode), then improper flushing of the column can lead to increased baseline noise associated with dewetting phenomenon.
Equilibrate the column for several hours at the method flow rate in order to completely flush the column of any immiscible solvents.
References
Tony Taylor is the technical director of Crawford Scientific and ChromAcademy. He comes from a pharmaceutical background and has many years research and development experience in small molecule analysis and bioanalysis using LC, GC, and hyphenated MS techniques. Taylor is actively involved in method development within the analytical services laboratory at Crawford Scientific and continues to research in LC-MS and GC-MS methods for structural characterization. As the technical director of the CHROMacademy, Taylor has spent the past 12 years as a trainer and developing online education materials in analytical chemistry techniques.
The LCGC Blog: Historical (Analytical) Chemistry Landmarks
November 1st 2024The American Chemical Society’s National Historic Chemical Landmarks program highlights sites and people that are important to the field of chemistry. How are analytical chemistry and separation science recognized within this program?